Synapse CV 6. with AR. A web-based application as the primary user interface for the processing of medical images.
Published February 28, 2024 · Recall initiated: January 3, 2024
Summary
FDA records medical device Class II recall Z-1172-2024, published February 28, 2024 and initiated January 3, 2024. Recalling firm: FUJIFILM Healthcare Americas Corporation; Status: Ongoing; Reported quantity: 8; Lot: UDI-DI 00854904006015 Software versions 6.2.1, 6.2.0, 6.1.1, 6.1.0, 6.0.1, and 6.0.0. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- FUJIFILM Healthcare Americas Corporation
- Distributor
- US domestic distribution nationwide. International distribution to Israel.
- Lot
- UDI-DI 00854904006015 Software versions 6.2.1, 6.2.0, 6.1.1, 6.1.0, 6.0.1, and 6.0.0
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFUJIFILM Synapse PACS software v7.4.310; radiological image processing systemSeptember 2, 2026
- FDAThe device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2October 2, 2024
- FDAFDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinalJune 5, 2024
- FDANoblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24May 29, 2024
- FDASynapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface forFebruary 28, 2024
- FDAOPTIMIZER model CCM X11 implantable pulse generator (IPG) devices - Smart Mini and LiteFebruary 28, 2024
Questions about this recall
Was Synapse CV 6. with AR. A web-based application as the primary user interface for the processing of medical images. recalled?
Yes. FDA published recall Z-1172-2024 on February 28, 2024. Reported hazard: The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.
What should I do if I have this product?
Voluntary: Firm initiated · Letter