GE Healthcare, Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ. Cardiovascular X-ray imaging systems. For use in generating fluoroscopic images of human anatomy.

Published March 20, 2013 · Recall initiated: February 4, 2011

Summary

FDA records medical device Class II recall Z-0934-2013, published March 20, 2013 and initiated February 4, 2011. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 48; Lot: SERIAL NUMBER 592043BU4 592676BU1 593459BU1 593465BU8 592458BU4 592673BU8 592682BU9 593059BU9 593051BU6 593475BU7 593476BU5 592049BU1 593468BU2 593054BU0 592047BU5 592042BU6 592046BU7 592459BU2 593048BU2 593056BU5 593049BU0 593063BU1 593062BU3 593464BU1 594762BU7 593473BU2 592451BU9 593460BU9 597954BU7 592456BU8 592674BU6 593052BU4 593058BU1 590867BU8 592050BU9 593064BU9 592460BU0 592457BU6 592680BU3 592455BU0 593462BU5 592678BU7 594763BU5 593055BU7 592675BU3 592452BU7 593463BU3 594385BU7. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Exposed high voltage wires inside the cabinet . On a cable located in the system cabinet, two of the shrink tubings covering the 400V live cable leave an area potentially unprotected.

What to do

Voluntary: Firm initiated · Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
GE Healthcare, LLC
Distributor
Worldwide Distribution-USA including the states of FL, IL, IN, NJ, NC, OK, TX, and WI, and the countries of UNITED KINGDOM, TUNISIA, SERBIA, RUSSIAN FEDERATION, NORWAY, NIGERIA, LITHUANIA, LEBANON, KOREA, REPUBLIC OF KOREA, ITALY, INDONESIA, INDIA, FRANCE, FINLAND, COLOMBIA, CHINA, and BRAZIL.
Lot
SERIAL NUMBER 592043BU4 592676BU1 593459BU1 593465BU8 592458BU4 592673BU8 592682BU9 593059BU9 593051BU6 593475BU7 593476BU5 592049BU1 593468BU2 593054BU0 592047BU5 592042BU6 592046BU7 592459BU2 593048BU2 593056BU5 593049BU0 593063BU1 593062BU3 593464BU1 594762BU7 593473BU2 592451BU9 593460BU9 597954BU7 592456BU8 592674BU6 593052BU4 593058BU1 590867BU8 592050BU9 593064BU9 592460BU0 592457BU6 592680BU3 592455BU0 593462BU5 592678BU7 594763BU5 593055BU7 592675BU3 592452BU7 593463BU3 594385BU7
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was GE Healthcare, Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ. Cardiovascular X-ray imaging systems. For use in generating fluoroscopic images of human anatomy. recalled?

Yes. FDA published recall Z-0934-2013 on March 20, 2013. Reported hazard: Exposed high voltage wires inside the cabinet . On a cable located in the system cabinet, two of the shrink tubings covering the 400V live cable leave an area potentially unprotected.

What should I do if I have this product?

Voluntary: Firm initiated · Visit