FDAFDA · Food and drugs·Recall number Z-1036-2017
Olympus Scientific Solutions Americas Corporation (OSSA) VANTA¿ XRF Analyzer. Model Vanta VCR and Vanta VMR Analytical X-ray systems
Published February 8, 2017 · Recall initiated: January 7, 2017
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It was discovered under rare circumstances the LED warning light circuit could fail. If this happens, while the instrument is emitting X-rays during at test, the LED warning lights will not illuminate. The test runs normally and will execute as before, but the light will not be on.
What to do
FDA Mandated · Letter
Recall details
- Manufacturer
- Olympus Scientific Solutions Americas
- Distributor
- US Distribution to the states of : PA, MA, CO, OH, NJ, GA, CA, TX, IL, VA, MN, NH, WA, SC, AZ, LA, MD, OK, IN, NY, NC and KS.
- Lot
- affects all units.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.