LOCATOR¿ Implant Attachment System. The LOCATOR¿ Implant Attachment System is a universal hinge, resilient attachment for endosseous implants. The LOCATOR¿ Implant Attachment System is designed for use with overdentures or partial dentures, retained in whole or in part, by endosseous implants in the mandible or maxilla. FDA 510(k): K994257, K072878

Published February 12, 2014 · Recall initiated: January 16, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Zest Anchors, LLC is voluntarily recalling specific lots of LOCATOR¿ Implant Attachment Systems because it is currently provided with a label that indicates that the final torque applied to the abutment should be 30 N-cm. It was recently noted by Zest that in order to provide an optimal performance by increasing the factor of safety between the failure load of the abutment screw and the recommend

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Zest Anchors LLC
Distributor
Distributed in Ohio and Switzerland.
Lot
Part Numbers: 8875-SPI, 8875-SPI-TMS, 8876-SPI, 8876-SPI-TMS, 08877-SPI, 08877-SPI-TMS, 08878-SPI, 08878-SPI-TMS, 08879-SPI, 08880-SPI; Lot Numbers: 11052, 15306, 20665, 21351, 12025,12259, 22440, I0K7X, I0K7Y, 11963,15307,17676,18934, 25294, 25567, I0JQ7, 12258,12941,15000, 19735, 20666, I0K7Z, I0KY7, 12703, 14568, 19832, 22170, I0LCG, 11093, 13364, 14151, 15624, 20043, 25026, 13530, I0LE4, 10055, 14152, 15001, 25027, 10892, 10893. Since the LOCATOR¿ Implant Attachment System (including the restorative components) is not sold by the Zest Anchors as a sterilized product, it does not have a designated shelf-life.