AMSCO 5000 Series Washer/Disinfector Model # 5052
Published March 20, 2019 · Recall initiated: December 20, 2018
Summary
FDA records medical device Class II recall Z-0996-2019, published March 20, 2019 and initiated December 20, 2018. Recalling firm: Steris Corporation; Status: Terminated; Reported quantity: 68; Lot: Serial Numbers 3627418005 3601518006 3612718013 3633817009 3632318011 3626718014 3625114008 3600515028 3600515029 3600515030 3628516022 3626718013 3631618008 3627418007 3624514006 3628818001 3608518007 3634517006 3630218008 3605316016 3600917004 3628218007 3628818003 3629518009 3626718009 3626718011 3627418001 3627418002 3630218012 3630218013 3630918013 3632318012 3626718016 3634517005 3631414004 3603618005 3607217008 3630318013 3626817008 3628317012 3616615026 3630918014 3633018006 3628218009 3632318009 3628218011 3623615011A 3601518012 3612817007 3627418009 3628218006 3627418008 3631618007 3617315011 3614818009 3624017008 3629215012 3614516007 3628218010 3610216025 3634313010 3611715016 3628818005 3615617008 3624017009 3616516020 3627418004 3627418006. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The software in the systems may not process the cycle originally intended. This could result in damage to the various medical devices and instruments processed in the washer/disinfector or potential for inadequate cleaning or disinfection.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Steris Corporation
- Distributor
- The products were distributed to the following US states: AL, AR, CA, CO, FL, GA, IL, LA, MA, MD, ME, MI, NC, NJ, NM, NY, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV. The products were distributed to the following foreign countries: Bahrain, Brazil, Canada, China, Saudi Arabia, South Korea, and Taiwan.
- Lot
- Serial Numbers 3627418005 3601518006 3612718013 3633817009 3632318011 3626718014 3625114008 3600515028 3600515029 3600515030 3628516022 3626718013 3631618008 3627418007 3624514006 3628818001 3608518007 3634517006 3630218008 3605316016 3600917004 3628218007 3628818003 3629518009 3626718009 3626718011 3627418001 3627418002 3630218012 3630218013 3630918013 3632318012 3626718016 3634517005 3631414004 3603618005 3607217008 3630318013 3626817008 3628317012 3616615026 3630918014 3633018006 3628218009 3632318009 3628218011 3623615011A 3601518012 3612817007 3627418009 3628218006 3627418008 3631618007 3617315011 3614818009 3624017008 3629215012 3614516007 3628218010 3610216025 3634313010 3611715016 3628818005 3615617008 3624017009 3616516020 3627418004 3627418006
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116October 21, 2026
- FDAVERIFY Biological Steam Test Pack Model/Catalog Number: S3069October 14, 2026
- FDADurex Pleasure Pack (60 ct), synthetic polisoprene male condom, SKU 02340-98729-00 02340-98729-00March 20, 2019
- FDAOssur Skull Pin, Sterile Titanium Pin, Part/Description: 516TS/ Skull Pin 2-1/4in TI Sterl 1PK; 516V2S/ SkullMarch 20, 2019
Questions about this recall
Was AMSCO 5000 Series Washer/Disinfector Model # 5052 recalled?
Yes. FDA published recall Z-0996-2019 on March 20, 2019. Reported hazard: The software in the systems may not process the cycle originally intended. This could result in damage to the various medical devices and instruments processed in the washer/disinfector or potential for inadequate cleaning or disinfection.
What should I do if I have this product?
Voluntary: Firm initiated · Letter