GE Healthcare - Giraffe and Panda Bag and Mask Resuscitation System;Giraffe Stand-alone Infant Resuscitation System. Product Usage: The Bag and Mask Resuscitation System provides the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant. These are clinical practices that represent the established standard of care.

Published March 5, 2014 · Recall initiated: October 1, 2013

Summary

FDA records medical device Class I recall Z-1071-2014, published March 5, 2014 and initiated October 1, 2013. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 12, 373 units total, 8,807 = US, 3,394 = OUS, 172 = unknown; Lot: QABS50026, QABR50066, QABR50068, QABR50133, QABS50023, QABR50102, QABR50103, QABR50064, QABR50065, QABR50067, QABS50070, QABS50069, QABR50275, QABS50027, QABR50274. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
GE Healthcare, LLC
Distributor
Worldwide Distribution - USA Nationwide including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL,GA. HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and the countries of Albania, Austria, Australia, Bahrain, Belarus, Belgium, Canada, China, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Greece, Guatemala, Germany, Honduras, Hong Kong, Hungry, India, Indonesia, Ireland, Italy, Kazakhstan, Korea, Lithuania, Malaysia, Mexico, Netherlands, Norway, Peru, Poland, Romania, Russia, Saudi Arabia, South Africia, Spain, Sweden, Switzerland Turkey, United Arab, United Kingdom, and Venezuela.
Lot
QABS50026, QABR50066, QABR50068, QABR50133, QABS50023, QABR50102, QABR50103, QABR50064, QABR50065, QABR50067, QABS50070, QABS50069, QABR50275, QABS50027, QABR50274
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was GE Healthcare - Giraffe and Panda Bag and Mask Resuscitation System;Giraffe Stand-alone Infant Resuscitation System. Product Usage: The Bag and Mask Resuscitation System provides the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant. These are clinical practices that represent the established standard of care. recalled?

Yes. FDA published recall Z-1071-2014 on March 5, 2014. Reported hazard: Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.

What should I do if I have this product?

Voluntary: Firm initiated · Letter