FDAFDA · Food and drugs·Recall number Z-0993-2014

Superstand stand-up wheelchair HPS-2 A standup wheelchair is a device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. The device incorporates an external manually controlled mechanical system that is intended to raise a paraplegic to an upright position by means of an elevating seat.

Published February 26, 2014 · Recall initiated: May 16, 2013

Summary

FDA records medical device Class II recall Z-0993-2014, published February 26, 2014 and initiated May 16, 2013. Recalling firm: The Standing Company; Status: Terminated; Reported quantity: approximately 119; Lot: Serial No: 1306, 1313, 1342, 1349, 1362, 1358, 1359, 1392, 1391, 1406, 1402, 1407, 1426, 1441, 1451, 1018, 1453, 1399, 1523, 1477, 1480, 1482, 1497, 1494, 1505, 1516, 1533, 1537, 1536, 1535, 1542, 1544, 1549, 1543, 1550, 1562, 1563, 1566, 1580, 1572, 1573, 1574, 1571, 1578, 1583, 1641, 1593, 1591, 1592, 1599, 1604, 1600, 1615, 1616, 1614, 1622, 1631, 1637, 1643, 1635, 1640, 1634, 1648, 1666, 1652, 1657, 1663, 1664, 1676, 1689, 1665, 1683, 1691, 1688, 1701, 1703, 1711, 1698, 1712, 1719, 1728, 1731, 1726, 1740, 2165, 2172, 2175, 2177, 2257, 2182, 2197, 2205, 2258, 2259, 2262, 2270, 2283, 2279, 2281, 2298, 2306, 2310, 2312, 2330, 2334, 2503, 2515, 2328, 2514, 2523, 2525, 2550, 2561, 2568, 2581, 2573, 1684, and 1427. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Power Superstand Standing Wheel chair, flex shaft (the flexible wand which supports the joystick control module) may fracture if it is bent repeatedly and /or beyond 90 degrees. A newly designed Flexshaft which corrects this problem is available at no charge.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
The Standing Company
Distributor
Nationwide Distribution including AL, CA, DE, IA, IL , IN, KS, KY, MI, MO, NJ, OH, OK, TX, WA, WI,and WV.
Lot
Serial No: 1306, 1313, 1342, 1349, 1362, 1358, 1359, 1392, 1391, 1406, 1402, 1407, 1426, 1441, 1451, 1018, 1453, 1399, 1523, 1477, 1480, 1482, 1497, 1494, 1505, 1516, 1533, 1537, 1536, 1535, 1542, 1544, 1549, 1543, 1550, 1562, 1563, 1566, 1580, 1572, 1573, 1574, 1571, 1578, 1583, 1641, 1593, 1591, 1592, 1599, 1604, 1600, 1615, 1616, 1614, 1622, 1631, 1637, 1643, 1635, 1640, 1634, 1648, 1666, 1652, 1657, 1663, 1664, 1676, 1689, 1665, 1683, 1691, 1688, 1701, 1703, 1711, 1698, 1712, 1719, 1728, 1731, 1726, 1740, 2165, 2172, 2175, 2177, 2257, 2182, 2197, 2205, 2258, 2259, 2262, 2270, 2283, 2279, 2281, 2298, 2306, 2310, 2312, 2330, 2334, 2503, 2515, 2328, 2514, 2523, 2525, 2550, 2561, 2568, 2581, 2573, 1684, and 1427.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Superstand stand-up wheelchair HPS-2 A standup wheelchair is a device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. The device incorporates an external manually controlled mechanical system that is intended to raise a paraplegic to an upright position by means of an elevating seat. recalled?

Yes. FDA published recall Z-0993-2014 on February 26, 2014. Reported hazard: Power Superstand Standing Wheel chair, flex shaft (the flexible wand which supports the joystick control module) may fracture if it is bent repeatedly and /or beyond 90 degrees. A newly designed Flexshaft which corrects this problem is available at no charge.

What should I do if I have this product?

Voluntary: Firm initiated · Letter