Smith & Nephew RENASYS Soft Port REF: 66800799 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)
Published March 19, 2014 · Recall initiated: January 9, 2014
Summary
FDA records medical device Class II recall Z-1157-2014, published March 19, 2014 and initiated January 9, 2014. Recalling firm: Smith & Nephew Inc.; Status: Terminated; Reported quantity: 162,629 invididual units; Lot: ALL LOT #s 2013010311 and lower. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Smith & Nephew Inc.
- Distributor
- Worldwide Distribution-USA (nationwide) including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, KY, LA, MA, MS, MT, NC, NJ, NV, NY, OH, OR, PA, TN, TX, VA, WA, WI, WV, HI and the countries Dubai, Curacao, Argentina, Australia, Chile, Germany, France, England, Hong Kong, Kuwait, Lebanon, Libya, Sri Lanka, Malaysia, New Zealand, Manila, Saudi Arabia, South Africa, Singapore, Turkey, Taipei and Vietnam.
- Lot
- ALL LOT #s 2013010311 and lower
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABIORAPTOR Suture Anchors. Tendon/ligament, non-bioabsorbable bone anchor.December 11, 2024
- FDAROTATION MEDICAL TENDON STAPLES (8) (broad indication)-Intended for fixation of prosthetic material to softSeptember 18, 2024
- FDAROTATION MEDICAL TENDON STAPLES (8) (narrow indication)-Intended for fixation of prosthetic material to softSeptember 18, 2024
- FDASmith & Nephew RENASYS-F Foam Dressing Kit with Soft Port REF #'s: 66800794, 66800795, and 66800796 (TheMarch 19, 2014
- FDAPanaroma Central Station Product Usage: This device can view real time, store, print, graph, and trend patientMarch 19, 2014
- FDAPhilips HeartStart MRx Monitor/Defibrillator Model numbers: M3535A, M3536A,March 19, 2014
Questions about this recall
Was Smith & Nephew RENASYS Soft Port REF: 66800799 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps) recalled?
Yes. FDA published recall Z-1157-2014 on March 19, 2014. Reported hazard: Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo
What should I do if I have this product?
Voluntary: Firm initiated