Intellispace Portal (ISP) Philips Medical Systems (Cleveland), Inc., IntelliSpace Portal (ISP) is a multimodality (CT, NM, and MR) thin-client applications server that delivers full diagnostic viewing and clinical applications to the enterprise. IntelliSpace Portal (ISP) is a medical software system that allows multiple users to remotely access IntelliSpace Workspace Portal (ISP) from compatible computers on a network. The system allows networking, selection, processing and filming of multimodality DICOM images. Both the client and server software are only for use with off the shelf hardware technology that meets defined minimum specifications. trained physicians may use the images as a basis for diagnosis upon ensuring that monitor quality, ambient light conditions and image compression ratios are consistent with clinical application.
Published April 24, 2013 · Recall initiated: February 12, 2013
Summary
FDA records medical device Class II recall Z-1102-2013, published April 24, 2013 and initiated February 12, 2013. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 313 units; Lot: Model #881001, Serial #s: 79916, 79932, 79933, 79935, 79936, 79937, 79939, 79940, 79943, 80071, 80914, 80916, 80925, 80926, 80928, 80932, 80939, 80963, 80965, 80977, 80979, 80994, 80995, 81004, 81006, 81008, 81010, 81026, 81027, 81028, 81034, 81035, 81036, 81039, 81063, 81065, 81076, 81079, 81081, 81085, 81089, 81097, 81098, 81108, 81110, 81117, 81119, 81127, 81130, 81132, 81133, 81150, 81153, 81154, 81160, 81161, 81163, 81165, 81169, 81172, 81173, 81174, 81175, 81178, 81179, 81182, 81183, 81184, 81187, 81189, 81190, 81192, 81194, 81196, 81200, 82001, 82002, 82003, 82004, 82006, 82007, 82008, 82009, 82010, 82011, 82012, 82013, 82014, 82015, 82016, 82017, 82018, 82019, 82020, 82021, 82022, 82023, 82024, 82025, 82026, 82027, 82028, 82029, 82030, 82031, 82032, 82033, 82034, 82035, 82036, 82037, 82038, 82050, 82051, 82052, 82053, 82054, 82055, 82056, 82057, 82058, 82059, 82060, 82061, 82062, 82063, 82064, 82065, 82066, 82067, 82068, 82069, 82070, 82071, 82072, 82073, 82074, 82084, 82086, 82087, 82088, 82089, 82090, 82091, 82092, 82095, 82097, 82098, 82099, 82100, 82101, 82102, 82103, 82104, 82105, 82106, 82107, 82108, 82109, 82110, 82112, 82116, 82118, 82120, 82121, 82123, 82125, 82126, 82128, 82129, 82130, 82131, 82132, 82133, 82138, 82142, 82143, 82144, 82146, 82148, 82201, 82202, 82203, 82236, 82238, 82240, 82241, 82250, 82251, 85005, 85006, 85007, 85008, 85009, 85010, 85011, 85014, 85025, 85026, 85027, 85028, 85029, 85031, 85032, 85033, 85040, 85041, 85042, 85043, 85044, 85046, 85047, 85049, 85050, 85066, 85067, 85068, 85069, 85080, 85081, 85082, 85083, 85084, 85086, 85088, 85090, 85096, 85103, 85104, 85105, 85106, 85107, 85108, 85110, 85117, 85141, 85143, 85144, 85145, 85149, 85164, 85165, 85167, 85168, 85169, 85170, 85171, 85172, 85174, 85175, 85176, 85177, 85178, 85179, 85180, 85213, 85214, 85215, 85216, 85217, 85218, 85219, 85238, 85239, 85243, 85244, 85245, 85246, 85247, 85248, 85249, 85250, 85252, 85253, 85254, 85255, 85257, 85260, 85296, 85299, 85300, 85302, 85305, 85306, 85307, 85308, 85309, 85312, 85315, 85316, 85317, 85319, 85320, 85322, 85329, 85330, 85332, 85333, 85334, 85335, 85337, 85338, 85355, 85373, 85475, 95001, 95007, 95008, 95009, 820001, 822002, 822003, 822004, 822005, 822009, 822017, 850004, 850015, 920003 & 950001. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Philips was notified of a problem using the IntelliSpace Portal with software version 4.0.2 with the URL PACS integration. Philips was informed that when loading a study to an analysis application while using URL PACS integration configuration, after another study was opened by a review application, the Analysis monitor is updated with the study of the current patient, but the Review monitor conti
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Distributor
- Worldwide Distribution - (USA) Nationwide Distribution including the states of AL, AR, AZ, CA, FL, GA, IA, IN, LA, MI, MN, MO, NC, NJ, NY, OH, OK, OR, PA, TN, TX, WA, WI & WV. and the countries of Australia, Austria, Belgium, Canada, China, Czech Republic, Denmark, Egypt, Finland, France, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Latvia, Lebanon, Monaco, Netherlands, Norway, Poland, Portugal, Russia, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand & United Kingdom
- Lot
- Model #881001, Serial #s: 79916, 79932, 79933, 79935, 79936, 79937, 79939, 79940, 79943, 80071, 80914, 80916, 80925, 80926, 80928, 80932, 80939, 80963, 80965, 80977, 80979, 80994, 80995, 81004, 81006, 81008, 81010, 81026, 81027, 81028, 81034, 81035, 81036, 81039, 81063, 81065, 81076, 81079, 81081, 81085, 81089, 81097, 81098, 81108, 81110, 81117, 81119, 81127, 81130, 81132, 81133, 81150, 81153, 81154, 81160, 81161, 81163, 81165, 81169, 81172, 81173, 81174, 81175, 81178, 81179, 81182, 81183, 81184, 81187, 81189, 81190, 81192, 81194, 81196, 81200, 82001, 82002, 82003, 82004, 82006, 82007, 82008, 82009, 82010, 82011, 82012, 82013, 82014, 82015, 82016, 82017, 82018, 82019, 82020, 82021, 82022, 82023, 82024, 82025, 82026, 82027, 82028, 82029, 82030, 82031, 82032, 82033, 82034, 82035, 82036, 82037, 82038, 82050, 82051, 82052, 82053, 82054, 82055, 82056, 82057, 82058, 82059, 82060, 82061, 82062, 82063, 82064, 82065, 82066, 82067, 82068, 82069, 82070, 82071, 82072, 82073, 82074, 82084, 82086, 82087, 82088, 82089, 82090, 82091, 82092, 82095, 82097, 82098, 82099, 82100, 82101, 82102, 82103, 82104, 82105, 82106, 82107, 82108, 82109, 82110, 82112, 82116, 82118, 82120, 82121, 82123, 82125, 82126, 82128, 82129, 82130, 82131, 82132, 82133, 82138, 82142, 82143, 82144, 82146, 82148, 82201, 82202, 82203, 82236, 82238, 82240, 82241, 82250, 82251, 85005, 85006, 85007, 85008, 85009, 85010, 85011, 85014, 85025, 85026, 85027, 85028, 85029, 85031, 85032, 85033, 85040, 85041, 85042, 85043, 85044, 85046, 85047, 85049, 85050, 85066, 85067, 85068, 85069, 85080, 85081, 85082, 85083, 85084, 85086, 85088, 85090, 85096, 85103, 85104, 85105, 85106, 85107, 85108, 85110, 85117, 85141, 85143, 85144, 85145, 85149, 85164, 85165, 85167, 85168, 85169, 85170, 85171, 85172, 85174, 85175, 85176, 85177, 85178, 85179, 85180, 85213, 85214, 85215, 85216, 85217, 85218, 85219, 85238, 85239, 85243, 85244, 85245, 85246, 85247, 85248, 85249, 85250, 85252, 85253, 85254, 85255, 85257, 85260, 85296, 85299, 85300, 85302, 85305, 85306, 85307, 85308, 85309, 85312, 85315, 85316, 85317, 85319, 85320, 85322, 85329, 85330, 85332, 85333, 85334, 85335, 85337, 85338, 85355, 85373, 85475, 95001, 95007, 95008, 95009, 820001, 822002, 822003, 822004, 822005, 822009, 822017, 850004, 850015, 920003 & 950001.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.September 21, 2022
- FDABrilliance 64 Model # 728231, computed tomography x-ray systemAugust 14, 2019
- FDAIngenuity Core128 Model # 728323, computed tomography x-ray systemAugust 14, 2019
- FDABrightView XCT , Model Number 882482July 31, 2019
- FDAAlaris PC unit model 8015 with software version 9.12 Product Usage: The Alaris PC unit is the centralApril 24, 2013
- FDAIntegra XKnife® Stereotactic Radiosurgery and Radiotherapy Treatment Planning, Software Versions 5.0.1 andApril 24, 2013
Questions about this recall
Was Intellispace Portal (ISP) Philips Medical Systems (Cleveland), Inc., IntelliSpace Portal (ISP) is a multimodality (CT, NM, and MR) thin-client applications server that delivers full diagnostic viewing and clinical applications to the enterprise. IntelliSpace Portal (ISP) is a medical software system that allows multiple users to remotely access IntelliSpace Workspace Portal (ISP) from compatible computers on a network. The system allows networking, selection, processing and filming of multimodality DICOM images. Both the client and server software are only for use with off the shelf hardware technology that meets defined minimum specifications. trained physicians may use the images as a basis for diagnosis upon ensuring that monitor quality, ambient light conditions and image compression ratios are consistent with clinical application. recalled?
Yes. FDA published recall Z-1102-2013 on April 24, 2013. Reported hazard: Philips was notified of a problem using the IntelliSpace Portal with software version 4.0.2 with the URL PACS integration. Philips was informed that when loading a study to an analysis application while using URL PACS integration configuration, after another study was opened by a review application, the Analysis monitor is updated with the study of the current patient, but the Review monitor conti
What should I do if I have this product?
Voluntary: Firm initiated · Letter