FDAFDA · Food and drugs·Recall number Z-0990-2014

Orbit soft Infusion sets, 9 mm soft cannula, P10 (42") tubing; Single-Use Set Orbit Infusion Sets are sterile, non-pyrogenic, single-use devices intended to be used for subcutaneous delivery of medication , for example insulin, from an external infusion pump.

Published February 26, 2014 · Recall initiated: December 20, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Perrigo was notified by our supplier, Ypsomed AG, that they are recalling these batches due to being manufactured with a non-conforming female luer-lock connector. The luer-lock connectors are incompatible with standard male luer-lock connectors for use in connection to compatible infusion pumps. Perrigo became was notified of a quality complaint related to this issue on November 20, 2013 from a

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Perrigo Diabetes Care
Distributor
Nationwide Distribution including GA & TX
Lot
400064, 400069, 400070, & 400072 UPC: 7613117009502
Risk classification
Class II
Area (FDA)
device
Country
United States