MicroScan WalkAway-40 plus Instrument REF B1018-283
Published May 25, 2022 · Recall initiated: October 21, 2019
Summary
FDA records medical device Class II recall Z-1083-2022, published May 25, 2022 and initiated October 21, 2019. Recalling firm: Beckman Coulter Inc.; Status: Ongoing; Reported quantity: 22 systems; Lot: Catalog Number: B1018-283 UDI-DI Code: 1509959065866 Serial Numbers: 34010003 34010004 34010005 34010006 34010007 34010008 34010009 34010010 34010011 34010012 34010013 34010014 34010015 34010016 34010017 34010018 34010019 34010020 34010021 34010022 34010023 34010024. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Beckman Coulter Inc.
- Distributor
- Worldwide distribution - U.S. Nationwide distribution in the states of Alaska, Arkansas, California, Colorado, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Missouri, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Pennsylvania, Tennessee, Texas, Utah, Vermont, Virginia, and Washington. The countries of Belgium, Brazil, Colombia, Germany, Hong Kong, Indonesia, Italy, Japan, Paraguay, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Spain, Switzerland, Taiwan, United Arab Emirates, Vietnam and Yemen.
- Lot
- Catalog Number: B1018-283 UDI-DI Code: 1509959065866 Serial Numbers: 34010003 34010004 34010005 34010006 34010007 34010008 34010009 34010010 34010011 34010012 34010013 34010014 34010015 34010016 34010017 34010018 34010019 34010020 34010021 34010022 34010023 34010024
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520October 14, 2026
- FDADxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AUOctober 14, 2026
- FDABeckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU ModuleAugust 26, 2026
- FDABeckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE,August 26, 2026
- FDAChocolate Marshmallow Premium Ice Cream, 48 ounce, Paper Carton and Paper Lid, 6 units per shrink-wrappedMay 25, 2022
- FDADxM 1096 MicroScan WalkAway Instrument REF B1018-496May 25, 2022
Questions about this recall
Was MicroScan WalkAway-40 plus Instrument REF B1018-283 recalled?
Yes. FDA published recall Z-1083-2022 on May 25, 2022. Reported hazard: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
What should I do if I have this product?
Voluntary: Firm initiated · Letter