FDA records medical device Class II recall Z-1106-2017, published February 1, 2017 and initiated December 27, 2016. Recalling firm: Philips Electronics North America Corporation; Status: Terminated; Reported quantity: 107; Lot: InfiniBand cables shipped between June 22, 2016 and October 6, 2016 in use with the Cisco 3850 switch. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
May experience a loss of centralized monitoring.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Electronics North America Corporation
- Distributor
- Worldwide Distribution - US Distribution to the states of : AK CA CO FL GA KY NC NE NH NJ NY OH PA SC TN TX VA WA and WV., and to the countries of : Canada, Germany, Japan, New Zealand, Poland and United Kingdom
- Lot
- InfiniBand cables shipped between June 22, 2016 and October 6, 2016 in use with the Cisco 3850 switch.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAINTEGRIS CV, System Code 722030 Vascular, cardiovascular and neurovascular imaging applications, CardiacOctober 3, 2018
- FDAAllura Xper FD2O, System Code 722006 Vascular, cardiovascular and neurovascular imaging applications, CardiacOctober 3, 2018
- FDAAIIuraXperFDlO/10, System Code 722011 Vascular, cardiovascular and neurovascular imaging applications, CardiacOctober 3, 2018
- FDAAIIura Xper F010 DR Table, System Code 722022 Vascular, cardiovascular and neurovascular imaging applications,October 3, 2018
- FDAFlucanozole Tablets, USP, 50 mg, a) 30 count (NDC 55111-143-30) and b) 100 count (NDC 55111-143-01) bottles,February 8, 2017
- FDAAccu-Chek Connect Diabetes Management AppFebruary 1, 2017
Questions about this recall
Was Philips Cisco 3850 layer 3 switch, 865339 model WS-C3850-12S-E in use with IntelliVue Information Center iX or Classic Information Center recalled?
Yes. FDA published recall Z-1106-2017 on February 1, 2017. Reported hazard: May experience a loss of centralized monitoring.
What should I do if I have this product?
Voluntary: Firm initiated · Letter