VITROS 3600 Immunodiagnostic System, Catalog Number 6802783, IVD --- Ortho Clinical Diagnostics. For use in the in vitro quantitative, semi-quantitative and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents.

Published January 21, 2015 · Recall initiated: November 19, 2014

Summary

FDA records medical device Class II recall Z-0967-2015, published January 21, 2015 and initiated November 19, 2014. Recalling firm: Ortho-Clinical Diagnostics; Status: Terminated; Reported quantity: 658 total: USA - 144, Foreign - 514; Lot: Software Version 3.1 & Below; Serial Numbers J36000101 - J36000834 (the serial numbers are sequential; J Numbers are analogous to serial numbers). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ortho-Clinical Diagnostics
Distributor
Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
Lot
Software Version 3.1 & Below; Serial Numbers J36000101 - J36000834 (the serial numbers are sequential; J Numbers are analogous to serial numbers)
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was VITROS 3600 Immunodiagnostic System, Catalog Number 6802783, IVD --- Ortho Clinical Diagnostics. For use in the in vitro quantitative, semi-quantitative and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents. recalled?

Yes. FDA published recall Z-0967-2015 on January 21, 2015. Reported hazard: Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit