ORTHOCORD One Violet and One Blue Braided Composite Suture with MO-7 Taper Needles, Catalog Number: 223114. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.
Published May 1, 2013 · Recall initiated: March 22, 2013
Summary
FDA records medical device Class II recall Z-1174-2013, published May 1, 2013 and initiated March 22, 2013. Recalling firm: DePuy Mitek, Inc., a Johnson & Johnson Co.; Status: Terminated; Reported quantity: 206540; Lot: ALL from Jan 1 2010 - Corrected IFU implementation date. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- DePuy Mitek, Inc., a Johnson & Johnson Co.
- Distributor
- Nationwide Distribution.
- Lot
- ALL from Jan 1 2010 - Corrected IFU implementation date
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHealix Knotless ADV BR 4.75 Suture AnchorDecember 18, 2019
- FDADePuy Mitek LATARJET EXPERIENCE Coracoid Drill Product Code: 288201 Product Usage: The Latarjet Cortical ScrewJune 28, 2017
- FDADePuy Mitek LATARJET EXPERIENCE Extraction Driver Product Code: 288216 Product Usage: The Latarjet CorticalJune 28, 2017
- FDADePuy Mitek LATARJET EXPERIENCE Top Hat Tap Product Code: 288203 Product Usage: The Latarjet Cortical ScrewJune 28, 2017
- FDAORTHOCORD ONE VIOLET AND ONE BLUE BRAIDED COMPOSITE SUTURE 2 (5 METRIC), Catalog Number: 223113. For use inMay 1, 2013
- FDAORTHOCORD One Violet and One Blue Braided Composite Suture with MO-6 Taper Needles, Product Code: 223115. ForMay 1, 2013
Questions about this recall
Was ORTHOCORD One Violet and One Blue Braided Composite Suture with MO-7 Taper Needles, Catalog Number: 223114. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures. recalled?
Yes. FDA published recall Z-1174-2013 on May 1, 2013. Reported hazard: The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six
What should I do if I have this product?
Voluntary: Firm initiated · Letter