Ingenuity TF PET/CT. Diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.
Published April 10, 2013 · Recall initiated: December 20, 2012
Summary
FDA records medical device Class II recall Z-1029-2013, published April 10, 2013 and initiated December 20, 2012. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 1; Lot: Model #: 882456; Serial #: 2004. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Issues have been detected in the Ingenuity TF PET/CT currently running software version 4.0.60605.0 that if they were to re-occur, could cause a risk for patients, users or service personnel, including inaccurate standard uptake values (SUV). Philips Medical Systems (Cleveland), Inc. is issuing a software update to address the issues.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Distributor
- Distributed to one customer in Ohio.
- Lot
- Model #: 882456; Serial #: 2004
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.September 21, 2022
- FDABrilliance 64 Model # 728231, computed tomography x-ray systemAugust 14, 2019
- FDAIngenuity Core128 Model # 728323, computed tomography x-ray systemAugust 14, 2019
- FDABrightView XCT , Model Number 882482July 31, 2019
- FDARoche Molecular Systems cobas x 480 System 50 mL reagent reservoir, Catalog number 05232732001. Used with theApril 10, 2013
- FDAIntegra Ojemann Cortical Stimulator; Product / Catalogue No: OCS2. Intended for intraoperative corticalApril 10, 2013
Questions about this recall
Was Ingenuity TF PET/CT. Diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. recalled?
Yes. FDA published recall Z-1029-2013 on April 10, 2013. Reported hazard: Issues have been detected in the Ingenuity TF PET/CT currently running software version 4.0.60605.0 that if they were to re-occur, could cause a risk for patients, users or service personnel, including inaccurate standard uptake values (SUV). Philips Medical Systems (Cleveland), Inc. is issuing a software update to address the issues.
What should I do if I have this product?
Voluntary: Firm initiated · Letter