Diagnostic imaging systems (GEMINI LXL Model #882412, TruFlight Select Model #882438, GEMINI TF PET/CT16 Model #882470, GEMINI TF PET/CT 64 Model #882471, GEMINI TF Ready Model #882473 & GEMINI TF Big Bore Model #882476)
Published February 12, 2014 · Recall initiated: December 4, 2013
Summary
FDA records medical device Class II recall Z-0932-2014, published February 12, 2014 and initiated December 4, 2013. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 386 units; Lot: Model #882412; Serial #'s:14701, 14702, 14704, 14705, 14706, 14707, 14708, 14709, 14710 & 14711. Model #882438; Serial #'s:17002, 17003, 17004, 17005, 17006, 17007, 17008, 17009, 17010, 17011, 17012, 17013, 17014, 17015, 17017, 17018, 17019, 17021, 17022, 17023, 17024, 17025, 17026 & 17028. Model #882470; Serial #'s: 7003, 7004, 7005, 7006, 7008, 7009, 7010, 7011, 7014, 7015, 7017, 7019, 7020, 7021, 7022, 7024, 7025, 7026, 7031, 7034, 7035, 7036, 7039, 7043, 7045, 7049, 7050, 7052, 7053, 7054, 7058, 7060, 7062, 7064, 7066, 7068, 7069, 7074, 7075, 7077, 7080, 7081, 7084, 7085, 7086, 7087, 7089, 7093, 7094, 7095, 7096, 7097, 7098, 7099, 7108, 7112, 7114, 7115, 7118, 7120, 7121, 7122, 7123, 7124, 7125, 7128, 7129, 7131, 7135, 7137, 7138, 7139, 7140, 7142, 7143, 7145, 7147, 7148, 7156, 7159, 7161, 7163, 7166, 7168, 7173, 7174, 7176, 7177, 7181, 7183, 7184, 7185, 7186, 7187, 7193, 7194, 7195, 7196, 7197, 7198, 7199, 7200, 7207, 7208, 7210, 7211, 7215, 7216, 7223, 7224, 7226, 7228, 7230, 7233, 7235, 7239, 7240, 7242, 7501, 7503, 7507, 7508, 7511, 7514, 7517, 7518, 7520, 7521, 7523, 7527, 7528, 7529, 7530, 7531, 7532, 7533, 7536, 7538, 7540, 7545, 7546, 7554, 7559, 7564, 7566, 7568, 7570, 7573, 7581, 7583, 7584, 7586, 7590, 7591, 7104M, 7106M, 7170M &7582M. Model #882471; Serial #'s: 7007, 7012, 7013, 7016, 7027, 7028, 7029, 7030, 7032, 7037, 7042, 7044, 7048, 7055, 7057, 7059, 7061, 7063, 7065, 7071, 7072, 7076, 7078, 7079, 7082, 7083, 7088, 7091, 7092, 7102, 7103, 7105, 7116, 7117, 7119, 7126, 7132, 7136, 7141, 7144, 7146, 7149, 7152, 7153, 7154, 7158, 7160, 7164, 7167, 7169, 7171, 7172, 7175, 7178, 7179, 7180, 7182, 7188, 7189, 7190, 7192, 7201, 7202, 7203, 7204, 7205, 7206, 7209, 7212, 7213, 7214, 7217, 7218, 7219, 7221, 7227, 7229, 7231, 7234, 7236, 7238, 7241, 7245, 7251, 7252, 7253, 7254, 7255, 7256, 7257, 7258, 7504, 7506, 7509, 7510, 7512, 7515, 7516, 7519, 7524, 7525, 7535, 7537, 7539, 7542, 7543, 7544, 7548, 7549, 7550, 7552, 7553, 7555, 7556, 7557, 7558, 7560, 7561, 7562, 7563, 7567, 7569, 7572, 7574, 7575, 7577, 7578, 7579, 7585, 7587, 7589 &7592. Model #882473; Serial #'s: 7040, 7048, 7070, 7095, 7111, 7113, 7127, 7130, 7134, 7150, 7151, 7155, 7157, 7162, 7237 & 7249. Model #:882476; Serial #'s: 9004, 9005, 9006, 9008, 9009, 9010, 9011, 9012, 9013, 9014, 9015, 9016, 9017, 9018, 9019, 9020, 9021, 9022, 9023, 9201, 9202, 9203, 9204, 9205, 9206, 9207, 9208, 9209, 9210, 9211, 9213, 9214, 9215, 9216, 9217, 9218, 9219, 9220, 9221, 9222, 9223, 9224, 9225, 9226, 9228, 9229 & 9231. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The PET Reconstruction Server (PRS) database may randomly lock up after the low dose CT acquisition is complete and will not allow the PET acquisition to begin, resulting in an unexpected incomplete study.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Distributor
- Product was shipped to the following states: AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, NY , OH, OR, PA, TN, TX, VA, VT, WA & WI. Product was also shipped to the following countries: Australia, Belgium, Brazil, Bulgaria, Canada, China, Croatia, Cuba, Denmark, Egypt, France, Germany, Hong Kong, Hungary, India, Indonesia, Italy, Japan, Kuwait, Lebanon, Malta, Mexico, Netherlands, Pakistan, Philippines, Poland, Portugal, Republic of Korea, Russia, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, Ukraine, United Kingdom, Venezuela & Vietnam.
- Lot
- Model #882412; Serial #'s:14701, 14702, 14704, 14705, 14706, 14707, 14708, 14709, 14710 & 14711. Model #882438; Serial #'s:17002, 17003, 17004, 17005, 17006, 17007, 17008, 17009, 17010, 17011, 17012, 17013, 17014, 17015, 17017, 17018, 17019, 17021, 17022, 17023, 17024, 17025, 17026 & 17028. Model #882470; Serial #'s: 7003, 7004, 7005, 7006, 7008, 7009, 7010, 7011, 7014, 7015, 7017, 7019, 7020, 7021, 7022, 7024, 7025, 7026, 7031, 7034, 7035, 7036, 7039, 7043, 7045, 7049, 7050, 7052, 7053, 7054, 7058, 7060, 7062, 7064, 7066, 7068, 7069, 7074, 7075, 7077, 7080, 7081, 7084, 7085, 7086, 7087, 7089, 7093, 7094, 7095, 7096, 7097, 7098, 7099, 7108, 7112, 7114, 7115, 7118, 7120, 7121, 7122, 7123, 7124, 7125, 7128, 7129, 7131, 7135, 7137, 7138, 7139, 7140, 7142, 7143, 7145, 7147, 7148, 7156, 7159, 7161, 7163, 7166, 7168, 7173, 7174, 7176, 7177, 7181, 7183, 7184, 7185, 7186, 7187, 7193, 7194, 7195, 7196, 7197, 7198, 7199, 7200, 7207, 7208, 7210, 7211, 7215, 7216, 7223, 7224, 7226, 7228, 7230, 7233, 7235, 7239, 7240, 7242, 7501, 7503, 7507, 7508, 7511, 7514, 7517, 7518, 7520, 7521, 7523, 7527, 7528, 7529, 7530, 7531, 7532, 7533, 7536, 7538, 7540, 7545, 7546, 7554, 7559, 7564, 7566, 7568, 7570, 7573, 7581, 7583, 7584, 7586, 7590, 7591, 7104M, 7106M, 7170M &7582M. Model #882471; Serial #'s: 7007, 7012, 7013, 7016, 7027, 7028, 7029, 7030, 7032, 7037, 7042, 7044, 7048, 7055, 7057, 7059, 7061, 7063, 7065, 7071, 7072, 7076, 7078, 7079, 7082, 7083, 7088, 7091, 7092, 7102, 7103, 7105, 7116, 7117, 7119, 7126, 7132, 7136, 7141, 7144, 7146, 7149, 7152, 7153, 7154, 7158, 7160, 7164, 7167, 7169, 7171, 7172, 7175, 7178, 7179, 7180, 7182, 7188, 7189, 7190, 7192, 7201, 7202, 7203, 7204, 7205, 7206, 7209, 7212, 7213, 7214, 7217, 7218, 7219, 7221, 7227, 7229, 7231, 7234, 7236, 7238, 7241, 7245, 7251, 7252, 7253, 7254, 7255, 7256, 7257, 7258, 7504, 7506, 7509, 7510, 7512, 7515, 7516, 7519, 7524, 7525, 7535, 7537, 7539, 7542, 7543, 7544, 7548, 7549, 7550, 7552, 7553, 7555, 7556, 7557, 7558, 7560, 7561, 7562, 7563, 7567, 7569, 7572, 7574, 7575, 7577, 7578, 7579, 7585, 7587, 7589 &7592. Model #882473; Serial #'s: 7040, 7048, 7070, 7095, 7111, 7113, 7127, 7130, 7134, 7150, 7151, 7155, 7157, 7162, 7237 & 7249. Model #:882476; Serial #'s: 9004, 9005, 9006, 9008, 9009, 9010, 9011, 9012, 9013, 9014, 9015, 9016, 9017, 9018, 9019, 9020, 9021, 9022, 9023, 9201, 9202, 9203, 9204, 9205, 9206, 9207, 9208, 9209, 9210, 9211, 9213, 9214, 9215, 9216, 9217, 9218, 9219, 9220, 9221, 9222, 9223, 9224, 9225, 9226, 9228, 9229 & 9231.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.September 21, 2022
- FDABrilliance 64 Model # 728231, computed tomography x-ray systemAugust 14, 2019
- FDAIngenuity Core128 Model # 728323, computed tomography x-ray systemAugust 14, 2019
- FDABrightView XCT , Model Number 882482July 31, 2019
- FDATrifurcated Extension Set, model no. ME 1224. Trifurcated Extension Set, model no. MP 9230.February 12, 2014
- FDAThe Gaymar Medi-Therm device provides a means of regulating patient temperature by supplyingFebruary 12, 2014
Questions about this recall
Was Diagnostic imaging systems (GEMINI LXL Model #882412, TruFlight Select Model #882438, GEMINI TF PET/CT16 Model #882470, GEMINI TF PET/CT 64 Model #882471, GEMINI TF Ready Model #882473 & GEMINI TF Big Bore Model #882476) recalled?
Yes. FDA published recall Z-0932-2014 on February 12, 2014. Reported hazard: The PET Reconstruction Server (PRS) database may randomly lock up after the low dose CT acquisition is complete and will not allow the PET acquisition to begin, resulting in an unexpected incomplete study.
What should I do if I have this product?
Voluntary: Firm initiated · Letter