133" 60 Drop 24 Hour Primary Set w/2 Bravo24®, 2 Gang 1o2® Manifold, Rotating Luer, 1 Ext, Item No. B59006 Product Usage: The intravascular administration set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein). Each set will be manufactured to a unique specifications using dimensions, components and configurations specified by the customer.

Published January 18, 2017 · Recall initiated: November 30, 2016

Summary

FDA records medical device Class II recall Z-0980-2017, published January 18, 2017 and initiated November 30, 2016. Recalling firm: ICU Medical, Inc.; Status: Terminated; Reported quantity: 225 units; Lot: Lot No. 3323650. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

ICU Medical Inc. has identified an isolated event involving a 15 drop drip chamber assembled on a device which should contain a 60 drop drip chamber.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
ICU Medical, Inc.
Distributor
US: NC
Lot
Lot No. 3323650
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was 133" 60 Drop 24 Hour Primary Set w/2 Bravo24®, 2 Gang 1o2® Manifold, Rotating Luer, 1 Ext, Item No. B59006 Product Usage: The intravascular administration set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein). Each set will be manufactured to a unique specifications using dimensions, components and configurations specified by the customer. recalled?

Yes. FDA published recall Z-0980-2017 on January 18, 2017. Reported hazard: ICU Medical Inc. has identified an isolated event involving a 15 drop drip chamber assembled on a device which should contain a 60 drop drip chamber.

What should I do if I have this product?

Voluntary: Firm initiated · Letter