Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003
Published February 8, 2017 · Recall initiated: December 27, 2016
Summary
FDA records medical device Class II recall Z-1109-2017, published February 8, 2017 and initiated December 27, 2016. Recalling firm: Immuno-Mycologics, Inc; Status: Terminated; Reported quantity: 630 units; Lot: 161028133653JH, 161003111544KH, and 161007105036JG. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Immuno-Mycologics, Inc
- Distributor
- Nationwide throughout the US
- Lot
- 161028133653JH, 161003111544KH, and 161007105036JG
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAImmy, Myco DDR Trident, Bulk Neutralization Buffer B (60 X 30 mL), REF: TBPN67-60 For processing of clinicalMay 6, 2026
- FDAOlympus Scientific Solutions Americas Corporation (OSSA) VANTA® XRF Analyzer. Model Vanta VCR and Vanta VMRFebruary 8, 2017
- FDASPS-1, Static Preservation Solution [1 or 2 liter bags], Sterile; Organ Recovery Systems, One Pierce Place,February 8, 2017
- FDAXambo, 80 boxes of bulk capsules. Each containing 12,053 capsules in each box.February 8, 2017
- FDASwishTapered ImplantFebruary 8, 2017
- FDAMikesell's Nacho Cheese FLAVORED TORTILLA CHIPS 2.25 OZ (64g) bag UPC 0 71104 00554 3February 8, 2017
Questions about this recall
Was Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003 recalled?
Yes. FDA published recall Z-1109-2017 on February 8, 2017. Reported hazard: The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail