Brilliance 6, 10, 16, 40, 64, Big Bore and the iCT, iCT SP, Ingenuity CT and MX8000 Dual v. EXP are Computed Tomography X-Ray Systems. These devices are Computed Tomography X-Ray Systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
Published March 20, 2013 · Recall initiated: December 20, 2012
Summary
FDA records medical device Class II recall Z-0945-2013, published March 20, 2013 and initiated December 20, 2012. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 4797 Units; Lot: System Code #728256, Serial #: 3001,3016,3019,3020,3025,3026,3030,3057,3058,3060,3071,3073,3074,3076,3080,3090,3095,3108,3114,3119,3124,3138,3151,3155,3167,3168,3174,3178,3186,3187,3192,3193,3196,3208,3210,3212,3213,3214,3218,3226,3245,3257,3264,3267,3269,3273,3274,3275,3281,3284,3289,3291,3296,3304,3306,3309,3313,3317,3327,3331,3336,3337,3339,3340,3341,3342,3344,3345,3346,3347,3348,3349,3350,3351,3352,3354,3355,3356,3357,3358,3359,3360,3361,3362,3363,3364,3365,3366,3367,3368,3369,3370,3371,3372,3373,3374,3375,3376,3377,3378,3379,3380,3381,3382,3383,3384,3385,3386,3387,3388,3389,3390,3391,3392,3393,3394,3395,3396,3397,3398,3399,3400,3402,3403,3404,3405,3406,3408,3409,3410,3411,3412,3413,3414,3415,3416,3417,3418,3419,3420,3421,3422,3423,3424,3425,3426,3427,3428,3429,3430,3431,3432,3433,3434,3435,3436,3437,3438,3439,3440,3441,3443,3444,3445,3447,3449,3450,3451,3452,3453,3454,3455,3456,3457,3458,3459,3460,3461,3462,3463,3464,3465,3466,3467,3468,3469,3470,3471,3472,3473,3474,3475,3476,3477,3478,3479,3480,3481,3482,3483,3484,3485,3486,3487,3488,3489,3490,3491,3492,3493,3494,3495,3496,3497,3498,3499,3500,3501,3502,3503,3504,3505,3506,3507,3508,3509,3510,3511,3512,3513,3514,3515,3516,3517,3518,3519,3520,3521,3522,3523,3524,3525,3526,3527,3528,3529,3530,3531,3532,3533,3534,3535,3536,3537,3538,3539,3540,3541,3542,3543,3544,3545,3546,3547,3548,3549,3550,3551,3552,3553,3554,3555,3556,3557,3558,3559,3560,3561,3562,3563,3564,3565,3566,3567,3568,3569,3570,3571,3572,3573,3574,3575,3576,3577,3578,3579,3580,3581,3582,3583,3584,3585,3586,3587,3588,3589,3590,3591,3592,3593,3594,3595,3596,3597,3598,3599,3600,3601,3602,3603,3604,3605,3606,3607,3608,3609,3610,3611,3612,3613,3614,3615,3616,3617,3618,3619,3620,3621,3622,3623,3624,3625,3626,3627,3628,3629,3630,3631,3632,3633,3634,3635,3636,3637,3638,3639,3640,3641,3642,3643,3644,3645,3646,3647,3648,3649,3650,3651,3652,3653,3654,3655,3656,3657,3658,3659,3660,3661,3663,3664,3665,3666,3667,3668,3669,3670,3671,3672,3673,3674,3675,3676,3677,3678,3679,3680,3681,3682,3683,3684,3685,3686,3687,3688,3689,3690,3691,3692,3693,3694,3695,3696,3697,3698,3699,3700,3701,3702,3703,3704,3705,3706,3707,3708,3709,3710,3711,3712,3713,3714,3715,3716,3717,3718,3719,3720,3721,3722,6039,6040,30003,30004,30006,30007,30008,30009,30012,30013,30015,30016,30021,30026,30028,30029,30030,30031,30036,30037,30038,30039,30046,30047,30050,30053,30054,30057,30060,30062,30065,30066,30072,30073,30075,30078,30079,30082,30083,30089,30090,30094,30100,30101,30103,30105,30109,30110,30111,30117,30118,30120,30122,30124,30125,30126,30128,30129,30130,30131 System Code #728251, Serial #: 3008,3098,4006,4007,4010,4011,4012,4016,4023,4031,4039,4050,4058,4061,4062,4063,4064,4065,4066,4067,40004,40005,40006,40007,40008,40009,40010,40011,40012,40014,40015,40016,40017,40018,40021,40025,40028,40030,40032 System Code #728246, Serial #: 3012,3013,3015,3017,3023,3027,3031,3043,3070,3072,3160,3185,3189,3191,3224,3295,3448,4018,4019,4024,4025,4043,4051,5001,5005,5006,5007,5013,5014,5016,5019,5022,5023,5024,5025,5028,5029,5032,5034,5036,5038,5046,5051,5052,5053,5055,5058,5059,5061,5063,5064,5069,5070,5074,5075,5077,5083,5086,5097,5102,5105,5106,5107,5108,5110,5111,5112,5113,5115,5116,5117,5119,5122,5123,5124,5125,5127,5130,5136,5137,5138,5139,5141,5143,5150,5151,5156,5158,5161,5169,5174,5179,5183,5184,5187,5188,5190,5195,5198,5202,5207,5211,5215,5217,5218,5220,5221,5224,5226,5228,5229,5230,5234,5249,5250,5252,5254,5265,5270,5274,5278,5282,5283,5285,5286,5287,5290,5291,5295,5296,5299,5309,5313,5315,5316,5317,5318,5319,5323,5328,5329,5331,5334,5343,5364,5365,5366,5367,5370,5371,5372,5375,5376,5377,5378,5379,5380,5381,5382,5383,5384,5385,5386,5387,5388,5389,5391,5392,5393,5394,5396,5397,5398,5399,5400,5401,5402,5403,5404,5405,5406,5407,5408,5409,5410,5411,5412,5414,5415,5416,5417,5418,5419,5420,5421,5422,5423,5424,5425,5426,5427,5428,5429,5432,5433,5434,5435,5437,5438,5439,5441,5442,5443,5444,5446,5447,5448,5449,5450,5451,5453,5457,5458,5459,5460,5461,5462,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Philips Healthcare received reports from the field stating the patient table on a system had an unexpected downward movement while a patient was on the table. Philips has determined the cause was a malfunction of the vertical brake.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Distributor
- Worldwide Distribution - Nationwide Distribution including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY.
- Lot
- System Code #728256, Serial #: 3001,3016,3019,3020,3025,3026,3030,3057,3058,3060,3071,3073,3074,3076,3080,3090,3095,3108,3114,3119,3124,3138,3151,3155,3167,3168,3174,3178,3186,3187,3192,3193,3196,3208,3210,3212,3213,3214,3218,3226,3245,3257,3264,3267,3269,3273,3274,3275,3281,3284,3289,3291,3296,3304,3306,3309,3313,3317,3327,3331,3336,3337,3339,3340,3341,3342,3344,3345,3346,3347,3348,3349,3350,3351,3352,3354,3355,3356,3357,3358,3359,3360,3361,3362,3363,3364,3365,3366,3367,3368,3369,3370,3371,3372,3373,3374,3375,3376,3377,3378,3379,3380,3381,3382,3383,3384,3385,3386,3387,3388,3389,3390,3391,3392,3393,3394,3395,3396,3397,3398,3399,3400,3402,3403,3404,3405,3406,3408,3409,3410,3411,3412,3413,3414,3415,3416,3417,3418,3419,3420,3421,3422,3423,3424,3425,3426,3427,3428,3429,3430,3431,3432,3433,3434,3435,3436,3437,3438,3439,3440,3441,3443,3444,3445,3447,3449,3450,3451,3452,3453,3454,3455,3456,3457,3458,3459,3460,3461,3462,3463,3464,3465,3466,3467,3468,3469,3470,3471,3472,3473,3474,3475,3476,3477,3478,3479,3480,3481,3482,3483,3484,3485,3486,3487,3488,3489,3490,3491,3492,3493,3494,3495,3496,3497,3498,3499,3500,3501,3502,3503,3504,3505,3506,3507,3508,3509,3510,3511,3512,3513,3514,3515,3516,3517,3518,3519,3520,3521,3522,3523,3524,3525,3526,3527,3528,3529,3530,3531,3532,3533,3534,3535,3536,3537,3538,3539,3540,3541,3542,3543,3544,3545,3546,3547,3548,3549,3550,3551,3552,3553,3554,3555,3556,3557,3558,3559,3560,3561,3562,3563,3564,3565,3566,3567,3568,3569,3570,3571,3572,3573,3574,3575,3576,3577,3578,3579,3580,3581,3582,3583,3584,3585,3586,3587,3588,3589,3590,3591,3592,3593,3594,3595,3596,3597,3598,3599,3600,3601,3602,3603,3604,3605,3606,3607,3608,3609,3610,3611,3612,3613,3614,3615,3616,3617,3618,3619,3620,3621,3622,3623,3624,3625,3626,3627,3628,3629,3630,3631,3632,3633,3634,3635,3636,3637,3638,3639,3640,3641,3642,3643,3644,3645,3646,3647,3648,3649,3650,3651,3652,3653,3654,3655,3656,3657,3658,3659,3660,3661,3663,3664,3665,3666,3667,3668,3669,3670,3671,3672,3673,3674,3675,3676,3677,3678,3679,3680,3681,3682,3683,3684,3685,3686,3687,3688,3689,3690,3691,3692,3693,3694,3695,3696,3697,3698,3699,3700,3701,3702,3703,3704,3705,3706,3707,3708,3709,3710,3711,3712,3713,3714,3715,3716,3717,3718,3719,3720,3721,3722,6039,6040,30003,30004,30006,30007,30008,30009,30012,30013,30015,30016,30021,30026,30028,30029,30030,30031,30036,30037,30038,30039,30046,30047,30050,30053,30054,30057,30060,30062,30065,30066,30072,30073,30075,30078,30079,30082,30083,30089,30090,30094,30100,30101,30103,30105,30109,30110,30111,30117,30118,30120,30122,30124,30125,30126,30128,30129,30130,30131 System Code #728251, Serial #: 3008,3098,4006,4007,4010,4011,4012,4016,4023,4031,4039,4050,4058,4061,4062,4063,4064,4065,4066,4067,40004,40005,40006,40007,40008,40009,40010,40011,40012,40014,40015,40016,40017,40018,40021,40025,40028,40030,40032 System Code #728246, Serial #: 3012,3013,3015,3017,3023,3027,3031,3043,3070,3072,3160,3185,3189,3191,3224,3295,3448,4018,4019,4024,4025,4043,4051,5001,5005,5006,5007,5013,5014,5016,5019,5022,5023,5024,5025,5028,5029,5032,5034,5036,5038,5046,5051,5052,5053,5055,5058,5059,5061,5063,5064,5069,5070,5074,5075,5077,5083,5086,5097,5102,5105,5106,5107,5108,5110,5111,5112,5113,5115,5116,5117,5119,5122,5123,5124,5125,5127,5130,5136,5137,5138,5139,5141,5143,5150,5151,5156,5158,5161,5169,5174,5179,5183,5184,5187,5188,5190,5195,5198,5202,5207,5211,5215,5217,5218,5220,5221,5224,5226,5228,5229,5230,5234,5249,5250,5252,5254,5265,5270,5274,5278,5282,5283,5285,5286,5287,5290,5291,5295,5296,5299,5309,5313,5315,5316,5317,5318,5319,5323,5328,5329,5331,5334,5343,5364,5365,5366,5367,5370,5371,5372,5375,5376,5377,5378,5379,5380,5381,5382,5383,5384,5385,5386,5387,5388,5389,5391,5392,5393,5394,5396,5397,5398,5399,5400,5401,5402,5403,5404,5405,5406,5407,5408,5409,5410,5411,5412,5414,5415,5416,5417,5418,5419,5420,5421,5422,5423,5424,5425,5426,5427,5428,5429,5432,5433,5434,5435,5437,5438,5439,5441,5442,5443,5444,5446,5447,5448,5449,5450,5451,5453,5457,5458,5459,5460,5461,5462,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.September 21, 2022
- FDABrilliance 64 Model # 728231, computed tomography x-ray systemAugust 14, 2019
- FDAIngenuity Core128 Model # 728323, computed tomography x-ray systemAugust 14, 2019
- FDABrightView XCT , Model Number 882482July 31, 2019
- FDAINTERNAL HEX RECON SCREW, TI-6AL-4V, REF 7162375, TRIGEN (R) INTERNAL HEX CAPTURED RECON SCREW, QTY: (1), 6.4March 20, 2013
- FDANicolet EEG Wireless Amplifier 32/64 Channel. Catalog/Part Numbers : 515-015400 for the 32 channel amplifierMarch 20, 2013
Questions about this recall
Was Brilliance 6, 10, 16, 40, 64, Big Bore and the iCT, iCT SP, Ingenuity CT and MX8000 Dual v. EXP are Computed Tomography X-Ray Systems. These devices are Computed Tomography X-Ray Systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories. recalled?
Yes. FDA published recall Z-0945-2013 on March 20, 2013. Reported hazard: Philips Healthcare received reports from the field stating the patient table on a system had an unexpected downward movement while a patient was on the table. Philips has determined the cause was a malfunction of the vertical brake.
What should I do if I have this product?
Voluntary: Firm initiated · Letter