Back to search
FDAFDA · Food and drugs·Recall number Z-1028-2017

Fresenius 2008K2 Series : Hemodialysis System

Published January 25, 2017 · Recall initiated: December 21, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Fresenius Medical Care Renal Therapies Group, LLC
Distributor
Nationwide Foreign: Canada, Mexico, Jamaica, Germany, Guam
Lot
All serial numbers
Risk classification
Class II
Area (FDA)
device
Country
United States