Trilogy EVO Ventilator, Iberia - Model Number: ES2110X15B Trilogy Evo is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and Pulse Rate data when integrated with the appropriate accessories.

Published February 26, 2020 · Recall initiated: November 19, 2019

Summary

FDA records medical device Class I recall Z-0943-2020, published February 26, 2020 and initiated November 19, 2019. Recalling firm: Philips Respironics, Inc.; Status: Terminated; Reported quantity: 8; Lot: Model/Material Number: ES2110X15B Serial Numbers: H25123324AA45, H2512383941DE, H251239768495, H25124050192C, H251250171648, H25125051141E, H2512510467D7. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing requirements TSRS1298 and PRD439.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Respironics, Inc.
Distributor
US: None OUS: CH DE DK IT FR IE GB KW QA PL ES AT PH AE SI CZ
Lot
Model/Material Number: ES2110X15B Serial Numbers: H25123324AA45, H2512383941DE, H251239768495, H25124050192C, H251250171648, H25125051141E, H2512510467D7
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Trilogy EVO Ventilator, Iberia - Model Number: ES2110X15B Trilogy Evo is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and Pulse Rate data when integrated with the appropriate accessories. recalled?

Yes. FDA published recall Z-0943-2020 on February 26, 2020. Reported hazard: Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing requirements TSRS1298 and PRD439.

What should I do if I have this product?

Voluntary: Firm initiated · Letter