Ossur Skull Pin, Sterile Titanium Pin, Part/Description: 516TS/ Skull Pin 2-1/4in TI Sterl 1PK; 516V2S/ Skull Pin 3in TI Sterl 1PK; 520TS/ Skull Pin 2-1/4in TI Sterl 5PK; 520V2S/ Skull Pin 3in TI Sterl 5 PK
Published March 20, 2019 · Recall initiated: June 21, 2018
Summary
FDA records medical device Class II recall Z-0997-2019, published March 20, 2019 and initiated June 21, 2018. Recalling firm: Ossur Americas; Status: Terminated; Reported quantity: 372; Lot: All lots of the products manufactured from 06/30/2009 - 12/13/2018 Titanium Pin Label Mistakenly lists Ceramic Tip, Part/Lot numbers: 516TS/ AL18021418 and AL17022109; 520TS/ AL18021417 and AL15091012. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line description should state titanium pin.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ossur Americas
- Distributor
- U.S.: HI, AL, CA, WA, AZ, TX, NM, NE, MI, IL, MN, MS, IN, OH, KY, TN, AK, AL, UT, SC, NC, GA, PA, FL, NY, DE, RI, VA, MA; Foreign: Australia, Germany, New Zealand, Belgium
- Lot
- All lots of the products manufactured from 06/30/2009 - 12/13/2018 Titanium Pin Label Mistakenly lists Ceramic Tip, Part/Lot numbers: 516TS/ AL18021418 and AL17022109; 520TS/ AL18021417 and AL15091012
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOssur OB Resolve, Sterile, Part/ Description: 505300D/ OB RESOLVE STERILE RING SML; 505400D/ OB RESOLVE RINGMarch 20, 2019
- FDAOssur Skull Pin, Sterile Ceramic Tip Pin, Part/ Description: 516CS/ Skull Pin 2-1/4in CT Sterl 1PK; 516DS/March 20, 2019
- FDAOssur, ReSolve Halo Vest, Part/ Description: 545700K/ ReSolve Halo Vest-shrt kodel; 545700L/ ReSolve HaloMarch 20, 2019
- FDAAMSCO 5000 Series Washer/Disinfector Model # 5052March 20, 2019
- FDADurex Pleasure Pack (60 ct), synthetic polisoprene male condom, SKU 02340-98729-00 02340-98729-00March 20, 2019
- FDAVITROS Chemistry Products Na+ Slides, Catalog Number 8379034 For in vitro diagnostic use only. VITROS Na+March 20, 2019
Questions about this recall
Was Ossur Skull Pin, Sterile Titanium Pin, Part/Description: 516TS/ Skull Pin 2-1/4in TI Sterl 1PK; 516V2S/ Skull Pin 3in TI Sterl 1PK; 520TS/ Skull Pin 2-1/4in TI Sterl 5PK; 520V2S/ Skull Pin 3in TI Sterl 5 PK recalled?
Yes. FDA published recall Z-0997-2019 on March 20, 2019. Reported hazard: Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line description should state titanium pin.
What should I do if I have this product?
Voluntary: Firm initiated · Letter