Back to search
FDAFDA · Food and drugs·Recall number Z-0962-2026

BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyxis Pro 7-Drawer Auxiliary, REF: 1149-00, BD Pyxis Pro 7- Drawer Auxiliary with Secure Tower, REF: 1152-00, BD Pyxis Pro MedStation Main, REF: 1155-00, BD Pyxis CII Safe ES, REF: 1116-00

Published December 31, 2025 · Recall initiated: November 20, 2025

Summary

FDA records medical device Class II recall Z-0962-2026, published December 31, 2025 and initiated November 20, 2025. Recalling firm: CareFusion 303, Inc.; Status: Ongoing; Reported quantity: 53669; Lot: Firmware Version: 1.48. REF/UDI-DI: 323/10885403512667, 324/10885403512681, 352/10885403512674, 1149-00/10885403517822, 1152-00/10885403517839, 1155-00/10885403521300, 1116-00/10885403512605. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
CareFusion 303, Inc.
Distributor
Worldwide - US Nationwide distribution including in the states of CO, IN, OH, TX, KS, FL, GA, NC, CA, HI, IL, NY, MI, VA, AL, LA, NE, KY, OK, AR, TN, WY, NJ, WI, SD, AZ, NV, MO, MA, ID, WV, DE, PA, WA, MT, CT, IA, MD, PR, MN, NM, OR, VT, MS, ME, AK, UT, DC, GU, SC, NH, RI, ND, AS and the countries of CH, CA, KW, QA, NL, GB, UY, AU, NZ, BH, TH, SG, CN, FR, LB, BS, HK, TR, PL, AR, BR, CL, CO, MX, ES, BM, SA, IL.
Lot
Firmware Version: 1.48. REF/UDI-DI: 323/10885403512667, 324/10885403512681, 352/10885403512674, 1149-00/10885403517822, 1152-00/10885403517839, 1155-00/10885403521300, 1116-00/10885403512605
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyxis Pro 7-Drawer Auxiliary, REF: 1149-00, BD Pyxis Pro 7- Drawer Auxiliary with Secure Tower, REF: 1152-00, BD Pyxis Pro MedStation Main, REF: 1155-00, BD Pyxis CII Safe ES, REF: 1116-00 recalled?

Yes. FDA published recall Z-0962-2026 on December 31, 2025. Reported hazard: Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs.

What should I do if I have this product?

Voluntary: Firm initiated · Letter