ROTATION MEDICAL TENDON STAPLES (8) (broad indication)-Intended for fixation of prosthetic material to soft tissues. Product Number: 2504-1
Published September 18, 2024 · Recall initiated: August 12, 2024
Summary
FDA records medical device Class II recall Z-3096-2024, published September 18, 2024 and initiated August 12, 2024. Recalling firm: Smith & Nephew Inc.; Status: Ongoing; Reported quantity: US: 6305 units; OUS : 160 units; Lot: UDI-DI: 00854501006067 Batch Number 51180055 51180056 51180057 51180058 51180059 51180060 51180061 51180062 51180063 51180064 51180065 51184280 51184281 51184282 51184283 51184284 51184285 51184286 51184287 51184288 51184289 51184290 51185286 51185287 51185288 51185289 Batch Number 51185290 51185291 51185292 51185293 51185294 51185295 51185296 51189247 51189249 51189250 51189251 51189252 51189254 51189255 51189256 51189257 51189258 51189259 51193811 51193816 51193818 51193820. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Product packaging process may result in an improper or incomplete seal of the outer foil pouch surrounding the inner Tyvek pouch, compromising product sterility
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Smith & Nephew Inc.
- Distributor
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, Chile, China, France, Germany, Great Britain, Greece, Hong Kong, India, Italy, Japan, Lithuania, Malaysia, Malta, New Zealand, Poland, Portugal, Puerto Rico, Romania, Singapore, Slovenia, South Africa, Spain, Switzerland, United Arab Emirates/Dubai (UAE).
- Lot
- UDI-DI: 00854501006067 Batch Number 51180055 51180056 51180057 51180058 51180059 51180060 51180061 51180062 51180063 51180064 51180065 51184280 51184281 51184282 51184283 51184284 51184285 51184286 51184287 51184288 51184289 51184290 51185286 51185287 51185288 51185289 Batch Number 51185290 51185291 51185292 51185293 51185294 51185295 51185296 51189247 51189249 51189250 51189251 51189252 51189254 51189255 51189256 51189257 51189258 51189259 51193811 51193816 51193818 51193820
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABIORAPTOR Suture Anchors. Tendon/ligament, non-bioabsorbable bone anchor.December 11, 2024
- FDAROTATION MEDICAL TENDON STAPLES (8) (narrow indication)-Intended for fixation of prosthetic material to softSeptember 18, 2024
- FDAMedline ReNewal Reprocessed Ethicon Harmonic ACE+7 Shears w/ Adv. Hemostasis Use Only w/GEN11 SW v2018-1 andSeptember 18, 2024
- FDAMedline ReNewal Reprocessed Covidien LigaSure Impact, Sealer/Divider, Nano-Coated, Compatible w/ FT10 18cmSeptember 18, 2024
- FDASmith & Nephew RENASYS Soft Port REF: 66800799 (The RENASYS Soft Port Dressing Kits are used in conjunctionMarch 19, 2014
- FDASmith & Nephew RENASYS-F Foam Dressing Kit with Soft Port REF #'s: 66800794, 66800795, and 66800796 (TheMarch 19, 2014
Questions about this recall
Was ROTATION MEDICAL TENDON STAPLES (8) (broad indication)-Intended for fixation of prosthetic material to soft tissues. Product Number: 2504-1 recalled?
Yes. FDA published recall Z-3096-2024 on September 18, 2024. Reported hazard: Product packaging process may result in an improper or incomplete seal of the outer foil pouch surrounding the inner Tyvek pouch, compromising product sterility
What should I do if I have this product?
Voluntary: Firm initiated · Letter