iPlan RT is a radiation treatment planning system that is intended for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck, and extracranial lesions.
Published January 21, 2015 · Recall initiated: November 19, 2014
Summary
FDA records medical device Class II recall Z-0957-2015, published January 21, 2015 and initiated November 19, 2014. Recalling firm: Brainlab AG; Status: Terminated; Reported quantity: 1412 systems total; Lot: Software Versions: 1) 21213 IPLAN RT DOSE 3.0 PLATFORM; 2) 21213A IPLAN RT DOSE 3.0.1 PLATFORM; 3) 21213B IPLAN RT DOSE 3.0.2 PLATFORM; 4) 21213D IPLAN RT DOSE 4.1.0 PLATFORM; 5) 21213E IPLAN RT DOSE 4.1.1 PLATFORM; 6) 21213F IPLAN RT DOSE 4.1.2 PLATFORM; 7) 21213G IPLAN RT DOSE 4.1.3 PLATFORM; 8) 21213H IPLAN RT DOSE 4.1.4 PLATFORM; 9) 21381 IPLAN RT DOSE 4.5.0 PLATFORM; 10) 21381A IPLAN RT DOSE 4.5.1 PLATFORM; 11) 21381B IPLAN RT DOSE 4.5.2 PLATFORM; 12) 21381C IPLAN RT DOSE 4.5.3 PLATFORM; 13) 21385 UPGRADE IPLAN RT DOSE 4.X TO 4.5.0; 14) 21385A UPGRADE IPLAN RT DOSE 4.X TO 4.5.1; 15) 21385B UPGRADE IPLAN RT DOSE 4.X TO 4.5.2; 16) 21385C UPGRADE IPLAN RT DOSE 4.X TO 4.5.3; 17) 21387 iPLAN RT DOSE PLATFORM 4.1 (21213); 18) 21387A iPLAN RT DOSE PLATFORM 4.1.2 (21213); 19) 21387B iPLAN RT DOSE PLATFORM 4.5.1 (21381); 20) 21387C iPLAN RT DOSE PLATFORM 4.5.2 (21381); 21) 21387D IPLAN RT DOSE PLATFORM 4.5.3(21381) V; 22) 70213 UPGRADE BRAINSCAN 5.X TO iPLAN RT DOSE 3; 23) 70213A UPGRADE BRAINSCAN 5.X TO iPLAN RT DOSE 3; 24) 70213B UPG BRAINSCAN 5.X TO iPLAN RT DOSE 3.0; 25) 70213C UPG BRAINSCAN 5.X TO iPLAN RT DOSE4.1.0; 26) 70213D UPG BRAINSCAN 5.X TO iPLAN RT DOSE 4.1.1; 27) 70213E UPG BRAINSCAN 5.X TO iPLAN RT DOSE 4.1.2; 28) 70213F UPG BRAINSCAN 5.X TO iPLAN RT DOSE 4.5.1; 29) 70214 UPGRADE BRAINSCAN 3.X/ 4.X TO IPLAN RT D; 30) 70214A UPGRADE BRAINSCAN 3.X/ 4.X TO IPLAN RT D; 31) 70214B UPG BRAINSC 3.X/4.X TO IPLAN RT DOSE 3.0; 32) 70214C UPG BRAINSC. 3/4.X - iPLAN RT DOSE 4.1.0; 33) 70214D UPG BRAINSC. 3/4.X - iPLAN RT DOSE 4.1.1; 34) 70214E UPG BRAINSC. 3/4.X - iPLAN RT DOSE 4.1.2; 35) 70214F UPG BRAINSC. 3/4.X - iPLAN RT DOSE 4.5.1; 36) 70226 UPGRADE IPLAN RT DOSE 3.0 TO 3.0.1; 37) 70226A UPGRADE IPLAN RT DOSE TO 3.0.2; 38) 70228 UPGRADE IPLAN RT DOSE 3.X TO 4.0; 39) 70228A UPG IPLAN RT DOSE 3.X TO 4.1.0; 40) 70228B UPG IPLAN RT DOSE 3.X TO 4.1.1; 41) 70228C UPG IPLAN RT DOSE 3.X TO 4.1.2; 42) 70228D UPG IPLAN RT DOSE 3.X TO 4.5.1; 43) 70228E UPG IPLAN RT DOSE 3.X TO 4.5.2; 44) 70228F UPG IPLAN RT DOSE 3.X TO 4.5.3; 45) 70235 UPGRADE BRAINSCAN TO IPLAN RT DOSE 4.5; 46) 70235A UPGRADE BRAINSCAN TO IPLAN RT DOSE 4.5.2; 47) 70235B UPGRADE BRAINSCAN TO IPLAN RT DOSE 4.5.3; 48) 70237 UPD IPLAN RT DOSE 4.x TO 4.5 (0TRANSFER); 49) 70237A UPD IPLAN RT DOSE 4.x TO 4.5.2 (O TRANS); 50) 70237B UPD IPLAN RT DOSE 4.x TO 4.5.3(O TRANS); 51) 70263 UPD IPLAN RT DOSE 4.1.x to 4.1.3.(0TRA); 52) 70440 UPD iPLAN RT DOSE 4.1.0 TO 4.1.1; 53) 70440A UPD iPLAN RT DOSE 4.1 TO 4.1.2; 54) 20146 2ND LICENCE SW - FOR ONE ADDITIONAL WS; 55) 20147 RT SRS 2ND LICENCE SW FOR ONE ADD SITE; 56) 20148 NOVALIS TX 2ND LICENCE SW ONE ADD SITE. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
iPlan RT Radiation Treatment Planning Software: Potentially incorrect patient positioning when using multiple localized CT image data sets.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Brainlab AG
- Distributor
- Worldwide Distribution - USA including Alabama, Arizona, Arkansas, California, Colorado, Connecticut, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, Wisconsin, Wyoming and Internationally to Algeria, Angola, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Latvia, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, Norway, Oman, Panama, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Kingdom, Uruguay, and Venezuela.
- Lot
- Software Versions: 1) 21213 IPLAN RT DOSE 3.0 PLATFORM; 2) 21213A IPLAN RT DOSE 3.0.1 PLATFORM; 3) 21213B IPLAN RT DOSE 3.0.2 PLATFORM; 4) 21213D IPLAN RT DOSE 4.1.0 PLATFORM; 5) 21213E IPLAN RT DOSE 4.1.1 PLATFORM; 6) 21213F IPLAN RT DOSE 4.1.2 PLATFORM; 7) 21213G IPLAN RT DOSE 4.1.3 PLATFORM; 8) 21213H IPLAN RT DOSE 4.1.4 PLATFORM; 9) 21381 IPLAN RT DOSE 4.5.0 PLATFORM; 10) 21381A IPLAN RT DOSE 4.5.1 PLATFORM; 11) 21381B IPLAN RT DOSE 4.5.2 PLATFORM; 12) 21381C IPLAN RT DOSE 4.5.3 PLATFORM; 13) 21385 UPGRADE IPLAN RT DOSE 4.X TO 4.5.0; 14) 21385A UPGRADE IPLAN RT DOSE 4.X TO 4.5.1; 15) 21385B UPGRADE IPLAN RT DOSE 4.X TO 4.5.2; 16) 21385C UPGRADE IPLAN RT DOSE 4.X TO 4.5.3; 17) 21387 iPLAN RT DOSE PLATFORM 4.1 (21213); 18) 21387A iPLAN RT DOSE PLATFORM 4.1.2 (21213); 19) 21387B iPLAN RT DOSE PLATFORM 4.5.1 (21381); 20) 21387C iPLAN RT DOSE PLATFORM 4.5.2 (21381); 21) 21387D IPLAN RT DOSE PLATFORM 4.5.3(21381) V; 22) 70213 UPGRADE BRAINSCAN 5.X TO iPLAN RT DOSE 3; 23) 70213A UPGRADE BRAINSCAN 5.X TO iPLAN RT DOSE 3; 24) 70213B UPG BRAINSCAN 5.X TO iPLAN RT DOSE 3.0; 25) 70213C UPG BRAINSCAN 5.X TO iPLAN RT DOSE4.1.0; 26) 70213D UPG BRAINSCAN 5.X TO iPLAN RT DOSE 4.1.1; 27) 70213E UPG BRAINSCAN 5.X TO iPLAN RT DOSE 4.1.2; 28) 70213F UPG BRAINSCAN 5.X TO iPLAN RT DOSE 4.5.1; 29) 70214 UPGRADE BRAINSCAN 3.X/ 4.X TO IPLAN RT D; 30) 70214A UPGRADE BRAINSCAN 3.X/ 4.X TO IPLAN RT D; 31) 70214B UPG BRAINSC 3.X/4.X TO IPLAN RT DOSE 3.0; 32) 70214C UPG BRAINSC. 3/4.X - iPLAN RT DOSE 4.1.0; 33) 70214D UPG BRAINSC. 3/4.X - iPLAN RT DOSE 4.1.1; 34) 70214E UPG BRAINSC. 3/4.X - iPLAN RT DOSE 4.1.2; 35) 70214F UPG BRAINSC. 3/4.X - iPLAN RT DOSE 4.5.1; 36) 70226 UPGRADE IPLAN RT DOSE 3.0 TO 3.0.1; 37) 70226A UPGRADE IPLAN RT DOSE TO 3.0.2; 38) 70228 UPGRADE IPLAN RT DOSE 3.X TO 4.0; 39) 70228A UPG IPLAN RT DOSE 3.X TO 4.1.0; 40) 70228B UPG IPLAN RT DOSE 3.X TO 4.1.1; 41) 70228C UPG IPLAN RT DOSE 3.X TO 4.1.2; 42) 70228D UPG IPLAN RT DOSE 3.X TO 4.5.1; 43) 70228E UPG IPLAN RT DOSE 3.X TO 4.5.2; 44) 70228F UPG IPLAN RT DOSE 3.X TO 4.5.3; 45) 70235 UPGRADE BRAINSCAN TO IPLAN RT DOSE 4.5; 46) 70235A UPGRADE BRAINSCAN TO IPLAN RT DOSE 4.5.2; 47) 70235B UPGRADE BRAINSCAN TO IPLAN RT DOSE 4.5.3; 48) 70237 UPD IPLAN RT DOSE 4.x TO 4.5 (0TRANSFER); 49) 70237A UPD IPLAN RT DOSE 4.x TO 4.5.2 (O TRANS); 50) 70237B UPD IPLAN RT DOSE 4.x TO 4.5.3(O TRANS); 51) 70263 UPD IPLAN RT DOSE 4.1.x to 4.1.3.(0TRA); 52) 70440 UPD iPLAN RT DOSE 4.1.0 TO 4.1.1; 53) 70440A UPD iPLAN RT DOSE 4.1 TO 4.1.2; 54) 20146 2ND LICENCE SW - FOR ONE ADDITIONAL WS; 55) 20147 RT SRS 2ND LICENCE SW FOR ONE ADD SITE; 56) 20148 NOVALIS TX 2ND LICENCE SW ONE ADD SITE
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Germany
Related recalls
- FDAUltrasound Navigation Software 1.0 and Ultrasound Integration Software for Cranial/ENT Navigation 3.0.x andOctober 14, 2020
- FDASpine & Trauma 3D Navigation 1.0 (subpart of the system Navigation Software Spine & Trauma 3D, Version 3.0 )April 24, 2019
- FDAExacTrac versions 6.5 through 6.5 intended to be used to place patients at an accurately defined point withinMarch 21, 2018
- FDAAIRO Mobile CT System, Model Number: MobiCT-32 Product Usage: The AIRO is intended to be used for X-rayFebruary 14, 2018
- FDAGE Healthcare, Proteus XR/a with wall stand model number 600-0301.January 21, 2015
- FDAHeartStart MRx Monitor/Defibrillator; Model. Numbers: M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2,January 21, 2015
Questions about this recall
Was iPlan RT is a radiation treatment planning system that is intended for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck, and extracranial lesions. recalled?
Yes. FDA published recall Z-0957-2015 on January 21, 2015. Reported hazard: iPlan RT Radiation Treatment Planning Software: Potentially incorrect patient positioning when using multiple localized CT image data sets.
What should I do if I have this product?
Voluntary: Firm initiated · Letter