Siemens Dimension(R) EXL(TM) 200 (Siemens Material Number 10636929). Intended to measure a variety of analytes in human body fluids.
Published April 10, 2013 · Recall initiated: February 27, 2013
Summary
FDA records medical device Class III recall Z-1048-2013, published April 10, 2013 and initiated February 27, 2013. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Terminated; Reported quantity: 2207 total; Lot: Siemens Material Number 10636929. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to the current limit set for System Check LOCI results. The errors have no impact on results, but they may cause customer inconvenience. Siemens h
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- Worldwide Distribution.
- Lot
- Siemens Material Number 10636929
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAtellica CH Revised C-Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223;October 7, 2026
- FDABeckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mLAugust 26, 2026
- FDASyva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mLAugust 26, 2026
- FDASyva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219ULAugust 12, 2026
- FDASiemens Dimension(R) EXL(TM) LM STM (Siemens Material Number 106393377). Intended to measure a variety ofApril 10, 2013
- FDASiemens Dimension(R) Phosphorus (PHOS) Flex(R) reagent cartridge (DF61) The Phosphorus (PHOS) method used onApril 10, 2013
Questions about this recall
Was Siemens Dimension(R) EXL(TM) 200 (Siemens Material Number 10636929). Intended to measure a variety of analytes in human body fluids. recalled?
Yes. FDA published recall Z-1048-2013 on April 10, 2013. Reported hazard: Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to the current limit set for System Check LOCI results. The errors have no impact on results, but they may cause customer inconvenience. Siemens h
What should I do if I have this product?
Voluntary: Firm initiated · Letter