Back to search
FDAFDA · Food and drugs·Recall number Z-1036-2017

Olympus Scientific Solutions Americas Corporation (OSSA) VANTA¿ XRF Analyzer. Model Vanta VCR and Vanta VMR Analytical X-ray systems

Published February 8, 2017 · Recall initiated: January 7, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

It was discovered under rare circumstances the LED warning light circuit could fail. If this happens, while the instrument is emitting X-rays during at test, the LED warning lights will not illuminate. The test runs normally and will execute as before, but the light will not be on.

What to do

FDA Mandated · Letter

Recall details

Manufacturer
Olympus Scientific Solutions Americas
Distributor
US Distribution to the states of : PA, MA, CO, OH, NJ, GA, CA, TX, IL, VA, MN, NH, WA, SC, AZ, LA, MD, OK, IN, NY, NC and KS.
Lot
affects all units.
Risk classification
Class II
Area (FDA)
device
Country
United States