VITROS 5600 Integrated System, Product Code 6802413. Intended for use in the measurement of a variety of chemistry analytes.
Published April 24, 2013 · Recall initiated: February 7, 2013
Summary
FDA records medical device Class II recall Z-1108-2013, published April 24, 2013 and initiated February 7, 2013. Recalling firm: Ortho-Clinical Diagnostics; Status: Terminated; Reported quantity: USA: 775 units, Foreign: 800 units; Lot: Serial Numbers 56000118 through 56001583. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Ortho-Clinical Diagnostics, Inc. (OCD) determined that extension springs on some VITROS® Systems may be out of its manufacturing specifications. Extension springs are part of the slide alignment guide assembly. During an internal evaluation in its manufacturing facility, it was discovered that an extension spring had insufficient force. If the extension spring does not provide sufficient force, MicroSlides may not be properly positioned within the MicroSlide incubator which has the potential to cause imprecise results. Extension springs are just one component in the slide alignment assembly. OCD's evaluation determined that the probability of an out of specification extension spring alone adversely affecting results is unlikely.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ortho-Clinical Diagnostics
- Distributor
- Worldwide Distribution- USA (nationwide) including Puerto Rico and countries of: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Republic of Panama, Singapore, Spain, and Venezuela.
- Lot
- Serial Numbers 56000118 through 56001583
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVITROS Performance Verifier IIApril 3, 2024
- FDAVITROS Performance Verifier IApril 3, 2024
- FDAVITROS XT Chemistry Products TBIL-ALKP Slides (ALKP is not affected) In Vitro Diagnostic use only - The TBILSeptember 2, 2020
- FDAVITROS Chemistry Product Performance Verifier I, Product Code 8231474, UDI # 10758750004577 - Product Usage:June 24, 2020
- FDAVITROS 4600 Chemistry System (VITROS 5,1 FS System family member), Product Code 6802445. Intended for use inApril 24, 2013
- FDAIS2000 USER MANUAL for Intuitive Surgical da Vinci S Surgical System, Model IS2000. Arm drapes protectApril 24, 2013
Questions about this recall
Was VITROS 5600 Integrated System, Product Code 6802413. Intended for use in the measurement of a variety of chemistry analytes. recalled?
Yes. FDA published recall Z-1108-2013 on April 24, 2013. Reported hazard: Ortho-Clinical Diagnostics, Inc. (OCD) determined that extension springs on some VITROS® Systems may be out of its manufacturing specifications. Extension springs are part of the slide alignment guide assembly. During an internal evaluation in its manufacturing facility, it was discovered that an extension spring had insufficient force. If the extension spring does not provide sufficient force, MicroSlides may not be properly positioned within the MicroSlide incubator which has the potential to cause imprecise results. Extension springs are just one component in the slide alignment assembly. OCD's evaluation determined that the probability of an out of specification extension spring alone adversely affecting results is unlikely.
What should I do if I have this product?
Voluntary: Firm initiated · Letter