Olympus Scientific Solutions Americas Corporation (OSSA) VANTA® XRF Analyzer. Model Vanta VCR and Vanta VMR Analytical X-ray systems

Published February 8, 2017 · Recall initiated: January 7, 2017

Summary

FDA records medical device Class II recall Z-1036-2017, published February 8, 2017 and initiated January 7, 2017. Recalling firm: Olympus Scientific Solutions Americas; Status: Terminated; Reported quantity: US 74; Lot: affects all units. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

It was discovered under rare circumstances the LED warning light circuit could fail. If this happens, while the instrument is emitting X-rays during at test, the LED warning lights will not illuminate. The test runs normally and will execute as before, but the light will not be on.

What to do

FDA Mandated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Olympus Scientific Solutions Americas
Distributor
US Distribution to the states of : PA, MA, CO, OH, NJ, GA, CA, TX, IL, VA, MN, NH, WA, SC, AZ, LA, MD, OK, IN, NY, NC and KS.
Lot
affects all units.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Olympus Scientific Solutions Americas Corporation (OSSA) VANTA® XRF Analyzer. Model Vanta VCR and Vanta VMR Analytical X-ray systems recalled?

Yes. FDA published recall Z-1036-2017 on February 8, 2017. Reported hazard: It was discovered under rare circumstances the LED warning light circuit could fail. If this happens, while the instrument is emitting X-rays during at test, the LED warning lights will not illuminate. The test runs normally and will execute as before, but the light will not be on.