FDAFDA · Food and drugs·Recall number Z-0997-2019
Ossur Skull Pin, Sterile Titanium Pin, Part/Description: 516TS/ Skull Pin 2-1/4in TI Sterl 1PK; 516V2S/ Skull Pin 3in TI Sterl 1PK; 520TS/ Skull Pin 2-1/4in TI Sterl 5PK; 520V2S/ Skull Pin 3in TI Sterl 5 PK
Published March 20, 2019 · Recall initiated: June 21, 2018
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line description should state titanium pin.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Ossur Americas
- Distributor
- U.S.: HI, AL, CA, WA, AZ, TX, NM, NE, MI, IL, MN, MS, IN, OH, KY, TN, AK, AL, UT, SC, NC, GA, PA, FL, NY, DE, RI, VA, MA; Foreign: Australia, Germany, New Zealand, Belgium
- Lot
- All lots of the products manufactured from 06/30/2009 - 12/13/2018 Titanium Pin Label Mistakenly lists Ceramic Tip, Part/Lot numbers: 516TS/ AL18021418 and AL17022109; 520TS/ AL18021417 and AL15091012
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States