ACCU-CHEK Inform II Blood Glucose Monitoring System bar code scanner. Used with the ACCU-CHEK Inform II System to quantitatively measure glucose (sugar).

Published April 24, 2013 · Recall initiated: March 6, 2013

Summary

FDA records medical device Class II recall Z-1113-2013, published April 24, 2013 and initiated March 6, 2013. Recalling firm: Roche Diagnostics Operations, Inc.; Status: Terminated; Reported quantity: Both devices 4467; Lot: part number 05060311001, 05060303001. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Roche Diagnostics Operations, Inc.
Distributor
Nationwide Distribution.
Lot
part number 05060311001, 05060303001
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ACCU-CHEK Inform II Blood Glucose Monitoring System bar code scanner. Used with the ACCU-CHEK Inform II System to quantitatively measure glucose (sugar). recalled?

Yes. FDA published recall Z-1113-2013 on April 24, 2013. Reported hazard: Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards.