LOCATOR® Implant Attachment System. The LOCATOR® Implant Attachment System is a universal hinge, resilient attachment for endosseous implants. The LOCATOR® Implant Attachment System is designed for use with overdentures or partial dentures, retained in whole or in part, by endosseous implants in the mandible or maxilla. FDA 510(k): K994257, K072878
Published February 12, 2014 · Recall initiated: January 16, 2014
Summary
FDA records medical device Class II recall Z-0935-2014, published February 12, 2014 and initiated January 16, 2014. Recalling firm: Zest Anchors LLC; Status: Terminated; Reported quantity: 3,130; Lot: Part Numbers: 8875-SPI, 8875-SPI-TMS, 8876-SPI, 8876-SPI-TMS, 08877-SPI, 08877-SPI-TMS, 08878-SPI, 08878-SPI-TMS, 08879-SPI, 08880-SPI; Lot Numbers: 11052, 15306, 20665, 21351, 12025,12259, 22440, I0K7X, I0K7Y, 11963,15307,17676,18934, 25294, 25567, I0JQ7, 12258,12941,15000, 19735, 20666, I0K7Z, I0KY7, 12703, 14568, 19832, 22170, I0LCG, 11093, 13364, 14151, 15624, 20043, 25026, 13530, I0LE4, 10055, 14152, 15001, 25027, 10892, 10893. Since the LOCATOR® Implant Attachment System (including the restorative components) is not sold by the Zest Anchors as a sterilized product, it does not have a designated shelf-life. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Zest Anchors, LLC is voluntarily recalling specific lots of LOCATOR® Implant Attachment Systems because it is currently provided with a label that indicates that the final torque applied to the abutment should be 30 N-cm. It was recently noted by Zest that in order to provide an optimal performance by increasing the factor of safety between the failure load of the abutment screw and the recommend
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zest Anchors LLC
- Distributor
- Distributed in Ohio and Switzerland.
- Lot
- Part Numbers: 8875-SPI, 8875-SPI-TMS, 8876-SPI, 8876-SPI-TMS, 08877-SPI, 08877-SPI-TMS, 08878-SPI, 08878-SPI-TMS, 08879-SPI, 08880-SPI; Lot Numbers: 11052, 15306, 20665, 21351, 12025,12259, 22440, I0K7X, I0K7Y, 11963,15307,17676,18934, 25294, 25567, I0JQ7, 12258,12941,15000, 19735, 20666, I0K7Z, I0KY7, 12703, 14568, 19832, 22170, I0LCG, 11093, 13364, 14151, 15624, 20043, 25026, 13530, I0LE4, 10055, 14152, 15001, 25027, 10892, 10893. Since the LOCATOR® Implant Attachment System (including the restorative components) is not sold by the Zest Anchors as a sterilized product, it does not have a designated shelf-life.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOxcarbazepine Tablets, 300 mg, 100 tablet bottles, Rx Only. Distributed by: Caraco PharmaceuticalFebruary 19, 2014
- FDA3.5mm Locking Cortical Bone Screw, T15 Hexalobe, Self Tapping, 20mm Part No. 00-0903-2620 OrthoPediatricsFebruary 12, 2014
- FDAThe Symbionic Leg is an integrated prosthetic leg for transfemoral and knee disarticulation amputees. TheFebruary 12, 2014
- FDAThe Gaymar Medi-Therm device provides a means of regulating patient temperature by supplyingFebruary 12, 2014
- FDAHemochron Jr. Citrate APTT Cuvette 510(k) K014008 A unitized microcoagulation test intended to be used inFebruary 12, 2014
- FDADiagnostic imaging systems (GEMINI LXL Model #882412, TruFlight Select Model #882438, GEMINI TF PET/CT16 ModelFebruary 12, 2014
Questions about this recall
Was LOCATOR® Implant Attachment System. The LOCATOR® Implant Attachment System is a universal hinge, resilient attachment for endosseous implants. The LOCATOR® Implant Attachment System is designed for use with overdentures or partial dentures, retained in whole or in part, by endosseous implants in the mandible or maxilla. FDA 510(k): K994257, K072878 recalled?
Yes. FDA published recall Z-0935-2014 on February 12, 2014. Reported hazard: Zest Anchors, LLC is voluntarily recalling specific lots of LOCATOR® Implant Attachment Systems because it is currently provided with a label that indicates that the final torque applied to the abutment should be 30 N-cm. It was recently noted by Zest that in order to provide an optimal performance by increasing the factor of safety between the failure load of the abutment screw and the recommend
What should I do if I have this product?
Voluntary: Firm initiated · Letter