Fresenius 2008K@home Series : Hemodialysis System

Published January 25, 2017 · Recall initiated: December 21, 2016

Summary

FDA records medical device Class II recall Z-1029-2017, published January 25, 2017 and initiated December 21, 2016. Recalling firm: Fresenius Medical Care Renal Therapies Group, LLC; Status: Terminated; Reported quantity: 1400 units (US) 556 units (OUS); Lot: All serial numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Fresenius Medical Care Renal Therapies Group, LLC
Distributor
Nationwide Foreign: Canada, Mexico, Jamaica, Germany, Guam
Lot
All serial numbers
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Fresenius 2008K@home Series : Hemodialysis System recalled?

Yes. FDA published recall Z-1029-2017 on January 25, 2017. Reported hazard: When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine

What should I do if I have this product?

Voluntary: Firm initiated · Letter