Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003

Published February 8, 2017 · Recall initiated: December 27, 2016

Summary

FDA records medical device Class II recall Z-1109-2017, published February 8, 2017 and initiated December 27, 2016. Recalling firm: Immuno-Mycologics, Inc; Status: Terminated; Reported quantity: 630 units; Lot: 161028133653JH, 161003111544KH, and 161007105036JG. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Immuno-Mycologics, Inc
Distributor
Nationwide throughout the US
Lot
161028133653JH, 161003111544KH, and 161007105036JG
Risk classification
Class II
Area (FDA)
device
Country
United States