CARESCAPE Monitor B850, a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility. Product Usage: The CARESCAPE Monitor B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B850 system is indicated for monitoring of Hemodynamic (including ECG, ST Segment, Arrhythmia Detection, ECG Diagnostic Analysis and Measurement, Invasive Pressure, Noninvasive Blood Pressure, Pulse Oximetry, Cardiac Output, Temperature, Impedance Respiration and SvO2 (Mixed Venous Oxygen Saturation)), Airway Gases (Fi/Et C02, 02, N20 and Anesthetic Agent), Spirometry, Gas Exchange (02 Consumption (VO2), C02 production (VCO2), energy expenditure (BE), and respiratory quotient (RQ)) and neurophysiological (including electroencephalography (EEG), Entropy, Bispectral Index (BIS) and Neuromuscular Transmission (NMT) Monitoring) status. The CARESCAPE Monitor B850 provides alarms, trends, snapshots and events, and calculations and can be connected to displays, printers and recording devices. The CARESCAPE Monitor B850 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for bed to bed viewing and to data management software devices via a network. The CARESCAPE Monitor B850 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility. In addition to the healthcare practitioner, the CARESCAPE Monitor B850 is designed to provide configuration and troubleshooting capabilities to qualified service personnel.
Published February 12, 2014 · Recall initiated: December 20, 2013
Summary
FDA records medical device Class II recall Z-0927-2014, published February 12, 2014 and initiated December 20, 2013. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 12, 945 units (6,239 US, 6,706 OUS); Lot: serial numbers: SEG09270517HX, SEG09270549HX, SEG09270551HX, SEG09270553HX, SEG09270554HX, SEG09270557HX, SEG09270559HX, SEG09270561HX, SEG09270562HX, SEG09270563HX, SEG09270564HX, SEG09270565HX, SEG09270566HX, SEG09270567HX, SEG09270568HX, SEG09270569HX, SEG09270583HX, SEG09270584HX, SEG09270585HX, SEG09270587HX, SEG09270588HX, SEG09270589HX, SEG09270591HX, SEG09270592HX, SEG09280596HX, SEG09280597HX, SEG09280598HX, SEG09280599HX, SEG09280600HX, SEG09280601HX, SEG09280602HX, SEG09280603HX, SEG09280604HX, SEG09280605HX, SEG09280606HX, SEG09280607HX, SEG09280608HX, SEG09280609HX, SEG09280610HX, SEG09280611HX, SEG09280612HX, SEG09280613HX, SEG09280614HX, SEG09280615HX, SEG09280616HX, SEG09280617HX, SEG09280618HX, SEG09280619HX, SEG09280620HX, SEG09280621HX, SEG09280622HX, SEG09280623HX, SEG09280624HX, SEG09280625HX, SEG09310656HX, SEG09310657HX, SEG09310658HX, SEG09310659HX, SEG09310660HX, SEG09310661HX, SEG09310662HX, SEG09310663HX, SEG09310664HX, SEG09310665HX, SEG09310666HX, SEG09310667HX, SEG09310668HX, SEG09310669HX, SEG09310670HX, SEG09310671HX, SEG09310672HX, SEG09310673HX, SEG09310674HX, SEG09310675HX, SEG09310696HX, SEG09310697HX, SEG09310698HX, SEG09310699HX, SEG09310700HX, SEG09310701HX, SEG09310702HX, SEG09310703HX, SEG09310704HX, SEG09310705HX, SEG09310706HX, SEG09310707HX, SEG09310708HX, SEG09310709HX, SEG09310710HX, SEG09310711HX, SEG09310712HX, SEG09310713HX, SEG09310714HX, SEG09310715HX, SEG09310716HX, SEG09310717HX, SEG09310718HX, SEG09310719HX, SEG09310720HX, SEG09310721HX, SEG09310722HX, SEG09310723HX, SEG09310724HX, SEG09310725HX, SEG09310726HX, SEG09310727HX, SEG09310728HX, SEG09310729HX, SEG09310730HX, SEG09310731HX, SEG09310732HX, SEG09310733HX, SEG09310734HX, SEG09310735HX, SEG09310736HX, SEG09310737HX, SEG09310738HX, SEG09310739HX, SEG09310740HX, SEG09310741HX, SEG09310742HX, SEG09310743HX, SEG09310744HX, SEG09310745HX, SEG09320797HX, SEG09320798HX, SEG09320799HX, SEG09320800HX, SEG09320801HX, SEG09320802HX, SEG09320803HX, SEG09320804HX, SEG09320805HX, SEG09320806HX, SEG09320807HX, SEG09320808HX, SEG09320809HX, SEG09320810HX, SEG09320811HX, SEG09320812HX, SEG09320813HX, SEG09320814HX, SEG09320815HX, SEG09320816HX, SEG09320817HX, SEG09320818HX, SEG09320819HX, SEG09320820HX, SEG09320821HX, SEG09320822HX, SEG09320823HX, SEG09320824HX, SEG09320825HX, SEG09320826HX, SEG09320827HX, SEG09320828HX, SEG09320829HX, SEG09320830HX, SEG09320831HX, SEG09320832HX, SEG09320833HX, SEG09320834HX, SEG09320835HX, SEG09320836HX, SEG09320837HX, SEG09320838HX, SEG09320839HX, SEG09320840HX, SEG09320841HX, SEG09320842HX, SEG09320843HX, SEG09320844HX, SEG09320845HX, SEG09320846HX, SEG09320847HX, SEG09320848HX, SEG09320849HX, SEG09320850HX, SEG09320851HX, SEG09320852HX, SEG09320853HX, SEG09320854HX, SEG09320855HX, SEG09320856HX, SEG09320857HX, SEG09320858HX, SEG09320859HX, SEG09320860HX, SEG09320861HX, SEG09320862HX, SEG09320863HX, SEG09320864HX, SEG09320865HX, SEG09320866HX, SEG09320867HX, SEG09320868HX, SEG09320869HX, SEG09320870HX, SEG09320871HX, SEG09320872HX, SEG09320873HX, SEG09320874HX, SEG09320875HX, SEG09320876HX, SEG09320877HX, SEG09320878HX, SEG09320879HX, SEG09320880HX, SEG09320881HX, SEG09320882HX, SEG09320883HX, SEG09320884HX, SEG09320885HX, SEG09320886HX, SEG09320887HX, SEG09320888HX, SEG09320889HX, SEG09320890HX, SEG09320891HX, SEG09320892HX, SEG09320893HX, SEG09320894HX, SEG09320895HX, SEG09320896HX, SEG09360898HX, SEG09360899HX, SEG09360900HX, SEG09360901HX, SEG09360902HX, SEG09360903HX, SEG09360904HX, SEG09360905HX, SEG09360906HX, SEG09360907HX, SEG09360908HX, SEG09360909HX, SEG09360910HX, SEG09360911HX, SEG09360912HX, SEG09360913HX, SEG09360914HX, SEG09360915HX, SEG09360916HX, SEG09360917HX, SEG09360918HX, SEG09360919HX, SEG09360920HX, SEG09360921HX, SEG09360922HX, SEG09360923HX, SEG09360924HX, SEG09360925HX, SEG09360926HX, SEG09360927HX, SEG09360928HX, SEG09360929HX, SEG09360930HX, SEG09360931HX, SEG09360932HX, SEG09360933HX, SEG09360934HX, SEG09360935HX, SEG09360936HX, SEG09360937HX, SEG09360938HX, SEG09360. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a potential for communication loss associated with the F5-01 Frame when connected to CARESCAPE Patient Data Module (PDM) in the CARESCAPE Monitor B850. The PDM patient parameter acquisition module can sometimes spontaneously lose the communication with the host patient monitor when attached to the F5-01 Frame due to an issue with the PDM Ethernet communication hardware. The communicatio
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Worldwide Distribution - US Nationwide including Puerto Rico and the countries of: Australia, Austria, Azerbaijan, Bahrain, Belgium, BRAZIL, Canada, CHILE, COSTA RICA, Cyprus, Czech Republic, Denmark, ECUADOR, Egypt, Estonia, Finland, France, Germany, Greece, GUAM, India, INDONESIA, Iraq, Ireland, Israel, Italy, Japan, Kazakhstan, Korea, Republic of, Kuwait, Lithuania, Malaysia, Malta, Mexico, NEPAL, Netherlands, New Zealand, Norway, Pakistan, PANAMA, PARAGUAY, PERU, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tanzania, United Republic of, Turkey, United Arab Emirates, United Kingdom, URUGUAY.
- Lot
- serial numbers: SEG09270517HX, SEG09270549HX, SEG09270551HX, SEG09270553HX, SEG09270554HX, SEG09270557HX, SEG09270559HX, SEG09270561HX, SEG09270562HX, SEG09270563HX, SEG09270564HX, SEG09270565HX, SEG09270566HX, SEG09270567HX, SEG09270568HX, SEG09270569HX, SEG09270583HX, SEG09270584HX, SEG09270585HX, SEG09270587HX, SEG09270588HX, SEG09270589HX, SEG09270591HX, SEG09270592HX, SEG09280596HX, SEG09280597HX, SEG09280598HX, SEG09280599HX, SEG09280600HX, SEG09280601HX, SEG09280602HX, SEG09280603HX, SEG09280604HX, SEG09280605HX, SEG09280606HX, SEG09280607HX, SEG09280608HX, SEG09280609HX, SEG09280610HX, SEG09280611HX, SEG09280612HX, SEG09280613HX, SEG09280614HX, SEG09280615HX, SEG09280616HX, SEG09280617HX, SEG09280618HX, SEG09280619HX, SEG09280620HX, SEG09280621HX, SEG09280622HX, SEG09280623HX, SEG09280624HX, SEG09280625HX, SEG09310656HX, SEG09310657HX, SEG09310658HX, SEG09310659HX, SEG09310660HX, SEG09310661HX, SEG09310662HX, SEG09310663HX, SEG09310664HX, SEG09310665HX, SEG09310666HX, SEG09310667HX, SEG09310668HX, SEG09310669HX, SEG09310670HX, SEG09310671HX, SEG09310672HX, SEG09310673HX, SEG09310674HX, SEG09310675HX, SEG09310696HX, SEG09310697HX, SEG09310698HX, SEG09310699HX, SEG09310700HX, SEG09310701HX, SEG09310702HX, SEG09310703HX, SEG09310704HX, SEG09310705HX, SEG09310706HX, SEG09310707HX, SEG09310708HX, SEG09310709HX, SEG09310710HX, SEG09310711HX, SEG09310712HX, SEG09310713HX, SEG09310714HX, SEG09310715HX, SEG09310716HX, SEG09310717HX, SEG09310718HX, SEG09310719HX, SEG09310720HX, SEG09310721HX, SEG09310722HX, SEG09310723HX, SEG09310724HX, SEG09310725HX, SEG09310726HX, SEG09310727HX, SEG09310728HX, SEG09310729HX, SEG09310730HX, SEG09310731HX, SEG09310732HX, SEG09310733HX, SEG09310734HX, SEG09310735HX, SEG09310736HX, SEG09310737HX, SEG09310738HX, SEG09310739HX, SEG09310740HX, SEG09310741HX, SEG09310742HX, SEG09310743HX, SEG09310744HX, SEG09310745HX, SEG09320797HX, SEG09320798HX, SEG09320799HX, SEG09320800HX, SEG09320801HX, SEG09320802HX, SEG09320803HX, SEG09320804HX, SEG09320805HX, SEG09320806HX, SEG09320807HX, SEG09320808HX, SEG09320809HX, SEG09320810HX, SEG09320811HX, SEG09320812HX, SEG09320813HX, SEG09320814HX, SEG09320815HX, SEG09320816HX, SEG09320817HX, SEG09320818HX, SEG09320819HX, SEG09320820HX, SEG09320821HX, SEG09320822HX, SEG09320823HX, SEG09320824HX, SEG09320825HX, SEG09320826HX, SEG09320827HX, SEG09320828HX, SEG09320829HX, SEG09320830HX, SEG09320831HX, SEG09320832HX, SEG09320833HX, SEG09320834HX, SEG09320835HX, SEG09320836HX, SEG09320837HX, SEG09320838HX, SEG09320839HX, SEG09320840HX, SEG09320841HX, SEG09320842HX, SEG09320843HX, SEG09320844HX, SEG09320845HX, SEG09320846HX, SEG09320847HX, SEG09320848HX, SEG09320849HX, SEG09320850HX, SEG09320851HX, SEG09320852HX, SEG09320853HX, SEG09320854HX, SEG09320855HX, SEG09320856HX, SEG09320857HX, SEG09320858HX, SEG09320859HX, SEG09320860HX, SEG09320861HX, SEG09320862HX, SEG09320863HX, SEG09320864HX, SEG09320865HX, SEG09320866HX, SEG09320867HX, SEG09320868HX, SEG09320869HX, SEG09320870HX, SEG09320871HX, SEG09320872HX, SEG09320873HX, SEG09320874HX, SEG09320875HX, SEG09320876HX, SEG09320877HX, SEG09320878HX, SEG09320879HX, SEG09320880HX, SEG09320881HX, SEG09320882HX, SEG09320883HX, SEG09320884HX, SEG09320885HX, SEG09320886HX, SEG09320887HX, SEG09320888HX, SEG09320889HX, SEG09320890HX, SEG09320891HX, SEG09320892HX, SEG09320893HX, SEG09320894HX, SEG09320895HX, SEG09320896HX, SEG09360898HX, SEG09360899HX, SEG09360900HX, SEG09360901HX, SEG09360902HX, SEG09360903HX, SEG09360904HX, SEG09360905HX, SEG09360906HX, SEG09360907HX, SEG09360908HX, SEG09360909HX, SEG09360910HX, SEG09360911HX, SEG09360912HX, SEG09360913HX, SEG09360914HX, SEG09360915HX, SEG09360916HX, SEG09360917HX, SEG09360918HX, SEG09360919HX, SEG09360920HX, SEG09360921HX, SEG09360922HX, SEG09360923HX, SEG09360924HX, SEG09360925HX, SEG09360926HX, SEG09360927HX, SEG09360928HX, SEG09360929HX, SEG09360930HX, SEG09360931HX, SEG09360932HX, SEG09360933HX, SEG09360934HX, SEG09360935HX, SEG09360936HX, SEG09360937HX, SEG09360938HX, SEG09360
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDAThe Rotoprone therapy System is an advanced patient care system for the treatment and prevention of pulmonaryFebruary 12, 2014
- FDAXiO RTP System. Used to create treatment plans for any cancer patient for whom external beam radiation therapyFebruary 12, 2014
Questions about this recall
Was CARESCAPE Monitor B850, a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility. Product Usage: The CARESCAPE Monitor B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B850 system is indicated for monitoring of Hemodynamic (including ECG, ST Segment, Arrhythmia Detection, ECG Diagnostic Analysis and Measurement, Invasive Pressure, Noninvasive Blood Pressure, Pulse Oximetry, Cardiac Output, Temperature, Impedance Respiration and SvO2 (Mixed Venous Oxygen Saturation)), Airway Gases (Fi/Et C02, 02, N20 and Anesthetic Agent), Spirometry, Gas Exchange (02 Consumption (VO2), C02 production (VCO2), energy expenditure (BE), and respiratory quotient (RQ)) and neurophysiological (including electroencephalography (EEG), Entropy, Bispectral Index (BIS) and Neuromuscular Transmission (NMT) Monitoring) status. The CARESCAPE Monitor B850 provides alarms, trends, snapshots and events, and calculations and can be connected to displays, printers and recording devices. The CARESCAPE Monitor B850 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for bed to bed viewing and to data management software devices via a network. The CARESCAPE Monitor B850 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility. In addition to the healthcare practitioner, the CARESCAPE Monitor B850 is designed to provide configuration and troubleshooting capabilities to qualified service personnel. recalled?
Yes. FDA published recall Z-0927-2014 on February 12, 2014. Reported hazard: There is a potential for communication loss associated with the F5-01 Frame when connected to CARESCAPE Patient Data Module (PDM) in the CARESCAPE Monitor B850. The PDM patient parameter acquisition module can sometimes spontaneously lose the communication with the host patient monitor when attached to the F5-01 Frame due to an issue with the PDM Ethernet communication hardware. The communicatio
What should I do if I have this product?
Voluntary: Firm initiated · Letter