EndoWrist One Vessel Sealer, IS3000, 6 pack used in conjunction with the da Vinci Surgical System Si IS3000 The EndoWrist One Vessel Sealer is a bipolar electrosurgical instrument for use with the da Vinci Surgical System and the ERBE VIO 300 D electrosurgical generator. It is intended for bipolar coagulation and mechanical transection of vessels up to 7mm in diameter and tissue bundles that fit the jaws of the instrument. PN 410322-05
Published March 4, 2015 · Recall initiated: February 11, 2015
Summary
FDA records medical device Class II recall Z-1141-2015, published March 4, 2015 and initiated February 11, 2015. Recalling firm: Intuitive Surgical, Inc.; Status: Terminated; Reported quantity: 3282 total units, all part numbers; Lot: PN 410322-05Serial numbers: M10140807 M10140926 M10141028 M10141119 M11140917 M11141119 M10140812 M10141001 M10141029 M10141125 M11140922 M11141125 M10140814 M10141004 M10141031 M10141126 M11140926 M11141126 M10140818 M10141006 M10141104 M10141128 M11141001 M11141201 M10140826 M10141009 M10141105 M10141201 M11141004 M10140828 M10141015 M10141107 M11140808 M11141006 M10140902 M10141020 M10141111 M11140818 M11141009 M10140905 M10141022 M10141113 M11140826 M11141015 M10140917 M10141023 M10141114 M11140828 M11141020 M10140922 M10141024 M10141117 M11140904 M11141024. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Intuitive Surgical has identified the potential for a certain, small portion of Vessel Sealers to exhibit interference in the back of the instrument grips due to manufacturing variability in some jaws.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Intuitive Surgical, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to Australia, Belgium, China, Denmark, Ecuador, Finland, France, Germany, Greece, Italy, Netherlands, Norway, Puerto Rico, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
- Lot
- PN 410322-05Serial numbers: M10140807 M10140926 M10141028 M10141119 M11140917 M11141119 M10140812 M10141001 M10141029 M10141125 M11140922 M11141125 M10140814 M10141004 M10141031 M10141126 M11140926 M11141126 M10140818 M10141006 M10141104 M10141128 M11141001 M11141201 M10140826 M10141009 M10141105 M10141201 M11141004 M10140828 M10141015 M10141107 M11140808 M11141006 M10140902 M10141020 M10141111 M11140818 M11141009 M10140905 M10141022 M10141113 M11140826 M11141015 M10140917 M10141023 M10141114 M11140828 M11141020 M10140922 M10141024 M10141117 M11140904 M11141024
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAda Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15July 22, 2026
- FDADa Vinci ASSY, PSS, SP1098, Part Number: 380601July 1, 2026
- FDABrand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/AJune 10, 2026
- FDAIntuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06May 13, 2026
- FDAMetasul Femoral Head - A modular head component is used in conjunction with a femoral stem component forMarch 4, 2015
- FDAThe Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in theMarch 4, 2015
Questions about this recall
Was EndoWrist One Vessel Sealer, IS3000, 6 pack used in conjunction with the da Vinci Surgical System Si IS3000 The EndoWrist One Vessel Sealer is a bipolar electrosurgical instrument for use with the da Vinci Surgical System and the ERBE VIO 300 D electrosurgical generator. It is intended for bipolar coagulation and mechanical transection of vessels up to 7mm in diameter and tissue bundles that fit the jaws of the instrument. PN 410322-05 recalled?
Yes. FDA published recall Z-1141-2015 on March 4, 2015. Reported hazard: Intuitive Surgical has identified the potential for a certain, small portion of Vessel Sealers to exhibit interference in the back of the instrument grips due to manufacturing variability in some jaws.
What should I do if I have this product?
Voluntary: Firm initiated · Letter