Superstand stand-up wheelchair HPS-2 A standup wheelchair is a device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. The device incorporates an external manually controlled mechanical system that is intended to raise a paraplegic to an upright position by means of an elevating seat.
Published February 26, 2014 · Recall initiated: May 16, 2013
Summary
FDA records medical device Class II recall Z-0993-2014, published February 26, 2014 and initiated May 16, 2013. Recalling firm: The Standing Company; Status: Terminated; Reported quantity: approximately 119; Lot: Serial No: 1306, 1313, 1342, 1349, 1362, 1358, 1359, 1392, 1391, 1406, 1402, 1407, 1426, 1441, 1451, 1018, 1453, 1399, 1523, 1477, 1480, 1482, 1497, 1494, 1505, 1516, 1533, 1537, 1536, 1535, 1542, 1544, 1549, 1543, 1550, 1562, 1563, 1566, 1580, 1572, 1573, 1574, 1571, 1578, 1583, 1641, 1593, 1591, 1592, 1599, 1604, 1600, 1615, 1616, 1614, 1622, 1631, 1637, 1643, 1635, 1640, 1634, 1648, 1666, 1652, 1657, 1663, 1664, 1676, 1689, 1665, 1683, 1691, 1688, 1701, 1703, 1711, 1698, 1712, 1719, 1728, 1731, 1726, 1740, 2165, 2172, 2175, 2177, 2257, 2182, 2197, 2205, 2258, 2259, 2262, 2270, 2283, 2279, 2281, 2298, 2306, 2310, 2312, 2330, 2334, 2503, 2515, 2328, 2514, 2523, 2525, 2550, 2561, 2568, 2581, 2573, 1684, and 1427. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Power Superstand Standing Wheel chair, flex shaft (the flexible wand which supports the joystick control module) may fracture if it is bent repeatedly and /or beyond 90 degrees. A newly designed Flexshaft which corrects this problem is available at no charge.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- The Standing Company
- Distributor
- Nationwide Distribution including AL, CA, DE, IA, IL , IN, KS, KY, MI, MO, NJ, OH, OK, TX, WA, WI,and WV.
- Lot
- Serial No: 1306, 1313, 1342, 1349, 1362, 1358, 1359, 1392, 1391, 1406, 1402, 1407, 1426, 1441, 1451, 1018, 1453, 1399, 1523, 1477, 1480, 1482, 1497, 1494, 1505, 1516, 1533, 1537, 1536, 1535, 1542, 1544, 1549, 1543, 1550, 1562, 1563, 1566, 1580, 1572, 1573, 1574, 1571, 1578, 1583, 1641, 1593, 1591, 1592, 1599, 1604, 1600, 1615, 1616, 1614, 1622, 1631, 1637, 1643, 1635, 1640, 1634, 1648, 1666, 1652, 1657, 1663, 1664, 1676, 1689, 1665, 1683, 1691, 1688, 1701, 1703, 1711, 1698, 1712, 1719, 1728, 1731, 1726, 1740, 2165, 2172, 2175, 2177, 2257, 2182, 2197, 2205, 2258, 2259, 2262, 2270, 2283, 2279, 2281, 2298, 2306, 2310, 2312, 2330, 2334, 2503, 2515, 2328, 2514, 2523, 2525, 2550, 2561, 2568, 2581, 2573, 1684, and 1427.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASynthes Matrix Mandible Short Threaded Drill Guide intended for oral, maxillofacial surgery; trauma andFebruary 26, 2014
- FDAIntegra Cusa Excel Rx Only CEM" Nosecone for CUSA® Excel System Product Usage: Intended for use CUSAFebruary 26, 2014
- FDAMORPH ACCESSPRO Steerable Introducer; 6F introducer, 90 cm working length: Sterile in unopened, undamagedFebruary 26, 2014
- FDAMedtronic Select 3D Arterial Cannulae. Vented, models 78618 (18 Fr.) and 78620 (20 Fr.). Non-Vented Cannulae,February 26, 2014
- FDAOrbit soft Infusion sets, 9 mm soft cannula, P10 (42") tubing; Single-Use Set Orbit Infusion Sets are sterile,February 26, 2014
- FDAMedtronic Select 3D Arterial Cannulae. Vented, model 78622 (22 Fr.). Non-Vented Cannulae, model 78722 (22February 26, 2014
Questions about this recall
Was Superstand stand-up wheelchair HPS-2 A standup wheelchair is a device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. The device incorporates an external manually controlled mechanical system that is intended to raise a paraplegic to an upright position by means of an elevating seat. recalled?
Yes. FDA published recall Z-0993-2014 on February 26, 2014. Reported hazard: Power Superstand Standing Wheel chair, flex shaft (the flexible wand which supports the joystick control module) may fracture if it is bent repeatedly and /or beyond 90 degrees. A newly designed Flexshaft which corrects this problem is available at no charge.
What should I do if I have this product?
Voluntary: Firm initiated · Letter