Water Filter MAJ-2318. This MAJ-2318 water filter has been designed to be used with Olympus endoscope reprocessor for prevent contamination of the rinse water. The MAJ-2318 water filter is compatible with OER-ELITE and OER-MINI.
Published February 14, 2024 · Recall initiated: January 16, 2024
Summary
FDA records medical device Class II recall Z-1009-2024, published February 14, 2024 and initiated January 16, 2024. Recalling firm: Olympus Corporation of the Americas; Status: Ongoing; Reported quantity: 877 MAJ-2318, 119 OER-Elite, 3 OER-MINI; Lot: UDI-DI 04953170401473 Lots 36A, 37A, 38A Affected product may have also been packaged within OER boxes (OER-Elite and OER-MINI). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Water filter may have been assembled incorrectly, resulting in unfiltered tap water being used for reprocessing. This may expose a patient to a potentially contaminated device, presenting a risk of infection or sepsis.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Olympus Corporation of the Americas
- Distributor
- Domestic US distribution nationwide.
- Lot
- UDI-DI 04953170401473 Lots 36A, 37A, 38A Affected product may have also been packaged within OER boxes (OER-Elite and OER-MINI)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOlympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.October 7, 2026
- FDAOlympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.October 7, 2026
- FDAOlympus Tripp-Lite P580-025 25ft DisplayPort Cable, Model Number P580-025, an accessory used in conjunctionOctober 7, 2026
- FDAOlympus BiCOAG Hemostasis Probe 10Fr Coaxial Connector. Model/Catalog Number: CD-B610LA.September 30, 2026
- FDABrasseler 8S RA SHP RD (Super-Sharp Round) H1S.21.023, REF 001156T0), carbide dental bur, 10 burs/package, Rx.February 14, 2024
- FDAPulseSpray Infusion System devices including: PULSE SPRAY 5F/135/20, UPN: H787125004045, REF: 12500404; PulseFebruary 14, 2024
Questions about this recall
Was Water Filter MAJ-2318. This MAJ-2318 water filter has been designed to be used with Olympus endoscope reprocessor for prevent contamination of the rinse water. The MAJ-2318 water filter is compatible with OER-ELITE and OER-MINI. recalled?
Yes. FDA published recall Z-1009-2024 on February 14, 2024. Reported hazard: Water filter may have been assembled incorrectly, resulting in unfiltered tap water being used for reprocessing. This may expose a patient to a potentially contaminated device, presenting a risk of infection or sepsis.
What should I do if I have this product?
Voluntary: Firm initiated · Letter