FDAFDA · Food and drugs·Recall number Z-1047-2013
Siemens Dimension(R) EXL(TM) LM STM (Siemens Material Number 106393377). Intended to measure a variety of analytes in human body fluids.
Published April 10, 2013 · Recall initiated: February 27, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to the current limit set for System Check LOCI results. The errors have no impact on results, but they may cause customer inconvenience. Siemens h
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- Worldwide Distribution.
- Lot
- Siemens Material Number 106393377
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States