da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40
Published December 31, 2025 · Recall initiated: August 21, 2024
Summary
FDA records medical device Class II recall Z-0964-2026, published December 31, 2025 and initiated August 21, 2024. Recalling firm: Intuitive Surgical, Inc.; Status: Ongoing; Reported quantity: 144; Lot: UDI-DI: 00886874119747. Serial Numbers: 10476314, 10485693, 10496049, 10497992, 10503432, 10503433, 10507302, 10509551, 10511044, 10511047, 10516467, 10518960, 10518961, 10522438, 10522439, 10522441, 10523771, 10523773, 10523775, 10523776, 10523777, 10523778, 10523779, 10523780, 10523781, 10529917, 10529918, 10529919, 10529920, 10529922, 10534426, 10534428, 10534429, 10534430, 10534752, 10534753, 10534754, 10534755, 10534757, 10543027, 10543028, 10543048, 10543051, 10544670, 10544676, 10544677, 10544678, 10547713, 10547717, 10547724, 10547728, 10551553, 10551554, 10552616, 10552618, 10552620, 10552621, 10556681, 10556682, 10558195, 10558197, 10558198, 10558202, 10560603, 10560604, 10560607, 10562584, 10562585, 10562586, 10562588, 10565609, 10565610, 10565614, 10567141, 10567143, 10567145, 10569365, 10569367, 10569368, 10569369, 10570704, 10570705, 10570706, 10570707, 10573677, 10573678, 10573679, 10573680, 10573681, 10573682, 10573683, 10573684, 10573685, 10573686, 10576019, 10576021, 10576022, 10576023, 10576024, 10576025, 10576026, 10576027, 10576028, 10579442, 10579446, 10579453, 10579455, 10579456, 10579457, 10579461, 10579466, 10582427, 10582428, 10582429, 10591177, 10591179, 10591180, 10591184, 10591197, 10594614, 10594615, 10594616, 10594617, 10594618, 10594619, 10597107, 10597108, 10597109, 10597110, 10599042, 10599044, 10599045, 10599048, 10600750, 10600753, 10603334, 10603335, 10603343, 10603344, 10603345, 10603367, 10605468, 10606129, 10606133. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. Contact Technical Support to report the issue: Cancel-Disable" ergonomics locked until motor replaced; may lead to delay/aborted/alternate procedure
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Intuitive Surgical, Inc.
- Distributor
- US Nationwide distribution in the states of MN, FL, CO, CA, NC, KS, TX, SC, WV, LA, OH, NY, NH, GA, KY, OK, NJ, MA, OR, PA, NM, AZ, TN, ND, IL, AR, WA, IN, VA, CT.
- Lot
- UDI-DI: 00886874119747. Serial Numbers: 10476314, 10485693, 10496049, 10497992, 10503432, 10503433, 10507302, 10509551, 10511044, 10511047, 10516467, 10518960, 10518961, 10522438, 10522439, 10522441, 10523771, 10523773, 10523775, 10523776, 10523777, 10523778, 10523779, 10523780, 10523781, 10529917, 10529918, 10529919, 10529920, 10529922, 10534426, 10534428, 10534429, 10534430, 10534752, 10534753, 10534754, 10534755, 10534757, 10543027, 10543028, 10543048, 10543051, 10544670, 10544676, 10544677, 10544678, 10547713, 10547717, 10547724, 10547728, 10551553, 10551554, 10552616, 10552618, 10552620, 10552621, 10556681, 10556682, 10558195, 10558197, 10558198, 10558202, 10560603, 10560604, 10560607, 10562584, 10562585, 10562586, 10562588, 10565609, 10565610, 10565614, 10567141, 10567143, 10567145, 10569365, 10569367, 10569368, 10569369, 10570704, 10570705, 10570706, 10570707, 10573677, 10573678, 10573679, 10573680, 10573681, 10573682, 10573683, 10573684, 10573685, 10573686, 10576019, 10576021, 10576022, 10576023, 10576024, 10576025, 10576026, 10576027, 10576028, 10579442, 10579446, 10579453, 10579455, 10579456, 10579457, 10579461, 10579466, 10582427, 10582428, 10582429, 10591177, 10591179, 10591180, 10591184, 10591197, 10594614, 10594615, 10594616, 10594617, 10594618, 10594619, 10597107, 10597108, 10597109, 10597110, 10599042, 10599044, 10599045, 10599048, 10600750, 10600753, 10603334, 10603335, 10603343, 10603344, 10603345, 10603367, 10605468, 10606129, 10606133
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAda Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15July 22, 2026
- FDADa Vinci ASSY, PSS, SP1098, Part Number: 380601July 1, 2026
- FDABrand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/AJune 10, 2026
- FDAIntuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06May 13, 2026
- FDABD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ESDecember 31, 2025
- FDALifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02December 31, 2025
Questions about this recall
Was da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40 recalled?
Yes. FDA published recall Z-0964-2026 on December 31, 2025. Reported hazard: Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. Contact Technical Support to report the issue: Cancel-Disable" ergonomics locked until motor replaced; may lead to delay/aborted/alternate procedure
What should I do if I have this product?
Voluntary: Firm initiated · Letter