Brilliance 6, 10, 16, 40, 64, Big Bore and the iCT, iCT SP, Ingenuity CT and MX8000 Dual v. EXP are Computed Tomography X-Ray Systems. These devices are Computed Tomography X-Ray Systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.

Published March 20, 2013 · Recall initiated: December 20, 2012

Summary

FDA records medical device Class II recall Z-0945-2013, published March 20, 2013 and initiated December 20, 2012. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 4797 Units; Lot: System Code #728256, Serial #: 3001,3016,3019,3020,3025,3026,3030,3057,3058,3060,3071,3073,3074,3076,3080,3090,3095,3108,3114,3119,3124,3138,3151,3155,3167,3168,3174,3178,3186,3187,3192,3193,3196,3208,3210,3212,3213,3214,3218,3226,3245,3257,3264,3267,3269,3273,3274,3275,3281,3284,3289,3291,3296,3304,3306,3309,3313,3317,3327,3331,3336,3337,3339,3340,3341,3342,3344,3345,3346,3347,3348,3349,3350,3351,3352,3354,3355,3356,3357,3358,3359,3360,3361,3362,3363,3364,3365,3366,3367,3368,3369,3370,3371,3372,3373,3374,3375,3376,3377,3378,3379,3380,3381,3382,3383,3384,3385,3386,3387,3388,3389,3390,3391,3392,3393,3394,3395,3396,3397,3398,3399,3400,3402,3403,3404,3405,3406,3408,3409,3410,3411,3412,3413,3414,3415,3416,3417,3418,3419,3420,3421,3422,3423,3424,3425,3426,3427,3428,3429,3430,3431,3432,3433,3434,3435,3436,3437,3438,3439,3440,3441,3443,3444,3445,3447,3449,3450,3451,3452,3453,3454,3455,3456,3457,3458,3459,3460,3461,3462,3463,3464,3465,3466,3467,3468,3469,3470,3471,3472,3473,3474,3475,3476,3477,3478,3479,3480,3481,3482,3483,3484,3485,3486,3487,3488,3489,3490,3491,3492,3493,3494,3495,3496,3497,3498,3499,3500,3501,3502,3503,3504,3505,3506,3507,3508,3509,3510,3511,3512,3513,3514,3515,3516,3517,3518,3519,3520,3521,3522,3523,3524,3525,3526,3527,3528,3529,3530,3531,3532,3533,3534,3535,3536,3537,3538,3539,3540,3541,3542,3543,3544,3545,3546,3547,3548,3549,3550,3551,3552,3553,3554,3555,3556,3557,3558,3559,3560,3561,3562,3563,3564,3565,3566,3567,3568,3569,3570,3571,3572,3573,3574,3575,3576,3577,3578,3579,3580,3581,3582,3583,3584,3585,3586,3587,3588,3589,3590,3591,3592,3593,3594,3595,3596,3597,3598,3599,3600,3601,3602,3603,3604,3605,3606,3607,3608,3609,3610,3611,3612,3613,3614,3615,3616,3617,3618,3619,3620,3621,3622,3623,3624,3625,3626,3627,3628,3629,3630,3631,3632,3633,3634,3635,3636,3637,3638,3639,3640,3641,3642,3643,3644,3645,3646,3647,3648,3649,3650,3651,3652,3653,3654,3655,3656,3657,3658,3659,3660,3661,3663,3664,3665,3666,3667,3668,3669,3670,3671,3672,3673,3674,3675,3676,3677,3678,3679,3680,3681,3682,3683,3684,3685,3686,3687,3688,3689,3690,3691,3692,3693,3694,3695,3696,3697,3698,3699,3700,3701,3702,3703,3704,3705,3706,3707,3708,3709,3710,3711,3712,3713,3714,3715,3716,3717,3718,3719,3720,3721,3722,6039,6040,30003,30004,30006,30007,30008,30009,30012,30013,30015,30016,30021,30026,30028,30029,30030,30031,30036,30037,30038,30039,30046,30047,30050,30053,30054,30057,30060,30062,30065,30066,30072,30073,30075,30078,30079,30082,30083,30089,30090,30094,30100,30101,30103,30105,30109,30110,30111,30117,30118,30120,30122,30124,30125,30126,30128,30129,30130,30131 System Code #728251, Serial #: 3008,3098,4006,4007,4010,4011,4012,4016,4023,4031,4039,4050,4058,4061,4062,4063,4064,4065,4066,4067,40004,40005,40006,40007,40008,40009,40010,40011,40012,40014,40015,40016,40017,40018,40021,40025,40028,40030,40032 System Code #728246, Serial #: 3012,3013,3015,3017,3023,3027,3031,3043,3070,3072,3160,3185,3189,3191,3224,3295,3448,4018,4019,4024,4025,4043,4051,5001,5005,5006,5007,5013,5014,5016,5019,5022,5023,5024,5025,5028,5029,5032,5034,5036,5038,5046,5051,5052,5053,5055,5058,5059,5061,5063,5064,5069,5070,5074,5075,5077,5083,5086,5097,5102,5105,5106,5107,5108,5110,5111,5112,5113,5115,5116,5117,5119,5122,5123,5124,5125,5127,5130,5136,5137,5138,5139,5141,5143,5150,5151,5156,5158,5161,5169,5174,5179,5183,5184,5187,5188,5190,5195,5198,5202,5207,5211,5215,5217,5218,5220,5221,5224,5226,5228,5229,5230,5234,5249,5250,5252,5254,5265,5270,5274,5278,5282,5283,5285,5286,5287,5290,5291,5295,5296,5299,5309,5313,5315,5316,5317,5318,5319,5323,5328,5329,5331,5334,5343,5364,5365,5366,5367,5370,5371,5372,5375,5376,5377,5378,5379,5380,5381,5382,5383,5384,5385,5386,5387,5388,5389,5391,5392,5393,5394,5396,5397,5398,5399,5400,5401,5402,5403,5404,5405,5406,5407,5408,5409,5410,5411,5412,5414,5415,5416,5417,5418,5419,5420,5421,5422,5423,5424,5425,5426,5427,5428,5429,5432,5433,5434,5435,5437,5438,5439,5441,5442,5443,5444,5446,5447,5448,5449,5450,5451,5453,5457,5458,5459,5460,5461,5462,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Philips Healthcare received reports from the field stating the patient table on a system had an unexpected downward movement while a patient was on the table. Philips has determined the cause was a malfunction of the vertical brake.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Medical Systems (Cleveland) Inc
Distributor
Worldwide Distribution - Nationwide Distribution including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY.
Lot
System Code #728256, Serial #: 3001,3016,3019,3020,3025,3026,3030,3057,3058,3060,3071,3073,3074,3076,3080,3090,3095,3108,3114,3119,3124,3138,3151,3155,3167,3168,3174,3178,3186,3187,3192,3193,3196,3208,3210,3212,3213,3214,3218,3226,3245,3257,3264,3267,3269,3273,3274,3275,3281,3284,3289,3291,3296,3304,3306,3309,3313,3317,3327,3331,3336,3337,3339,3340,3341,3342,3344,3345,3346,3347,3348,3349,3350,3351,3352,3354,3355,3356,3357,3358,3359,3360,3361,3362,3363,3364,3365,3366,3367,3368,3369,3370,3371,3372,3373,3374,3375,3376,3377,3378,3379,3380,3381,3382,3383,3384,3385,3386,3387,3388,3389,3390,3391,3392,3393,3394,3395,3396,3397,3398,3399,3400,3402,3403,3404,3405,3406,3408,3409,3410,3411,3412,3413,3414,3415,3416,3417,3418,3419,3420,3421,3422,3423,3424,3425,3426,3427,3428,3429,3430,3431,3432,3433,3434,3435,3436,3437,3438,3439,3440,3441,3443,3444,3445,3447,3449,3450,3451,3452,3453,3454,3455,3456,3457,3458,3459,3460,3461,3462,3463,3464,3465,3466,3467,3468,3469,3470,3471,3472,3473,3474,3475,3476,3477,3478,3479,3480,3481,3482,3483,3484,3485,3486,3487,3488,3489,3490,3491,3492,3493,3494,3495,3496,3497,3498,3499,3500,3501,3502,3503,3504,3505,3506,3507,3508,3509,3510,3511,3512,3513,3514,3515,3516,3517,3518,3519,3520,3521,3522,3523,3524,3525,3526,3527,3528,3529,3530,3531,3532,3533,3534,3535,3536,3537,3538,3539,3540,3541,3542,3543,3544,3545,3546,3547,3548,3549,3550,3551,3552,3553,3554,3555,3556,3557,3558,3559,3560,3561,3562,3563,3564,3565,3566,3567,3568,3569,3570,3571,3572,3573,3574,3575,3576,3577,3578,3579,3580,3581,3582,3583,3584,3585,3586,3587,3588,3589,3590,3591,3592,3593,3594,3595,3596,3597,3598,3599,3600,3601,3602,3603,3604,3605,3606,3607,3608,3609,3610,3611,3612,3613,3614,3615,3616,3617,3618,3619,3620,3621,3622,3623,3624,3625,3626,3627,3628,3629,3630,3631,3632,3633,3634,3635,3636,3637,3638,3639,3640,3641,3642,3643,3644,3645,3646,3647,3648,3649,3650,3651,3652,3653,3654,3655,3656,3657,3658,3659,3660,3661,3663,3664,3665,3666,3667,3668,3669,3670,3671,3672,3673,3674,3675,3676,3677,3678,3679,3680,3681,3682,3683,3684,3685,3686,3687,3688,3689,3690,3691,3692,3693,3694,3695,3696,3697,3698,3699,3700,3701,3702,3703,3704,3705,3706,3707,3708,3709,3710,3711,3712,3713,3714,3715,3716,3717,3718,3719,3720,3721,3722,6039,6040,30003,30004,30006,30007,30008,30009,30012,30013,30015,30016,30021,30026,30028,30029,30030,30031,30036,30037,30038,30039,30046,30047,30050,30053,30054,30057,30060,30062,30065,30066,30072,30073,30075,30078,30079,30082,30083,30089,30090,30094,30100,30101,30103,30105,30109,30110,30111,30117,30118,30120,30122,30124,30125,30126,30128,30129,30130,30131 System Code #728251, Serial #: 3008,3098,4006,4007,4010,4011,4012,4016,4023,4031,4039,4050,4058,4061,4062,4063,4064,4065,4066,4067,40004,40005,40006,40007,40008,40009,40010,40011,40012,40014,40015,40016,40017,40018,40021,40025,40028,40030,40032 System Code #728246, Serial #: 3012,3013,3015,3017,3023,3027,3031,3043,3070,3072,3160,3185,3189,3191,3224,3295,3448,4018,4019,4024,4025,4043,4051,5001,5005,5006,5007,5013,5014,5016,5019,5022,5023,5024,5025,5028,5029,5032,5034,5036,5038,5046,5051,5052,5053,5055,5058,5059,5061,5063,5064,5069,5070,5074,5075,5077,5083,5086,5097,5102,5105,5106,5107,5108,5110,5111,5112,5113,5115,5116,5117,5119,5122,5123,5124,5125,5127,5130,5136,5137,5138,5139,5141,5143,5150,5151,5156,5158,5161,5169,5174,5179,5183,5184,5187,5188,5190,5195,5198,5202,5207,5211,5215,5217,5218,5220,5221,5224,5226,5228,5229,5230,5234,5249,5250,5252,5254,5265,5270,5274,5278,5282,5283,5285,5286,5287,5290,5291,5295,5296,5299,5309,5313,5315,5316,5317,5318,5319,5323,5328,5329,5331,5334,5343,5364,5365,5366,5367,5370,5371,5372,5375,5376,5377,5378,5379,5380,5381,5382,5383,5384,5385,5386,5387,5388,5389,5391,5392,5393,5394,5396,5397,5398,5399,5400,5401,5402,5403,5404,5405,5406,5407,5408,5409,5410,5411,5412,5414,5415,5416,5417,5418,5419,5420,5421,5422,5423,5424,5425,5426,5427,5428,5429,5432,5433,5434,5435,5437,5438,5439,5441,5442,5443,5444,5446,5447,5448,5449,5450,5451,5453,5457,5458,5459,5460,5461,5462,
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Brilliance 6, 10, 16, 40, 64, Big Bore and the iCT, iCT SP, Ingenuity CT and MX8000 Dual v. EXP are Computed Tomography X-Ray Systems. These devices are Computed Tomography X-Ray Systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories. recalled?

Yes. FDA published recall Z-0945-2013 on March 20, 2013. Reported hazard: Philips Healthcare received reports from the field stating the patient table on a system had an unexpected downward movement while a patient was on the table. Philips has determined the cause was a malfunction of the vertical brake.

What should I do if I have this product?

Voluntary: Firm initiated · Letter