Back to search
FDAFDA · Food and drugs·Recall number Z-1093-2017

Tryptic Soy Agar, Acumedia PN 7100 500g, 2Kg, 10Kg, and 50Kg sizes

Published February 1, 2017 · Recall initiated: April 22, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

light precipitate may form in prepared plates of Acumedia Tryptic Soy Agar (7100) Lot# 108166. While there is no change in growth performance of this medium, the formation of particulate matter may be misinterpreted as growth in poured plates thereby leading to inaccurate counts.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Acumedia Manufacturers, Inc.
Distributor
Worldwide Distribution - US Distribution to the states of : AR, CA, CT, FL, GA, ID, IL, IN, MD, ME, MI, NC, ND, NY, OH, OR, TX, TN, UT, VA, WI. and to the countries of : CANADA, UK, ISRAEL, MEXICO, AUSTRALIA, COLOMBIA, TAIWAN, BRAZIL, CHILE and INDONESIA.
Lot
7100A Lot 108166C 7100B, Lot 108166B 7100C, Lot 108166A 7100D, Lot 108166A 7100CAR, lot 108166B 108-1, Lot 108166B
Risk classification
Class II
Area (FDA)
device
Country
United States