ABX PENTRA Multical. ABX PENTRA N Control. ABX PENTRA P Control. The ABX PENTRA 400 and PC200 are benchtop clinical chemistry analyzers using two measuring principals: absorbance and ion selective electrodes. The ABX PENTRA Multical is a lyophilized human serum calibrator with chemical additives and materials of biological origin. The assigned values of the calibrator components are given in the enclosed annex, ensuring optimal calibration of the appropriate HORIBA ABX SAS methods on the ABX PENTRA Clinical Chemistry Analyzer. This calibrator is provided in ten vials of 3 ml.
Published March 12, 2014 · Recall initiated: January 20, 2014
Summary
FDA records medical device Class II recall Z-1146-2014, published March 12, 2014 and initiated January 20, 2014. Recalling firm: Horiba Instruments, Inc dba Horiba Medical; Status: Terminated; Reported quantity: 2,023 units; Lot: Model Numbers:Multical - A11A01652, N Control - A11A01653, P Control - A11A01654. Lot Numbers: MultiCal: 1204101 and 1304901, N Control: 1300801, P Control: 1300601. Expiration Date (s): MultiCal: 1204101 Expires on 05/31/14, MultiCal: 1304901 Expires on 07/15/15, N Control: 1300801 Expires on 11/05/14, P Control: 1300601 Expires on 11/05/14. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
HORIBA Medical is informing all ABX PENTRA MultiCal, N Control, and P Control customers running the Creatinine Rate Blank Method assay on the ABX PENTRA 400 and PC200 that the Target Values for Creatinine Rate Blank Method have been updated. This issue is only applicable to Creatinine Rate Blank Method values reported on the ABX PENTRA 400 and PC200 Chemistry Analyzers only. No other analyzers
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Horiba Instruments, Inc dba Horiba Medical
- Distributor
- United States nationwide distribution.
- Lot
- Model Numbers:Multical - A11A01652, N Control - A11A01653, P Control - A11A01654. Lot Numbers: MultiCal: 1204101 and 1304901, N Control: 1300801, P Control: 1300601. Expiration Date (s): MultiCal: 1204101 Expires on 05/31/14, MultiCal: 1304901 Expires on 07/15/15, N Control: 1300801 Expires on 11/05/14, P Control: 1300601 Expires on 11/05/14.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAABX Pentra Micro ALBUMIN 2 CP, Serum/Plasma, REF 1300032563 The device is intended for use in conjunction withJuly 4, 2018
- FDAABX PENTRA Magnesium RTU Reagent, Part No. A11A01646 Product Usage: ABX PENTRA Magnesium RTU reagent withOctober 1, 2014
- FDADevice Name: ABX PENTRA ML Model/Part #(s): A11A01652 Lot/Serial #(s): N/A Software Version(s): All versionsMarch 19, 2014
- FDASimplexa Flu AlB & RSV Direct Assay Kits, intended for use as an aid in the differential diagnosis ofMarch 12, 2014
- FDAGE Healthcare Solar 8000M and Solar 8000i Patient Monitor software version 5.4, 5.5, or 5.6 with theMarch 12, 2014
- FDAABX PENTRA N Control, Model No. A11A01653 The ABX PENTRA N Control is for use in quality control by monitoringFebruary 12, 2014
Questions about this recall
Was ABX PENTRA Multical. ABX PENTRA N Control. ABX PENTRA P Control. The ABX PENTRA 400 and PC200 are benchtop clinical chemistry analyzers using two measuring principals: absorbance and ion selective electrodes. The ABX PENTRA Multical is a lyophilized human serum calibrator with chemical additives and materials of biological origin. The assigned values of the calibrator components are given in the enclosed annex, ensuring optimal calibration of the appropriate HORIBA ABX SAS methods on the ABX PENTRA Clinical Chemistry Analyzer. This calibrator is provided in ten vials of 3 ml. recalled?
Yes. FDA published recall Z-1146-2014 on March 12, 2014. Reported hazard: HORIBA Medical is informing all ABX PENTRA MultiCal, N Control, and P Control customers running the Creatinine Rate Blank Method assay on the ABX PENTRA 400 and PC200 that the Target Values for Creatinine Rate Blank Method have been updated. This issue is only applicable to Creatinine Rate Blank Method values reported on the ABX PENTRA 400 and PC200 Chemistry Analyzers only. No other analyzers
What should I do if I have this product?
Voluntary: Firm initiated · Letter