Accuray CyberKnife System Medical Physicist, Robotic Radiosurgery System, Medical charged-particle radiation therapy system, Model: Iris Variable Aperture Collimator P/Ns 028986 and 031777. The CyberKnife System is indicated for radiation treatment.
Published April 24, 2013 · Recall initiated: March 19, 2013
Summary
FDA records medical device Class II recall Z-1126-2013, published April 24, 2013 and initiated March 19, 2013. Recalling firm: Accuray Incorporated; Status: Terminated; Reported quantity: 176 units; Lot: All units of these models. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A defect on the master-side connector block of the pneumatic tool changer could introduce a tilt in the collimator housings, resulting in a potential shift to the central axis of the radiation beam. The issue could pose significant clinical impact to plans created using the smallest collimators.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Accuray Incorporated
- Distributor
- Worldwide distribution, including Nationwide in the US and the countries of Korea, Myanmar, Canada, Belgium, England, Czech Republic, Finland, France, Germany, Greece, India, Italy, Poland, Russia, Spain, Switzerland, Turkey, Columbia, Mexico, Japan and Saudi Arabia.
- Lot
- All units of these models.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear AcceleratorJuly 8, 2026
- FDACyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear AcceleratorJuly 8, 2026
- FDACyberKnife Treatment Delivery System - Product Usage: indicated for image-guided stereotactic radiosurgery andMarch 24, 2021
- FDACyberKnife Robotic Radiosurgery Systems, models: G3, G4, VSI. Radiology: The CyberKnife Robotic RadiosurgeryDecember 28, 2016
- FDAInstrumentation Industries BE183-SU and BE 183 SU-12 Manual Jet Ventilators; BE 183-SUR and BE 183 SUR-12April 24, 2013
- FDAThe Oxford Partial Knee; Oxford Knee Resection Procedure 3 Pack CEMENTED For use with Stryker System 5, SystemApril 24, 2013
Questions about this recall
Was Accuray CyberKnife System Medical Physicist, Robotic Radiosurgery System, Medical charged-particle radiation therapy system, Model: Iris Variable Aperture Collimator P/Ns 028986 and 031777. The CyberKnife System is indicated for radiation treatment. recalled?
Yes. FDA published recall Z-1126-2013 on April 24, 2013. Reported hazard: A defect on the master-side connector block of the pneumatic tool changer could introduce a tilt in the collimator housings, resulting in a potential shift to the central axis of the radiation beam. The issue could pose significant clinical impact to plans created using the smallest collimators.
What should I do if I have this product?
Voluntary: Firm initiated · Letter