Greatbatch Medical, Offset Cup Impactor. Part Numbers: P9485, T10243, T10281, T10287, T10484, T10491, T10507, T10600, T10604, T10753, T10861, T11209, T11340, T12121, T12230, T12288, T12305, T12391, T12699, T12767, T12861, T13344, T13480, T13642, T13722, T13999, T14155, T14384, T15311, T15752, T15822, T15949, T16117, T16427, T16604, T16611, T16661, T16829, T16934, T17062, T17203, T17238, T17321, T6318, T7821, T8042, T8043, T8044, T8087, T8088, T8093, T8095, T8177, T8184, T8277, T8310, T8333, T8468, T8487, T9109, T9196, T9205, T9316, T9348, T9360, T9556, T9557, T9747, T9894, T9954, T9955, T9999. The Standard Offset Cup Impactor is a reusable instrument used during acetabular replacement surgery to implant an acetabular cup. The Standard Offset Cup Impactor is provided non-sterile and must be sterilized prior to use in a surgery. The Standard Offset Cup Impactor is used to implant cups during many minimally invasive surgical approaches to Total Hip Arthroplasty ( ri-IA). EXPANDED 3-28-2014: The following Part Numbers are also affected: P10270, P6427, P6464, P6501, P7611, P7640, P7794, P7927, P8233, P8590, T10006, T10261, T10606, T10835, T11095, T11506, T11622, T12081, T12083, T12124, T12125, T13435, T13527, T14174, T15262, T15515, T16418, T16551, T16552, T16553, T17272, T5709, T6744, T7751, T9195.
Published March 12, 2014 · Recall initiated: January 31, 2014
Summary
FDA records medical device Class I recall Z-1119-2014, published March 12, 2014 and initiated January 31, 2014. Recalling firm: Greatbatch Medical; Status: Terminated; Reported quantity: 4870 worldwide (2482 US), EXPANDED: 7245 devices; Lot: all lots. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Greatbatch Medical has initiated a global Field Corrective Action for the Standard Offset Cup Impactor since the product was first released in 2004. Greatbatch has identified through internal Sterility Assurance Level validation testing that a sterility assurance level of 10-6 is not achieved on the Standard Offset Cup Impactor when it is subjected to the steam sterilization cycles and associated dry times that are specified in the current Instructions For Use (IFU) provided with the device. The Standard Offset Cup Impactor is provided non-sterile and must be sterilized prior to use in a surgery. Greatbatch has not received any reports of deaths, illnesses, injuries, or other adverse effects associated with this issue.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Greatbatch Medical
- Distributor
- Worldwide Distribution - USA including CA, IN, KS, MA, NJ, TN,TX,UT and Internationally to Australia, Austria, Belgium, China, France, Germany, Greece, Ireland, Italy, Japan, Luxembourg, Netherlands, Norway, South Korea, Spain, Sweden, Switzerland, Taiwan, United Kingdom.
- Lot
- all lots
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABiosense Webster MobiCath Bi-Directional Guiding Sheath, Model D140010 (small curve) and D140011 (largeAugust 9, 2017
- FDAViaPeel PTFE Peelable Introducer 5FR, Part Number 10890-006August 9, 2017
- FDALinear Straight Broach Handle (Rasp Handle) (Greatbatch Part # 510397; DJO Surgical Inc., # 803-03-159) and EZMay 17, 2017
- FDAGreatbatch Stiffer Coaxial Micro-Introducer. Model #/ Name: 10904-001 Stiffer Coaxial Micro-Introducer 4FR;November 30, 2016
- FDATSX-101A Aquilion 32/64 ONE Computed Tomography X-Ray System; TXS-101/R Aquilion RXL Computed Tomography X-RayMarch 12, 2014
- FDATriton FP Infusion Pump, Part Number 400000, packaged in boxes. The Triton Volumetric Infusion Pump isMarch 12, 2014
Questions about this recall
Was Greatbatch Medical, Offset Cup Impactor. Part Numbers: P9485, T10243, T10281, T10287, T10484, T10491, T10507, T10600, T10604, T10753, T10861, T11209, T11340, T12121, T12230, T12288, T12305, T12391, T12699, T12767, T12861, T13344, T13480, T13642, T13722, T13999, T14155, T14384, T15311, T15752, T15822, T15949, T16117, T16427, T16604, T16611, T16661, T16829, T16934, T17062, T17203, T17238, T17321, T6318, T7821, T8042, T8043, T8044, T8087, T8088, T8093, T8095, T8177, T8184, T8277, T8310, T8333, T8468, T8487, T9109, T9196, T9205, T9316, T9348, T9360, T9556, T9557, T9747, T9894, T9954, T9955, T9999. The Standard Offset Cup Impactor is a reusable instrument used during acetabular replacement surgery to implant an acetabular cup. The Standard Offset Cup Impactor is provided non-sterile and must be sterilized prior to use in a surgery. The Standard Offset Cup Impactor is used to implant cups during many minimally invasive surgical approaches to Total Hip Arthroplasty ( ri-IA). EXPANDED 3-28-2014: The following Part Numbers are also affected: P10270, P6427, P6464, P6501, P7611, P7640, P7794, P7927, P8233, P8590, T10006, T10261, T10606, T10835, T11095, T11506, T11622, T12081, T12083, T12124, T12125, T13435, T13527, T14174, T15262, T15515, T16418, T16551, T16552, T16553, T17272, T5709, T6744, T7751, T9195. recalled?
Yes. FDA published recall Z-1119-2014 on March 12, 2014. Reported hazard: Greatbatch Medical has initiated a global Field Corrective Action for the Standard Offset Cup Impactor since the product was first released in 2004. Greatbatch has identified through internal Sterility Assurance Level validation testing that a sterility assurance level of 10-6 is not achieved on the Standard Offset Cup Impactor when it is subjected to the steam sterilization cycles and associated dry times that are specified in the current Instructions For Use (IFU) provided with the device. The Standard Offset Cup Impactor is provided non-sterile and must be sterilized prior to use in a surgery. Greatbatch has not received any reports of deaths, illnesses, injuries, or other adverse effects associated with this issue.
What should I do if I have this product?
Voluntary: Firm initiated · Letter