VITROS 3600 Immunodiagnostic System, Catalog Number 6802783, IVD --- Ortho Clinical Diagnostics. For use in the in vitro quantitative, semi-quantitative and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents.
Published January 21, 2015 · Recall initiated: November 19, 2014
Summary
FDA records medical device Class II recall Z-0967-2015, published January 21, 2015 and initiated November 19, 2014. Recalling firm: Ortho-Clinical Diagnostics; Status: Terminated; Reported quantity: 658 total: USA - 144, Foreign - 514; Lot: Software Version 3.1 & Below; Serial Numbers J36000101 - J36000834 (the serial numbers are sequential; J Numbers are analogous to serial numbers). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ortho-Clinical Diagnostics
- Distributor
- Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
- Lot
- Software Version 3.1 & Below; Serial Numbers J36000101 - J36000834 (the serial numbers are sequential; J Numbers are analogous to serial numbers)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVITROS Performance Verifier IIApril 3, 2024
- FDAVITROS Performance Verifier IApril 3, 2024
- FDAVITROS XT Chemistry Products TBIL-ALKP Slides (ALKP is not affected) In Vitro Diagnostic use only - The TBILSeptember 2, 2020
- FDAVITROS Chemistry Product Performance Verifier I, Product Code 8231474, UDI # 10758750004577 - Product Usage:June 24, 2020
- FDAMUM-1 (Multiple myeloma oncogene-1), catalog number PRM352 AA; Product Usage: For In Vitro Diagnostic Use.January 21, 2015
- FDAlnline Persuader- PedFuse Pedicle Screw System; Part Number: 11-80164January 21, 2015
Questions about this recall
Was VITROS 3600 Immunodiagnostic System, Catalog Number 6802783, IVD --- Ortho Clinical Diagnostics. For use in the in vitro quantitative, semi-quantitative and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents. recalled?
Yes. FDA published recall Z-0967-2015 on January 21, 2015. Reported hazard: Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit