The Plum A+ Infusion System is designed to meet the fluid delivery requirements of today's evolving healthcare environments. It is a cassette based multi-function infusion system. The Plum A+ allows two lines in and one line out. The pump can be used for standard, piggyback, or concurrent delivery. Therapy modes include: Standard Infusions, Multistep Programming, Loading Dose, and Dose Calculation. The Plum A+ is designed to deliver parenteral, enteral, or epidural infusions over a broad range of infusion rates from multiple fluid container types. Each system includes a pumping module and an assortment of disposable IV sets, optional accessories, and an operator's manual.
Published March 4, 2015 · Recall initiated: July 2, 2014
Summary
FDA records medical device Class I recall Z-1073-2015, published March 4, 2015 and initiated July 2, 2014. Recalling firm: Hospira Inc.; Status: Terminated; Reported quantity: 4,656 infusion pumps; Lot: 1) List Number: 20678: Serial Numbers: 15867001 , 15938721 , 15900791 , 15900971 , 15901051 , 15901101 , 15901301 , 15902021 , 15886121 , 15886191 , 15886201 , 15886331 , 15887441 , 15887451 , 15887501 , 15887541 , 15887551 , 15887561 , 15887571 , 15887591 , 15887601 , 15887621 , 15887721 , 15887781 , 15887801 , 15887811 , 15887841 , 15887851 , 15887861 , 15909721 , 15910521 , 15910551 , 15911281 , 15911441 , 15911501 , 15911631 , 15922951 , 15926291 , 15926731 , 15926961 , 15927281 , 15927621 , 15927731 , 15927781 , 15928041 , 15928061 , 15928131 , 15928171 , 15928221 , 15928231 , 15928291 , 15928951 , 15929051 , 15929151 , 15929351 , 15929531 , 15929901 , 15930141 , 15930191 , 15930341 , 15930351 , 15930641 , 15930691 , 15930711 , 15931171 , 15933301 , 15933501 , 15933821 , 15933941 , 15933991 , 15934001 , 15934171 , 15934261 , 15934681 , 15934711 , 15934841 , 15934891 , 15935011 , 15935071 , 15935141 , 15935191 , 15935221 , 15935261 , 15935361 , 15935381 , 15935411 , 15935431 , 15935451 , 15935471 , 15935521 , 15935541 , 15935551 , 15935561 , 15935601 , 15935891 , 15936011 , 15936181 , 15936401 , 15937241 , 15937311 , 15937381 , 15937451 , 15937461 , 15938101 , 15938601 , 15938621 , 15938731 , 15938741 , 15938751 , 15938761 , 15938871 , 15941441 , 15943391 , 15944341 , 15944421 , 15944721 , 15944751 , 15944861 , 15944881 , 15945021 , 15945091 , 15945121 , 15945381 , 15945531 , 15945921 , 15947671 , 15958991 , 15541871 , 15861851 , 15862021 , 15864801 , 15871841 , 15892651 , 15892671 , 15896621 , 15903101 , 15904551 , 15904961 , 15905391 , 15905491 , 15907531 , 15907591 , 15907751 , 15907791 , 15907801 , 15908091 , 15908471 , 15909671 , 15910131 , 15910371 , 15911701 , 15914421 , 15917421 , 15917721 , 15918431 , 15918731 , 15918951 , 15919281 , 15920761 , 15922001 , 15922761 , 15922801 , 15923281 , 15923551 , 15924971 , 15924991 , 15925081 , 15925701 , 15925731 , 15925741 , 15925791 , 15926141 , 15926191 , 15926271 , 15926281 , 15926301 , 15928071 , 15929601 , 15929621 , 15930171 , 15930631 , 15932561 , 15933451 , 15933561 , 15933721 , 15933811 , 15933831 , 15934781 , 15935321 , 15935761 , 15936121 , 15936361 , 15936561 , 15936601 , 15936951 , 15937031 , 15937041 , 15937061 , 15937191 , 15938121 , 15938471 , 15938481 , 15943001 , 15944541 , 15952631 , 15954621 , 15956511 , 15957721 , 15958091 , 15958161 , 15958241 , 15958511 , 15958691 , 15958941 , 15960891 , 15960971 , 19240471 , 19240541 , 19240771 , 19240801 , 19240861 , 19241171 , 19241441 , 15945821 , 15946021 , 15951911 , 19241901 , 19242221 , 19250221 , 19414401 , 19414451 , 19414521 , 19414571 , 19414581 , 19414601 , 19414631 , 19414641 , 19414651 , 19414661 , 19414691 , 19414711 , 19414721 , 19414751 , 19414781 , 19414811 , 19414821 , 19414841 , 19414881 , 19414901 , 19414921 , 19414981 , 19415011 , 19415041 , 19415061 , 19415071 , 19415081 , 19415091 , 19415101 , 19415111 , 19415131 , 19415191 , 19415241 , 19415251 , 19415261 , 19415321 , 19415351 , 19415361 , 19415391 , 19415421 , 19415461 , 19415491 , 19415511 , 19415521 , 19415551 , 19415581 , 19415701 , 19415731 , 19415741 , 19415751 , 19415801 , 19415831 , 19415851 , 19415891 , 19415951 , 19416331 , 19416341 , 19416361 , 19416371 , 19416381 , 19416401 , 19416411 , 19416441 , 19416451 , 19416461 , 19416471 , 19416491 , 19416521 , 19416551 , 19416571 , 19416591 , 19416611 , 19416621 , 19416661 , 19416681 , 19416751 , 19416771 , 19416781 , 19416801 , 19416811 , 19416821 , 19416831 , 19416911 , 19416941 , 19416961 , 19417061 , 19417181 , 19617581 , 15546971 , 15946081 , 15946181 , 15948941 , 15954341 , 19241451 , 19241471 , 19241731 , 19241801 , 19241891 , 19241921 , 19241971 , 19242061 , 1. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
One lot of alarm assemblies used in Plum A+ and Plum A+3 infusion pumps may fail to sound at all volume levels.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Hospira Inc.
- Distributor
- Worldwide Distribution - USA including AK, AL, AR, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and Brazil, Canada, Colombia, Ecuador, France, Israel, Italy, Mexico, Oman, Peru, Saudi Arabia, Spain, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Lot
- 1) List Number: 20678: Serial Numbers: 15867001 , 15938721 , 15900791 , 15900971 , 15901051 , 15901101 , 15901301 , 15902021 , 15886121 , 15886191 , 15886201 , 15886331 , 15887441 , 15887451 , 15887501 , 15887541 , 15887551 , 15887561 , 15887571 , 15887591 , 15887601 , 15887621 , 15887721 , 15887781 , 15887801 , 15887811 , 15887841 , 15887851 , 15887861 , 15909721 , 15910521 , 15910551 , 15911281 , 15911441 , 15911501 , 15911631 , 15922951 , 15926291 , 15926731 , 15926961 , 15927281 , 15927621 , 15927731 , 15927781 , 15928041 , 15928061 , 15928131 , 15928171 , 15928221 , 15928231 , 15928291 , 15928951 , 15929051 , 15929151 , 15929351 , 15929531 , 15929901 , 15930141 , 15930191 , 15930341 , 15930351 , 15930641 , 15930691 , 15930711 , 15931171 , 15933301 , 15933501 , 15933821 , 15933941 , 15933991 , 15934001 , 15934171 , 15934261 , 15934681 , 15934711 , 15934841 , 15934891 , 15935011 , 15935071 , 15935141 , 15935191 , 15935221 , 15935261 , 15935361 , 15935381 , 15935411 , 15935431 , 15935451 , 15935471 , 15935521 , 15935541 , 15935551 , 15935561 , 15935601 , 15935891 , 15936011 , 15936181 , 15936401 , 15937241 , 15937311 , 15937381 , 15937451 , 15937461 , 15938101 , 15938601 , 15938621 , 15938731 , 15938741 , 15938751 , 15938761 , 15938871 , 15941441 , 15943391 , 15944341 , 15944421 , 15944721 , 15944751 , 15944861 , 15944881 , 15945021 , 15945091 , 15945121 , 15945381 , 15945531 , 15945921 , 15947671 , 15958991 , 15541871 , 15861851 , 15862021 , 15864801 , 15871841 , 15892651 , 15892671 , 15896621 , 15903101 , 15904551 , 15904961 , 15905391 , 15905491 , 15907531 , 15907591 , 15907751 , 15907791 , 15907801 , 15908091 , 15908471 , 15909671 , 15910131 , 15910371 , 15911701 , 15914421 , 15917421 , 15917721 , 15918431 , 15918731 , 15918951 , 15919281 , 15920761 , 15922001 , 15922761 , 15922801 , 15923281 , 15923551 , 15924971 , 15924991 , 15925081 , 15925701 , 15925731 , 15925741 , 15925791 , 15926141 , 15926191 , 15926271 , 15926281 , 15926301 , 15928071 , 15929601 , 15929621 , 15930171 , 15930631 , 15932561 , 15933451 , 15933561 , 15933721 , 15933811 , 15933831 , 15934781 , 15935321 , 15935761 , 15936121 , 15936361 , 15936561 , 15936601 , 15936951 , 15937031 , 15937041 , 15937061 , 15937191 , 15938121 , 15938471 , 15938481 , 15943001 , 15944541 , 15952631 , 15954621 , 15956511 , 15957721 , 15958091 , 15958161 , 15958241 , 15958511 , 15958691 , 15958941 , 15960891 , 15960971 , 19240471 , 19240541 , 19240771 , 19240801 , 19240861 , 19241171 , 19241441 , 15945821 , 15946021 , 15951911 , 19241901 , 19242221 , 19250221 , 19414401 , 19414451 , 19414521 , 19414571 , 19414581 , 19414601 , 19414631 , 19414641 , 19414651 , 19414661 , 19414691 , 19414711 , 19414721 , 19414751 , 19414781 , 19414811 , 19414821 , 19414841 , 19414881 , 19414901 , 19414921 , 19414981 , 19415011 , 19415041 , 19415061 , 19415071 , 19415081 , 19415091 , 19415101 , 19415111 , 19415131 , 19415191 , 19415241 , 19415251 , 19415261 , 19415321 , 19415351 , 19415361 , 19415391 , 19415421 , 19415461 , 19415491 , 19415511 , 19415521 , 19415551 , 19415581 , 19415701 , 19415731 , 19415741 , 19415751 , 19415801 , 19415831 , 19415851 , 19415891 , 19415951 , 19416331 , 19416341 , 19416361 , 19416371 , 19416381 , 19416401 , 19416411 , 19416441 , 19416451 , 19416461 , 19416471 , 19416491 , 19416521 , 19416551 , 19416571 , 19416591 , 19416611 , 19416621 , 19416661 , 19416681 , 19416751 , 19416771 , 19416781 , 19416801 , 19416811 , 19416821 , 19416831 , 19416911 , 19416941 , 19416961 , 19417061 , 19417181 , 19617581 , 15546971 , 15946081 , 15946181 , 15948941 , 15954341 , 19241451 , 19241471 , 19241731 , 19241801 , 19241891 , 19241921 , 19241971 , 19242061 , 1
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAInfant 25% DEXTROSE Injection, USP, 2.5 g (250 mg/mL), 10 mL Unit of Use Single-dose Syringe per carton, RxMay 31, 2017
- FDAMinibore Extension Set, 59 Inch with Spin Lock Collar, Non-DEHP, Latex-Free Intended use: For theJanuary 18, 2017
- FDAFentanyl Citrate Injection, USP, 100 mcg Fentanyl/ 2mL, Single Dose, 50 mcg/mL, Rx only, Hospira, Inc., LakeNovember 30, 2016
- FDASymbiq Two Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids. List numberNovember 16, 2016
- FDAQueso Fresco, Latin Style Soft Cheese, packaged in plastic wrap, Net Wt. 1 lb. or 3 lbs. The UPC of 1 lb.March 4, 2015
- FDAHeartWare Ventricular Assist System (HeartWare Controller) Product Usage: For use as a bridge-to-cardiacMarch 4, 2015
Questions about this recall
Was The Plum A+ Infusion System is designed to meet the fluid delivery requirements of today's evolving healthcare environments. It is a cassette based multi-function infusion system. The Plum A+ allows two lines in and one line out. The pump can be used for standard, piggyback, or concurrent delivery. Therapy modes include: Standard Infusions, Multistep Programming, Loading Dose, and Dose Calculation. The Plum A+ is designed to deliver parenteral, enteral, or epidural infusions over a broad range of infusion rates from multiple fluid container types. Each system includes a pumping module and an assortment of disposable IV sets, optional accessories, and an operator's manual. recalled?
Yes. FDA published recall Z-1073-2015 on March 4, 2015. Reported hazard: One lot of alarm assemblies used in Plum A+ and Plum A+3 infusion pumps may fail to sound at all volume levels.
What should I do if I have this product?
Voluntary: Firm initiated · Letter