Siemens Total Iron Binding Capacity (IBCT) Flex Reagent Cartridge. An in vitro diagnostic test intended to quantitatively measure Total Iron Binding Capacity in human serum.
Published March 20, 2013 · Recall initiated: December 20, 2012
Summary
FDA records medical device Class III recall Z-0933-2013, published March 20, 2013 and initiated December 20, 2012. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Terminated; Reported quantity: 19, 503 cartons; Lot: Siemens Material Number (SMN) 10444944, Catalog number DF84 -- lot numbers BA3016, EXP 2013-01-16; EC3045, EXP 2013-02-14; EA3073, EXP 2013-03-14; DC3102, EXP 2013-04-12; FC3122, EXP 2013-05-02; EA3157, EXP 2013-06-06; BB3184, EXP 2013-07-03; EC3219, EXP 2013-08-07; EC3220, EXP 2013-08-08; GA3248, 2013-09-05; and FB3290, EXP 2013-10-17. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The analytical sensitivity claim of 6 ug/dL (1.1 umol/L) in the Instructions for Use (IFU) is no longer valid. Siemens is conducting a voluntary corrective action for Dimension Total Iron Binding Capacity (IBCT) to modify the analytical sensitivity claim to 36 ug/dL (6.44 umol/L).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- Nationwide Distribution
- Lot
- Siemens Material Number (SMN) 10444944, Catalog number DF84 -- lot numbers BA3016, EXP 2013-01-16; EC3045, EXP 2013-02-14; EA3073, EXP 2013-03-14; DC3102, EXP 2013-04-12; FC3122, EXP 2013-05-02; EA3157, EXP 2013-06-06; BB3184, EXP 2013-07-03; EC3219, EXP 2013-08-07; EC3220, EXP 2013-08-08; GA3248, 2013-09-05; and FB3290, EXP 2013-10-17
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAtellica CH Revised C-Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223;October 7, 2026
- FDABeckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mLAugust 26, 2026
- FDASyva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mLAugust 26, 2026
- FDASyva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219ULAugust 12, 2026
- FDACynch Lordotic T-PLlF Implant. Intervertebral fusion device with bone graft, for use with supplementalMarch 20, 2013
- FDAGE Healthcare, Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ. Cardiovascular X-ray imaging systems. For useMarch 20, 2013
Questions about this recall
Was Siemens Total Iron Binding Capacity (IBCT) Flex Reagent Cartridge. An in vitro diagnostic test intended to quantitatively measure Total Iron Binding Capacity in human serum. recalled?
Yes. FDA published recall Z-0933-2013 on March 20, 2013. Reported hazard: The analytical sensitivity claim of 6 ug/dL (1.1 umol/L) in the Instructions for Use (IFU) is no longer valid. Siemens is conducting a voluntary corrective action for Dimension Total Iron Binding Capacity (IBCT) to modify the analytical sensitivity claim to 36 ug/dL (6.44 umol/L).
What should I do if I have this product?
Voluntary: Firm initiated · Letter