FDAFDA · Food and drugs·Recall number Z-1099-2017

Accu-Chek Connect Diabetes Management App

Published February 1, 2017 · Recall initiated: December 20, 2016

Summary

FDA records medical device Class II recall Z-1099-2017, published February 1, 2017 and initiated December 20, 2016. Recalling firm: Roche Diabetes Care, Inc.; Status: Terminated; Reported quantity: 27243; Lot: Catalog number 07562462001 / GTIN number 00365702700000 Catalog number 07250452001 / GTIN number 00365702700017. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

iOS and Android: Under certain conditions the affected app versions may disregard historical bolus data potentially leading to an incorrect bolus insulin recommendation being provided to the user. iOS only: Pairing and using multiple meters with the Accu-Chek Connect app can under rare circumstances cause the bolus advisor to fail to offer a correction bolus recommendation within the eligible time window following a blood glucose measurement (10  15 minutes). Depending on the individual metabolic situation potentially incorrect bolus advice could lead to serious health consequences such as hypoglycemia. Both software issues may also cause the amount of active insulin displayed during the bolus calculation process to be incorrect and should not be used to manually calculate a bolus.

What to do

Voluntary: Firm initiated · Other

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Roche Diabetes Care, Inc.
Distributor
Worldwide Distribution - Nationwide Distribution The Accu-Chek Connect App for iOS is distributed directly to end users via the Apple App Store. The Accu-Chek Connect App for Android is distributed directly to end users via the Google Play App Store.
Lot
Catalog number 07562462001 / GTIN number 00365702700000 Catalog number 07250452001 / GTIN number 00365702700017
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Accu-Chek Connect Diabetes Management App recalled?

Yes. FDA published recall Z-1099-2017 on February 1, 2017. Reported hazard: iOS and Android: Under certain conditions the affected app versions may disregard historical bolus data potentially leading to an incorrect bolus insulin recommendation being provided to the user. iOS only: Pairing and using multiple meters with the Accu-Chek Connect app can under rare circumstances cause the bolus advisor to fail to offer a correction bolus recommendation within the eligible time window following a blood glucose measurement (10  15 minutes). Depending on the individual metabolic situation potentially incorrect bolus advice could lead to serious health consequences such as hypoglycemia. Both software issues may also cause the amount of active insulin displayed during the bolus calculation process to be incorrect and should not be used to manually calculate a bolus.