DxI 9000 Access Immunoassay Analyzer running software version 1.16 or prior,

Published February 14, 2024 · Recall initiated: November 21, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There are potential performance issues found in the DxI 9000 Access Immunoassay Analyzer, including: 1) Access Ultrasensitive Insulin users who configure SI units (pmol/L) applies an incorrect conversion factor; 2) System provides numerical results that are below the lowest reportable result if laboratory chooses to report assay results in units of measurement other than the defined default units; 3) A result of 0 incorrectly reported when the system utilizes an expression that utilizes non-numerical symbols (e.g., > or <) to derive a calculated test as part of result reporting; 4) Reserve volume is enabled and the assay LIS code is not the same as the assay Test ID. These issues could potentially lead to erroneous results or delay reporting results for multiple analytes.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Beckman Coulter, Inc.
Distributor
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Croatia, Czech Republic, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Kingdom.
Lot
UDI/DI 15099590732103, Serial Numbers: 300116, 300117, 300118, 300123, 300127, 300126, 300128, 300129, 300130, 300131, 300133, 300134, 300135, 300136, 300138, 300139, 300137, 300140, 300142, 300141, 300144, 300143, 300145, 300148 300146, 300147, 300149, 300150, 300151, 300152, 300155, 300153, 300154, 300156, 300158, 300160, 300163, 300161, 300162, 300164, 300167, 300170, 300169, 300171, 300168, 300166, 300172, 300176, 300173, 300178, 300181, 300179, 300183, 300182, 300188, 300189, 300190, 300191, 300192, 300195, 300196, 300193, 300194, 300197, 300199, 300198, 300201, 300200, 300202, 300203, 300205, 300204 300207, 300206, 300208, 300209, 300212, 300210, 300211, 300213, 300214, 300215, 300217, 300216, 300220, 300218, 300219, 300221, 300223, 300224, 300222, 300227, 300228, 300229, 300231, 300230, 300235, 300236, 300234, 300237, 300232, 300239, 300233, 300238, 300240, 300243, 300241, 300242, 300247, 300246, 300249, 300253, 300244, 300250, 300252, 300248, 300251, 300256, 300257, 300254, 300258, 300255, 300259, 300260, 300261, 300264, 300263, 300262; running software version 1.16 or prior
Risk classification
Class II
Area (FDA)
device