Back to search
FDAFDA · Food and drugs·Recall number Z-0982-2024

DxI 9000 Access Immunoassay Analyzer running software version 1.16 or prior,

Published February 14, 2024 · Recall initiated: November 21, 2023

Summary

FDA records medical device Class II recall Z-0982-2024, published February 14, 2024 and initiated November 21, 2023. Recalling firm: Beckman Coulter, Inc.; Status: Ongoing; Reported quantity: 128 analyzers; Lot: UDI/DI 15099590732103, Serial Numbers: 300116, 300117, 300118, 300123, 300127, 300126, 300128, 300129, 300130, 300131, 300133, 300134, 300135, 300136, 300138, 300139, 300137, 300140, 300142, 300141, 300144, 300143, 300145, 300148 300146, 300147, 300149, 300150, 300151, 300152, 300155, 300153, 300154, 300156, 300158, 300160, 300163, 300161, 300162, 300164, 300167, 300170, 300169, 300171, 300168, 300166, 300172, 300176, 300173, 300178, 300181, 300179, 300183, 300182, 300188, 300189, 300190, 300191, 300192, 300195, 300196, 300193, 300194, 300197, 300199, 300198, 300201, 300200, 300202, 300203, 300205, 300204 300207, 300206, 300208, 300209, 300212, 300210, 300211, 300213, 300214, 300215, 300217, 300216, 300220, 300218, 300219, 300221, 300223, 300224, 300222, 300227, 300228, 300229, 300231, 300230, 300235, 300236, 300234, 300237, 300232, 300239, 300233, 300238, 300240, 300243, 300241, 300242, 300247, 300246, 300249, 300253, 300244, 300250, 300252, 300248, 300251, 300256, 300257, 300254, 300258, 300255, 300259, 300260, 300261, 300264, 300263, 300262; running software version 1.16 or prior. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There are potential performance issues found in the DxI 9000 Access Immunoassay Analyzer, including: 1) Access Ultrasensitive Insulin users who configure SI units (pmol/L) applies an incorrect conversion factor; 2) System provides numerical results that are below the lowest reportable result if laboratory chooses to report assay results in units of measurement other than the defined default units; 3) A result of 0 incorrectly reported when the system utilizes an expression that utilizes non-numerical symbols (e.g., > or <) to derive a calculated test as part of result reporting; 4) Reserve volume is enabled and the assay LIS code is not the same as the assay Test ID. These issues could potentially lead to erroneous results or delay reporting results for multiple analytes.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Beckman Coulter, Inc.
Distributor
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Croatia, Czech Republic, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Kingdom.
Lot
UDI/DI 15099590732103, Serial Numbers: 300116, 300117, 300118, 300123, 300127, 300126, 300128, 300129, 300130, 300131, 300133, 300134, 300135, 300136, 300138, 300139, 300137, 300140, 300142, 300141, 300144, 300143, 300145, 300148 300146, 300147, 300149, 300150, 300151, 300152, 300155, 300153, 300154, 300156, 300158, 300160, 300163, 300161, 300162, 300164, 300167, 300170, 300169, 300171, 300168, 300166, 300172, 300176, 300173, 300178, 300181, 300179, 300183, 300182, 300188, 300189, 300190, 300191, 300192, 300195, 300196, 300193, 300194, 300197, 300199, 300198, 300201, 300200, 300202, 300203, 300205, 300204 300207, 300206, 300208, 300209, 300212, 300210, 300211, 300213, 300214, 300215, 300217, 300216, 300220, 300218, 300219, 300221, 300223, 300224, 300222, 300227, 300228, 300229, 300231, 300230, 300235, 300236, 300234, 300237, 300232, 300239, 300233, 300238, 300240, 300243, 300241, 300242, 300247, 300246, 300249, 300253, 300244, 300250, 300252, 300248, 300251, 300256, 300257, 300254, 300258, 300255, 300259, 300260, 300261, 300264, 300263, 300262; running software version 1.16 or prior
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was DxI 9000 Access Immunoassay Analyzer running software version 1.16 or prior, recalled?

Yes. FDA published recall Z-0982-2024 on February 14, 2024. Reported hazard: There are potential performance issues found in the DxI 9000 Access Immunoassay Analyzer, including: 1) Access Ultrasensitive Insulin users who configure SI units (pmol/L) applies an incorrect conversion factor; 2) System provides numerical results that are below the lowest reportable result if laboratory chooses to report assay results in units of measurement other than the defined default units; 3) A result of 0 incorrectly reported when the system utilizes an expression that utilizes non-numerical symbols (e.g., > or <) to derive a calculated test as part of result reporting; 4) Reserve volume is enabled and the assay LIS code is not the same as the assay Test ID. These issues could potentially lead to erroneous results or delay reporting results for multiple analytes.

What should I do if I have this product?

Voluntary: Firm initiated · Letter