Back to search
FDAFDA · Food and drugs·Recall number Z-0928-2024

Packaged, Tank Hose Assembly CMA-US, Product Catalog Number A001056

Published February 7, 2024 · Recall initiated: November 28, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

AtriCure has identified certain lots of tank hose assemblies where a second-tier supplier provided incorrect tank hose assemblies. These tank hose assemblies were provided with an outer sheathing that does not have required pin perforations and the inner tube of the hose is not the correct material.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
AtriCure, Inc.
Distributor
US Nationwide. Brazil, Canada, Japan, Taiwan
Lot
UDI-DI: 10840143903076; Lot Number: 129675
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card