Roche Molecular Systems cobas x 480 System 50 mL reagent reservoir, Catalog number 05232732001. Used with the cobas® CT/NG 4800 System, for in vitro diagnostics.

Published April 10, 2013 · Recall initiated: March 1, 2013

Summary

FDA records medical device Class II recall Z-1028-2013, published April 10, 2013 and initiated March 1, 2013. Recalling firm: Roche Diagnostics Operations, Inc.; Status: Terminated; Reported quantity: 189 units; Lot: Lot / SN : 477415,477416, 477898, and 477899. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Certain lots of Reagent reservoirs, may not have a separation in the lower part of the reservoir. As a consequence, pipetting errors can occur due to uneven distribution of reagent. These pipetting errors may lead to delayed results and wasted reagent.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Roche Diagnostics Operations, Inc.
Distributor
Nationwide Distribution including Puerto Rico and the states of MO, NE, CO, MT, NJ, FL, WA, TX, GA, NY, KS, NH, VA, HI, PA, MI, IA, CA, NC, AZ, UT, LA, IL, TN, ME, MN, KY, CT, OR, NV, AL and NM.
Lot
Lot / SN : 477415,477416, 477898, and 477899
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Roche Molecular Systems cobas x 480 System 50 mL reagent reservoir, Catalog number 05232732001. Used with the cobas® CT/NG 4800 System, for in vitro diagnostics. recalled?

Yes. FDA published recall Z-1028-2013 on April 10, 2013. Reported hazard: Certain lots of Reagent reservoirs, may not have a separation in the lower part of the reservoir. As a consequence, pipetting errors can occur due to uneven distribution of reagent. These pipetting errors may lead to delayed results and wasted reagent.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit