Cryo Module Accessories Domestic CMA-OUS, Product Catalog Number A001053
Published February 7, 2024 · Recall initiated: November 28, 2023
Summary
FDA records medical device Class II recall Z-0927-2024, published February 7, 2024 and initiated November 28, 2023. Recalling firm: AtriCure, Inc.; Status: Ongoing; Reported quantity: 35; Lot: UDI-DI: 10840143903038; Lot Numbers: 129673, 129674, 129882, 130314, 130750, 131614, 132669, 133672, 133673, 134270. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
AtriCure has identified certain lots of tank hose assemblies where a second-tier supplier provided incorrect tank hose assemblies. These tank hose assemblies were provided with an outer sheathing that does not have required pin perforations and the inner tube of the hose is not the correct material.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- AtriCure, Inc.
- Distributor
- US Nationwide. Brazil, Canada, Japan, Taiwan
- Lot
- UDI-DI: 10840143903038; Lot Numbers: 129673, 129674, 129882, 130314, 130750, 131614, 132669, 133672, 133673, 134270
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPackaged, Tank Hose Assembly CMA-US, Product Catalog Number A001056February 7, 2024
- FDATIBIAL BEARING INSERT SERIES P-S I ASSY Intended for knee replacement Product Number: 7291-0324February 7, 2024
- FDAPS LIPPED TIBIAL INSERT ASSY DURACON Intended for knee replacement Product Number: 6742-1-416February 7, 2024
- FDAVolar Bearing Plate. REF: VBEAL-3-7N, VBEAL-5-7N, VBEAL-3-7S, VBEAL-5-7S, VBEAL-7-7S, VBEAL-3-7W, VBEAL-5-7W,February 7, 2024
- FDAPS LIPPED TIBIAL INSERT ASSY DURACON Intended for knee replacement Product Number: 6742-1-413February 7, 2024
- FDAAtriClip" LAA Exclusion System with preloaded Gillinov-Cosgrove" Clip - A single PRO2 device is packaged in aFebruary 15, 2017
Questions about this recall
Was Cryo Module Accessories Domestic CMA-OUS, Product Catalog Number A001053 recalled?
Yes. FDA published recall Z-0927-2024 on February 7, 2024. Reported hazard: AtriCure has identified certain lots of tank hose assemblies where a second-tier supplier provided incorrect tank hose assemblies. These tank hose assemblies were provided with an outer sheathing that does not have required pin perforations and the inner tube of the hose is not the correct material.
What should I do if I have this product?
Voluntary: Firm initiated · Letter