Invivo Corporation Expression IP5 Information Portal Product Usage: The Expression Information Portal (Model IP5) a display and device controller for Philips MRI patient monitoring systems, is intended for use by trained health care professional to remotely monitor the vital signs of patients undergoing MRI procedures. Available monitoring from the IP5 can include ECG, SpO2, non-invasive and invasive blood pressure, CO2, respiration, anesthetic agents, gases, and temperature parameters.
Published March 5, 2014 · Recall initiated: October 3, 2013
Summary
FDA records medical device Class II recall Z-1096-2014, published March 5, 2014 and initiated October 3, 2013. Recalling firm: Invivo Corporation; Status: Terminated; Reported quantity: 271; Lot: US23200113, US23200120, US23200121, US23200122, US23200123, US23200124, US23200125, US23200126, US23200127, US23200128, US23200129, US23200130, US23200131, US23200132, US23200133, US23200134, US23200135, US23200136, US23200137, US23200138, US23200139, US23200140, US23200141, US23200142, US23200143, US23200144, US23200145, US23200146, US23200147, US23200148, US23200149, US23200150, US23200151, US23200152, US23200153, US23200154, US23200155, US23200156, US23200157, US23200158, US23200159, US23200160, US23200161, US23200162, US23200163, US23200164, US23200165, US23200166, US23200167, US23200168, US23200169, US23200170, US23200172, US23200173, US23200176, US23200177, US23200178, US23200179, US23200180, US23200181, US23200182, US23200183, US23200184, US23200185, US23200186, US23200187, US23200188, US23200189, US23200190, US23200191, US23200192, US23200193, US23200194, US23200195, US23200196, US23200197, US23200198, US23200199, US23200200, US23200201, US23200202, US23200203, US23200204, US23200205, US23200206, US23200207, US23200208, US23200209, US23200210, US23200211, US23200212, US23200213, US23200214, US23200215, US23200216, US23200217, US23200218, US23200219, US23200220, US23200221, US23200222, US23200223, US23200224, US23200225, US23200226, US23200227, US23200228, US23200229, US23200230, US23200231, US23200232, US23200233, US23200234, US23200235, US23200236, US23200237, US23200238, US23200239, US23200240, US23200241, US23200242, US23200243, US23200244, US23200245, US23200246, US23200247, US23200248, US23200249, US23200250, US23200251, US23200252, US23200253, US23200254, US23200255, US23200256, US23200258, US23200259, US23200260, US23200261, US23200262, US23200263, US23200264, US23200265, US23200266, US23200267, US23200268, US23200269, US23200270, US23200271, US23200272, US23200273, US23200274, US23200275, US23200276, US23200277, US23200278, US23200279, US23200280, US23200281, US23200282, US23200283, US23200284, US23200285, US23200286, US23200287, US23200288, US23200289, US23200290, US23200291, US23200292, US23200293, US23200294, US23200295, US23200296, US23200297, US23200298, US23200299, US23200300, US23200301, US23200302, US23200303, US23200304, US23200305, US23200306, US23200307, US23200308, US23200309, US23200310, US23200311, US23200312, US23200313, US23200314, US23200315, US23200316, US23200317, US23200318, US23200319, US23200320, US23200321, US23200322, US23200323, US23200324, US23200325, US23200326, US23200327, US23200328, US23200329, US23200330, US23200331, US23200332, US23200333, US23200334, US23200335, US23200336, US23200337, US23200338, US23200339, US23200340, US23200341, US23200342, US23200343, US23200344, US23200345, US23200346, US23200347, US23200349, US23200350, US23200351, US23200352, US23200353, US23200354, US23200355, US23200356, US23200357, US23200358, US23200359, US23200360, US23200361, US23200362, US23200363, US23200364, US23200365, US23200366, US23200367, US23200368, US23200369, US23200370, US23200371, US23200372, US23200373, US23200374, US23200375, US23200376, US23200377, US23200378, US23200379, US23200380, US23200381, US23200382, US23200383, US23200385, US23200386, US23200387, US23200388, US23200389, US23200390, US23200391, US23200393, US23200394, US23200395, US23200397, US23200398, US23200399, US23200400, US23200401, US23200402, US23200403, US23200405, US23200406, US23200407, US23200408, US23200410, US23200411, US23200413, US23200414, US23200415, US23200416, US23200417, US23200419, US23200421, US23200422, US23200423, US23200424, US23200425, US23200427, US23200428, US23200429, US23200431, US23200434. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Invivo Corporation located at 12151 Research Parkway, Orlando, FL 32826 is recalling their Expression Information Portal (Model IP5). Three different IP5 devices lost audio during use.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Invivo Corporation
- Distributor
- Worldwide Distribution - USA and the countries of Australia, Austria, Denmark, France, Germany, Netherlands, Poland, Singapore, Sweden, Switzerland, United Kingdom.
- Lot
- US23200113, US23200120, US23200121, US23200122, US23200123, US23200124, US23200125, US23200126, US23200127, US23200128, US23200129, US23200130, US23200131, US23200132, US23200133, US23200134, US23200135, US23200136, US23200137, US23200138, US23200139, US23200140, US23200141, US23200142, US23200143, US23200144, US23200145, US23200146, US23200147, US23200148, US23200149, US23200150, US23200151, US23200152, US23200153, US23200154, US23200155, US23200156, US23200157, US23200158, US23200159, US23200160, US23200161, US23200162, US23200163, US23200164, US23200165, US23200166, US23200167, US23200168, US23200169, US23200170, US23200172, US23200173, US23200176, US23200177, US23200178, US23200179, US23200180, US23200181, US23200182, US23200183, US23200184, US23200185, US23200186, US23200187, US23200188, US23200189, US23200190, US23200191, US23200192, US23200193, US23200194, US23200195, US23200196, US23200197, US23200198, US23200199, US23200200, US23200201, US23200202, US23200203, US23200204, US23200205, US23200206, US23200207, US23200208, US23200209, US23200210, US23200211, US23200212, US23200213, US23200214, US23200215, US23200216, US23200217, US23200218, US23200219, US23200220, US23200221, US23200222, US23200223, US23200224, US23200225, US23200226, US23200227, US23200228, US23200229, US23200230, US23200231, US23200232, US23200233, US23200234, US23200235, US23200236, US23200237, US23200238, US23200239, US23200240, US23200241, US23200242, US23200243, US23200244, US23200245, US23200246, US23200247, US23200248, US23200249, US23200250, US23200251, US23200252, US23200253, US23200254, US23200255, US23200256, US23200258, US23200259, US23200260, US23200261, US23200262, US23200263, US23200264, US23200265, US23200266, US23200267, US23200268, US23200269, US23200270, US23200271, US23200272, US23200273, US23200274, US23200275, US23200276, US23200277, US23200278, US23200279, US23200280, US23200281, US23200282, US23200283, US23200284, US23200285, US23200286, US23200287, US23200288, US23200289, US23200290, US23200291, US23200292, US23200293, US23200294, US23200295, US23200296, US23200297, US23200298, US23200299, US23200300, US23200301, US23200302, US23200303, US23200304, US23200305, US23200306, US23200307, US23200308, US23200309, US23200310, US23200311, US23200312, US23200313, US23200314, US23200315, US23200316, US23200317, US23200318, US23200319, US23200320, US23200321, US23200322, US23200323, US23200324, US23200325, US23200326, US23200327, US23200328, US23200329, US23200330, US23200331, US23200332, US23200333, US23200334, US23200335, US23200336, US23200337, US23200338, US23200339, US23200340, US23200341, US23200342, US23200343, US23200344, US23200345, US23200346, US23200347, US23200349, US23200350, US23200351, US23200352, US23200353, US23200354, US23200355, US23200356, US23200357, US23200358, US23200359, US23200360, US23200361, US23200362, US23200363, US23200364, US23200365, US23200366, US23200367, US23200368, US23200369, US23200370, US23200371, US23200372, US23200373, US23200374, US23200375, US23200376, US23200377, US23200378, US23200379, US23200380, US23200381, US23200382, US23200383, US23200385, US23200386, US23200387, US23200388, US23200389, US23200390, US23200391, US23200393, US23200394, US23200395, US23200397, US23200398, US23200399, US23200400, US23200401, US23200402, US23200403, US23200405, US23200406, US23200407, US23200408, US23200410, US23200411, US23200413, US23200414, US23200415, US23200416, US23200417, US23200419, US23200421, US23200422, US23200423, US23200424, US23200425, US23200427, US23200428, US23200429, US23200431, US23200434
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMR Coils The MR Coil is intended to be used in conjunction with a Magnetic Resonance Scanner to produceOctober 28, 2020
- FDAMonitor for the Functional Imaging System for Magnetic Resonance Imaging SystemMarch 11, 2020
- FDAPhilips Expression Information Portal (IP5) Model 865471. The IP5 is intended to be used as a secondarySeptember 28, 2016
- FDAPhilips Expression MR200 MRI Patient Monitoring SystemOctober 22, 2014
- FDACSI, Diamondback 360 Coronary Orbital Atherectomy System, Part Number 70058-02, Model Number DBEC-125 ProductMarch 5, 2014
- FDAThe SYMBIONIC LEG is an integrated prosthetic leg for transfemoral and knee disarticulation amputees. ProductMarch 5, 2014
Questions about this recall
Was Invivo Corporation Expression IP5 Information Portal Product Usage: The Expression Information Portal (Model IP5) a display and device controller for Philips MRI patient monitoring systems, is intended for use by trained health care professional to remotely monitor the vital signs of patients undergoing MRI procedures. Available monitoring from the IP5 can include ECG, SpO2, non-invasive and invasive blood pressure, CO2, respiration, anesthetic agents, gases, and temperature parameters. recalled?
Yes. FDA published recall Z-1096-2014 on March 5, 2014. Reported hazard: Invivo Corporation located at 12151 Research Parkway, Orlando, FL 32826 is recalling their Expression Information Portal (Model IP5). Three different IP5 devices lost audio during use.
What should I do if I have this product?
Voluntary: Firm initiated · Letter