FDAFDA · Food and drugs·Recall number Z-1176-2015

NAMIC Convenience Kit, RADIOLOGY KIT, UPN H749601506921, REF/Catalog No. 60150692, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.

Published March 4, 2015 · Recall initiated: January 9, 2015

Summary

FDA records medical device Class II recall Z-1176-2015, published March 4, 2015 and initiated January 9, 2015. Recalling firm: Navilyst Medical, Inc; Status: Terminated; Reported quantity: 22 units; Lot: Lots 4271315, 4310125, 4604147, 4655162, 4706282, 4745859, 4788727. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Navilyst Medical, Inc
Distributor
Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
Lot
Lots 4271315, 4310125, 4604147, 4655162, 4706282, 4745859, 4788727.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was NAMIC Convenience Kit, RADIOLOGY KIT, UPN H749601506921, REF/Catalog No. 60150692, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems. recalled?

Yes. FDA published recall Z-1176-2015 on March 4, 2015. Reported hazard: The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

What should I do if I have this product?

Voluntary: Firm initiated · Letter