Back to search
FDAFDA · Food and drugs·Recall number Z-0928-2024

Packaged, Tank Hose Assembly CMA-US, Product Catalog Number A001056

Published February 7, 2024 · Recall initiated: November 28, 2023

Summary

FDA records medical device Class II recall Z-0928-2024, published February 7, 2024 and initiated November 28, 2023. Recalling firm: AtriCure, Inc.; Status: Ongoing; Reported quantity: 1; Lot: UDI-DI: 10840143903076; Lot Number: 129675. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

AtriCure has identified certain lots of tank hose assemblies where a second-tier supplier provided incorrect tank hose assemblies. These tank hose assemblies were provided with an outer sheathing that does not have required pin perforations and the inner tube of the hose is not the correct material.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
AtriCure, Inc.
Distributor
US Nationwide. Brazil, Canada, Japan, Taiwan
Lot
UDI-DI: 10840143903076; Lot Number: 129675
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Packaged, Tank Hose Assembly CMA-US, Product Catalog Number A001056 recalled?

Yes. FDA published recall Z-0928-2024 on February 7, 2024. Reported hazard: AtriCure has identified certain lots of tank hose assemblies where a second-tier supplier provided incorrect tank hose assemblies. These tank hose assemblies were provided with an outer sheathing that does not have required pin perforations and the inner tube of the hose is not the correct material.

What should I do if I have this product?

Voluntary: Firm initiated · Letter