TSX-101A Aquilion 32/64 ONE Computed Tomography X-Ray System; TXS-101/R Aquilion RXL Computed Tomography X-Ray System; TSX-301A Aquilion One Computed Tomography X-Ray System; TSX-301B Aquilion Premium Computed Tomography X-Ray System; TSX-302A Aquilion Prime Computed Tomography X-Ray System. This device is indicated to acquire and display cross sectional volumes of the whole body, to include the head. The Aquilion RXL has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.
Published March 12, 2014 · Recall initiated: October 1, 2013
Summary
FDA records medical device Class II recall Z-1113-2014, published March 12, 2014 and initiated October 1, 2013. Recalling firm: Toshiba American Medical Systems Inc; Status: Terminated; Reported quantity: 338 units; Lot: SERIAL NUMBER JDA1062176, JDA1062177, JDA1072183, JDA1072184, JDA1072186, JDA1072187, JDA1082188, JDA1082189, JDA1082190, JDA10X2191, JDA10X2192, JDA10X2193, JDA10X2194, JDA10X2195, JDA10X2196, JDA10X2197, JDA10X2198, JDA10Y2199, JDA10Y2200, JDA10Y2201, JDA10Y2202, JDA10Y2203, JDA10Y2204, JDA10Y2205, JDA10Z2206, JDA10Z2207, JDA10Z2209, JDA10Z2210, JDA10Z2211, JDA10Z2212, JDA10Z2213, JDA10Z2214, JDA10Z2215, JDA1112216, JDA1112217, JDA1112218, JDA1112219, JDA1112221, JDA1112222, JDA1112224, JDA1112225, JDA1112226, JDA1112227, JDA1122231, JDA1122234, JDA1142236, JDA1162237, JDA1162238, JDA1162239, JDA1182240, JDA1182241, JDA1182242, JDA1182243, JDA1182244, JDA11X2245, JDA11X2246, JDA11X2247, JDA11Z2249, JDA11Z2250, JDA11Z2252, JDA11Z2253, JDA11Z2254, JDA11Z2255, JDA1222257, JDA1222258, JDA1222259, JDA1222260, JDA1222261, JDA1242264, JDA1252265, JDA1252266, JDA1252267, JDA12Z2271, JDA1312272, JGA1062178, JGA1062181, JGA1062182, JGA1112220, JGA1112228, JGA11Z2256, JGA1222262, JGA12X2269, JLA1062179, JLA1062180, JLA1072185, JLA10Z2208, JLA1122232, JLA1122233, JLA1132235, JLA11Y2248, JLA11Z2251, JLA1232263, JLA1262268, JLA12Y2270, KDA1062080, KDA1072081, KDA1072082, KDA1072083, KDA1072085, KDA1072086, KDA1072088, KDA1082090, KDA1082091, KDA1082092, KDA1092093, KDA10X2094, KDA10X2095, KDA10X2096, KDA10X2097, KDA10X2098, KDA10X2099, KDA10Y2100, KDA10Y2101, KDA10Y2102, KDA10Z2104, KDA1112107, KDA1122110, KDA1122111, KDA1132113, KDA1142114, KDA1152115, KDA1152116, KDA1152117, KDA1152118, KDA1162119, KDA1162120, KDA1162121, KDA1162122, KDA1162123, KDA1162124, KDA1172125, KDA1182126, KDA1182127, KDA1182128, KDA1182129, KDA1192130, KDA1192131, KDA11X2132, KDA11X2133, KDA11Y2134, KDA11Z2136, KDA11Z2137, KDA11Z2138, KDA11Z2139, KDA11Z2140, KDA1222141, KDA1222142, KDA1222143, KDA1232144, KDA1242145, KDA1242146, KDA1252147, KDA1272148, KDA1282149, KDA1292151, KDA12X2152, KDA12Y2153, KDA12Z2154, KDA1322155, KGA1082089, KGA10Z2103, KGA10Z2105, KGA1112106, KGA1112108, KGA1122109, KGA1122112, KGA11Z2135, KLA1072084, KLA1072087, KLA1292150, RDA1292019, RDA12X2023, RDA12X2024, RDB12Y2029, RDB12Z2032, RDB1312037, RDB1312038, RGA12X2025, RGB12Y2028, RGB1312034, 2DA0792002, 2DA0792004, 2DA07X2005, 2DA07X2008, 2DA07Y2010, 2DA0842023, 2DA0842026, 2DA0862034, 2DA0862036, 2DA0872038, 2DA0882048, 2DA0882049, 2DA0882051, 2DA0882052, 2DA0882053, 2DA08X2061, 2DA08X2067, 2DA08X2070, 2DA08X2071, 2DA08X2072, 2DA08Y2077, 2DA08Y2080, 2DA08Z2092, 2DA0952115, 2DA0962116, 2DA0972123, 2DA0982131, 2DA0982133, 2DA09X2139, 2DA09X2143, 2DA09Y2148, 2DA09Y2151, 2DA09Z2158, 2DA09Z2161, 2DA09Z2162, 2DA1012165, 2DA1012167, 2DA1012168, 2DA1012169, 2DA1012170, 2DA1012174, 2DA1062208, 2DA1072212, 2DB1082219, 2DB1082220, 2DB1082222, 2DB10X2241, 2DB10Y2253, 2DB10Y2257, 2DB10Z2265, 2DB1112275, 2DB1112283, 2DB1122284, 2DB1122291, 2DB1122294, 2DC1142308, 2DC1142311, 2DC1152316, 2DC1162324, 2DC1172328, 2DC1172332, 2DC1172333, 2DC11X2352, 2DC11Y2365, 2DC11Y2367, 2DC11Z2372, 2DC11Z2374, 2DC1212377, 2DC1212383, 2DC1232392, 2DC1232393, 2DC1232396, 2DC1232397, 2DC1232398, 2DC1242407, 2DC1242409, 2DC1252410, 2DC1262412, 2DC1272417, 2DC1272421, 2DC1282423, 2DC12X2427, 2DC12Y2430, 2DC12Y2431, 2DC12Z2433, 2DC1312435, 2DC1312436, 2DC1322437, 1DA0962006, 1DA0972007, 1DA0972008, 1DA0992011, 1DA09Y2014, 1DA1062015, 1DB1072016, 1DB1082017, 1DB1082018, 1DB1082019, 1DB10Y2020, 1DB10Y2021, 1DB10Y2022, 1DB1112025, 1DB1122027, 1DC1142030, 1DC1152031, 1DC1162032, 1DC1162033, 1DC1172035, 1DC11Y2037, 1DC11Y2038, 1DC11Z2040, 1DC1222042, 1DC1232043, 1DC1232044, 1DC1252045, 1DC1252046, 1DC1262048, 1DC1272050, 1DC1282051, 1DC1292052, 1DC12X2053, 1DC12Y2054, 1DC12Y2055, 1DC12Z2056, 1DC1312057, 1DC1312058, 1DC1342. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Toshiba intitiated this recall because their investigation revealed that due to a problem of the X-ray output control software, the X-ray output conditions may not be set correctly. As a result, X-ray exposure under the scan conditions used for contrast study may not start.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Toshiba American Medical Systems Inc
- Distributor
- Nationwide Distribution.
- Lot
- SERIAL NUMBER JDA1062176, JDA1062177, JDA1072183, JDA1072184, JDA1072186, JDA1072187, JDA1082188, JDA1082189, JDA1082190, JDA10X2191, JDA10X2192, JDA10X2193, JDA10X2194, JDA10X2195, JDA10X2196, JDA10X2197, JDA10X2198, JDA10Y2199, JDA10Y2200, JDA10Y2201, JDA10Y2202, JDA10Y2203, JDA10Y2204, JDA10Y2205, JDA10Z2206, JDA10Z2207, JDA10Z2209, JDA10Z2210, JDA10Z2211, JDA10Z2212, JDA10Z2213, JDA10Z2214, JDA10Z2215, JDA1112216, JDA1112217, JDA1112218, JDA1112219, JDA1112221, JDA1112222, JDA1112224, JDA1112225, JDA1112226, JDA1112227, JDA1122231, JDA1122234, JDA1142236, JDA1162237, JDA1162238, JDA1162239, JDA1182240, JDA1182241, JDA1182242, JDA1182243, JDA1182244, JDA11X2245, JDA11X2246, JDA11X2247, JDA11Z2249, JDA11Z2250, JDA11Z2252, JDA11Z2253, JDA11Z2254, JDA11Z2255, JDA1222257, JDA1222258, JDA1222259, JDA1222260, JDA1222261, JDA1242264, JDA1252265, JDA1252266, JDA1252267, JDA12Z2271, JDA1312272, JGA1062178, JGA1062181, JGA1062182, JGA1112220, JGA1112228, JGA11Z2256, JGA1222262, JGA12X2269, JLA1062179, JLA1062180, JLA1072185, JLA10Z2208, JLA1122232, JLA1122233, JLA1132235, JLA11Y2248, JLA11Z2251, JLA1232263, JLA1262268, JLA12Y2270, KDA1062080, KDA1072081, KDA1072082, KDA1072083, KDA1072085, KDA1072086, KDA1072088, KDA1082090, KDA1082091, KDA1082092, KDA1092093, KDA10X2094, KDA10X2095, KDA10X2096, KDA10X2097, KDA10X2098, KDA10X2099, KDA10Y2100, KDA10Y2101, KDA10Y2102, KDA10Z2104, KDA1112107, KDA1122110, KDA1122111, KDA1132113, KDA1142114, KDA1152115, KDA1152116, KDA1152117, KDA1152118, KDA1162119, KDA1162120, KDA1162121, KDA1162122, KDA1162123, KDA1162124, KDA1172125, KDA1182126, KDA1182127, KDA1182128, KDA1182129, KDA1192130, KDA1192131, KDA11X2132, KDA11X2133, KDA11Y2134, KDA11Z2136, KDA11Z2137, KDA11Z2138, KDA11Z2139, KDA11Z2140, KDA1222141, KDA1222142, KDA1222143, KDA1232144, KDA1242145, KDA1242146, KDA1252147, KDA1272148, KDA1282149, KDA1292151, KDA12X2152, KDA12Y2153, KDA12Z2154, KDA1322155, KGA1082089, KGA10Z2103, KGA10Z2105, KGA1112106, KGA1112108, KGA1122109, KGA1122112, KGA11Z2135, KLA1072084, KLA1072087, KLA1292150, RDA1292019, RDA12X2023, RDA12X2024, RDB12Y2029, RDB12Z2032, RDB1312037, RDB1312038, RGA12X2025, RGB12Y2028, RGB1312034, 2DA0792002, 2DA0792004, 2DA07X2005, 2DA07X2008, 2DA07Y2010, 2DA0842023, 2DA0842026, 2DA0862034, 2DA0862036, 2DA0872038, 2DA0882048, 2DA0882049, 2DA0882051, 2DA0882052, 2DA0882053, 2DA08X2061, 2DA08X2067, 2DA08X2070, 2DA08X2071, 2DA08X2072, 2DA08Y2077, 2DA08Y2080, 2DA08Z2092, 2DA0952115, 2DA0962116, 2DA0972123, 2DA0982131, 2DA0982133, 2DA09X2139, 2DA09X2143, 2DA09Y2148, 2DA09Y2151, 2DA09Z2158, 2DA09Z2161, 2DA09Z2162, 2DA1012165, 2DA1012167, 2DA1012168, 2DA1012169, 2DA1012170, 2DA1012174, 2DA1062208, 2DA1072212, 2DB1082219, 2DB1082220, 2DB1082222, 2DB10X2241, 2DB10Y2253, 2DB10Y2257, 2DB10Z2265, 2DB1112275, 2DB1112283, 2DB1122284, 2DB1122291, 2DB1122294, 2DC1142308, 2DC1142311, 2DC1152316, 2DC1162324, 2DC1172328, 2DC1172332, 2DC1172333, 2DC11X2352, 2DC11Y2365, 2DC11Y2367, 2DC11Z2372, 2DC11Z2374, 2DC1212377, 2DC1212383, 2DC1232392, 2DC1232393, 2DC1232396, 2DC1232397, 2DC1232398, 2DC1242407, 2DC1242409, 2DC1252410, 2DC1262412, 2DC1272417, 2DC1272421, 2DC1282423, 2DC12X2427, 2DC12Y2430, 2DC12Y2431, 2DC12Z2433, 2DC1312435, 2DC1312436, 2DC1322437, 1DA0962006, 1DA0972007, 1DA0972008, 1DA0992011, 1DA09Y2014, 1DA1062015, 1DB1072016, 1DB1082017, 1DB1082018, 1DB1082019, 1DB10Y2020, 1DB10Y2021, 1DB10Y2022, 1DB1112025, 1DB1122027, 1DC1142030, 1DC1152031, 1DC1162032, 1DC1162033, 1DC1172035, 1DC11Y2037, 1DC11Y2038, 1DC11Z2040, 1DC1222042, 1DC1232043, 1DC1232044, 1DC1252045, 1DC1252046, 1DC1262048, 1DC1272050, 1DC1282051, 1DC1292052, 1DC12X2053, 1DC12Y2054, 1DC12Y2055, 1DC12Z2056, 1DC1312057, 1DC1312058, 1DC1342
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAINFX-8000C Fluoroscopic X-Ray SystemsMarch 14, 2018
- FDAAPLIO 500 TUS-A500 diagnostic ultrasound systemJuly 19, 2017
- FDAXARIO 200 TUS-X200 diagnostic ultrasound systemJuly 19, 2017
- FDAToshiba Kalare DREX-KL80 X-ray generatorMarch 22, 2017
- FDAAgility. Version 3.1 of the Integrity interface and control software for the Elekta range of medical digitalMarch 12, 2014
- FDAGreatbatch Medical, Offset Cup Impactor. Part Numbers: P9485, T10243, T10281, T10287, T10484, T10491, T10507,March 12, 2014
Questions about this recall
Was TSX-101A Aquilion 32/64 ONE Computed Tomography X-Ray System; TXS-101/R Aquilion RXL Computed Tomography X-Ray System; TSX-301A Aquilion One Computed Tomography X-Ray System; TSX-301B Aquilion Premium Computed Tomography X-Ray System; TSX-302A Aquilion Prime Computed Tomography X-Ray System. This device is indicated to acquire and display cross sectional volumes of the whole body, to include the head. The Aquilion RXL has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician. recalled?
Yes. FDA published recall Z-1113-2014 on March 12, 2014. Reported hazard: Toshiba intitiated this recall because their investigation revealed that due to a problem of the X-ray output control software, the X-ray output conditions may not be set correctly. As a result, X-ray exposure under the scan conditions used for contrast study may not start.
What should I do if I have this product?
Voluntary: Firm initiated · Letter