UNiD" Adaptive Spine Intelligence UNiD Spine Analyzer Product Number SW3002 UDI-DI code 03613720286929 Two cloud-based software applications within UNiD Adaptive Spine Intelligence (ASI): the UNiD HUB and the UNiD Spine Analyzer. The UNiD HUB cloud-based software is a healthcare application to receive, transfer, display, store data needed for planning a spine surgery or for post-operative follow-up (patient information, X-ray image and recommendations for planning). The UNiD Spine Analyzer cloud based software is a healthcare application intended for assisting healthcare professionals in viewing and measuring images as well as planning spine surgeries.
Published February 12, 2025 · Recall initiated: December 3, 2024
Summary
FDA records medical device Class II recall Z-1118-2025, published February 12, 2025 and initiated December 3, 2024. Recalling firm: Medicrea International; Status: Ongoing; Reported quantity: 1,548 users; Lot: UNiD HUB is not a medical device and thus does not have a CFN or GTIN code. UNiD Spine Analyzer Product Number SW3002 UDI-DI code: 03613720286929 No lot numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to software anomalies that may impact on rod planning specifically, certain optional surgical parameters may have had errors that resulted i incorrect calculations displayed on system. Impacted parameters includes the following: Roussouly Classification, Real Lumbar Lordosis and Real Thoracic Kyphosis, Barrey Ratio, and Lenke Classification.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medicrea International
- Distributor
- U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. O.U.S.: Argentina Australia Belgium Brazil Bulgaria Chile Czech Republic Denmark Finland France Germany Greece Iceland Ireland Israel Italy Luxembourg Malaysia Mexico Monaco Morocco Netherlands New Zealand Poland Portugal Singapore South Korea Spain Sweden Switzerland United Kingdom
- Lot
- UNiD HUB is not a medical device and thus does not have a CFN or GTIN code. UNiD Spine Analyzer Product Number SW3002 UDI-DI code: 03613720286929 No lot numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- France
Related recalls
- FDAIrbesartan Tablets USP, 300 mg, 50 Tablets (5 x 10) Unit Dose per carton, Manufactured for: AvKARE, Pulaski,February 19, 2025
- FDANestle Toll House Valentine's Chocolate Chip cookie dough, Makes 20 cookies. Net wt 14 oz. UPC 0 28000-34408February 19, 2025
- FDAIntelliVue Patient Monitor MX550. Model Number: 866066February 12, 2025
- FDAThe Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN Specimen Collection kitFebruary 12, 2025
- FDAMedicrea International Universal Counter Torque for use with the Pass TULIP [Spinal] System Ref : A22020500February 26, 2020
- FDAMedicrea Pass LP, REF B02217545, Non Sterile, Material: Ti-6Al-4V ELI, VIS PEDICULAIRE POLYAXIALE 7.5MM xJanuary 8, 2020
Questions about this recall
Was UNiD" Adaptive Spine Intelligence UNiD Spine Analyzer Product Number SW3002 UDI-DI code 03613720286929 Two cloud-based software applications within UNiD Adaptive Spine Intelligence (ASI): the UNiD HUB and the UNiD Spine Analyzer. The UNiD HUB cloud-based software is a healthcare application to receive, transfer, display, store data needed for planning a spine surgery or for post-operative follow-up (patient information, X-ray image and recommendations for planning). The UNiD Spine Analyzer cloud based software is a healthcare application intended for assisting healthcare professionals in viewing and measuring images as well as planning spine surgeries. recalled?
Yes. FDA published recall Z-1118-2025 on February 12, 2025. Reported hazard: Due to software anomalies that may impact on rod planning specifically, certain optional surgical parameters may have had errors that resulted i incorrect calculations displayed on system. Impacted parameters includes the following: Roussouly Classification, Real Lumbar Lordosis and Real Thoracic Kyphosis, Barrey Ratio, and Lenke Classification.
What should I do if I have this product?
Voluntary: Firm initiated · Letter