VITROS 350 Chemistry System, Catalog Number 6802153, Unique Device Identifier Number 10758750002054; IVD. Product Usage: For in vitro diagnostic use. The VITROS 350 Chemistry System performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides.
Published March 16, 2016 · Recall initiated: December 21, 2015
Summary
FDA records medical device Class II recall Z-1072-2016, published March 16, 2016 and initiated December 21, 2015. Recalling firm: Ortho-Clinical Diagnostics; Status: Terminated; Reported quantity: US = 885; Foreign: 3221; Lot: Software Version 9.4 and below; Serial Numbers J27000001 - 27004539. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Increased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chemistry Systems when using Calibrator Kit 9, Lot 954. A trend of complaints regarding customers actions following U90-382 or 6LU condition codes was noted (See RES 72289 - VITROS Calibrator Kit 9, lot 954 recall).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ortho-Clinical Diagnostics
- Distributor
- Worldwide Distribution - US Nationwide including Puerto Rico, and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela
- Lot
- Software Version 9.4 and below; Serial Numbers J27000001 - 27004539
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVITROS Performance Verifier IIApril 3, 2024
- FDAVITROS Performance Verifier IApril 3, 2024
- FDAVITROS XT Chemistry Products TBIL-ALKP Slides (ALKP is not affected) In Vitro Diagnostic use only - The TBILSeptember 2, 2020
- FDAVITROS Chemistry Product Performance Verifier I, Product Code 8231474, UDI # 10758750004577 - Product Usage:June 24, 2020
- FDAConformis iTotal Cruciate Retaining (CR) Knee Replacement System Model: TCR-111-1111March 16, 2016
- FDAVITROS 5600 Chemistry System, Catalog Number 6802413, Unique Device Identifier Number 10758750002740; IVD.March 16, 2016
Questions about this recall
Was VITROS 350 Chemistry System, Catalog Number 6802153, Unique Device Identifier Number 10758750002054; IVD. Product Usage: For in vitro diagnostic use. The VITROS 350 Chemistry System performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides. recalled?
Yes. FDA published recall Z-1072-2016 on March 16, 2016. Reported hazard: Increased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chemistry Systems when using Calibrator Kit 9, Lot 954. A trend of complaints regarding customers actions following U90-382 or 6LU condition codes was noted (See RES 72289 - VITROS Calibrator Kit 9, lot 954 recall).
What should I do if I have this product?
Voluntary: Firm initiated · Letter